Elitech ENVOY 500 HDL Manual

Cholesterol reagent kit

Advertisement

Quick Links

ENVOY
®
500 HDL CHOLESTEROL REAGENT KIT
APPLICATION PARAMETERS
PRIMARY PARAMETERS
CHECK PARAMETERS
Reagent Limit (mABS)
Code
HDL
Curve Acceptance (%)
Bar-Code
Active
RE-RUN SERUM
Code for Bar-Code
310
Test Limit (Conc)
Test Methodology
Selective
Low Test Limit (Conc)
Method
Sample Blank (A)
Initial ABS (mABS)
Kind of Process
Linear
Final ABS (mABS)
1st Filter
578
Max ABS Delta (mABS)
2nd Filter
700
Prozone Check
Reaction direction
Increasing
Normal Range
REAGENTS
Man
Number of reagents
2
Woman
Reagent 1 Volume μL
300
Child
Concentrated
Inactive
Re-run hyperactive
Reagent 2 Volume μL
100
Re-run pathological
Concentrated
Inactive
RE-RUN URINE
Serum
Urine
SAMPLE
Test Limit (Conc)
Name
HDL
Low Test Limit (Conc)
Cholesterol
Initial ABS (mABS)
Sample μL
4
N/A
Final ABS (mABS)
Pre-Dilution 1:
1
N/A
Max ABS Delta (mABS)
Post-Dilution 1:
1
N/A
Prozone Check
TIMES
Normal Range
Sample Starter
Inactive
Man
Delay Time
0
Woman
Reading Time
60
Child
Reagent 1 Incubation Time
300
Re-run hyperactive
Reagent 2 Incubation Time
240
Re-run pathological
PROGRAMMING INSTRUCTIONS
Detailed instructions for programming reagent parameters are provided in
the Envoy 500 Operator manual and Envoy 500 Settings Table.
If the Envoy 500 Chemistry System is not pre-programmed, a HDL choles-
terol code must first be added before the parameters can be entered. On
the menu bar, select «Test  Test Directory.» A new window will open up
listing all the codes for the tests that are installed on the instrument. Click on
the «New Code» button, type «HDL» into the Code field and select «Save.»
To program the application parameters, check the box next to the code for
the HDL test, and select the «Parameters» button located at the bottom of
the window. To program standard information, click the «Standards» button
located at the bottom of the window.
For Technical questions, Please call or contact
ELITech Clinical Systems SAS
Zone Industrielle
(855) 354-8324 - www.elitechgroup.com
61500 SEES
27 Wellington Road
France
Lincoln, Rhode Island 02865 - U.S.A.
SECONDARY PARAMETERS
1
Unit Serum
mg/dL
200
st
2
nd
Unit Serum
Inactive
100
1
Unit Urine
N/A
st
2
nd
Unit Urine
Inactive
150
Dynamic Blank
Active
1.1
Needle washes
[From Settings Table]
N/A
Cuvette washes
[From Settings Table]
N/A
Special wash
[From Settings Table]
N/A
Instrumental Factor
1.000
Inactive
Shift
0.000
Min
Max
Reagent Blank
Every Day
[User defined]
Decimals
0
[User defined]
[User defined]
STANDARD PARAMETERS
Inactive
Factor
[Determined by calibration]
Inactive
Minimum
270
Maximum
900
N/A
No. of Samples
1
N/A
Max Var. (%)
10
N/A
Timed re-run
Inactive
N/A
N. replicates
3
N/A
Inactive
Reagents ABS
[Determined by Envoy]
Min
Max
Pos.
[From Settings Table]
N/A
Conc.
[From calibrator label]
N/A
ABS
[Determined by Envoy]
N/A
% last calibration
100
Inactive
Inactive
FTEVY-HDLL-v5
ENVOY
®
500 HDL CHOLESTEROL REAGENT KIT
Product no.
Product no. 55301
For in vitro diagnostic use
For in vitro diagnostic use
 CAUTION : Federal Law restricts this device to sale by or on the order of
a licensed healthcare practitioner (Rx ONLY)
INTENDED USE
Envoy® 500 HDL Cholesterol Reagent is for the quantitative determina-
tion of high density lipoprotein (HDL) cholesterol in serum and plasma
on Envoy 500 Series Analyzers.
SUMMARY
The principle role of high density lipoproteins (HDL) in lipid metabolism
is the uptake and transport of cholesterol from the peripheral tissues
to the liver through a process known as reverse cholesterol transport,
which is a proposed cardio-protective mechanism.
levels are associated with an increased risk of coronary heart disease
and coronary artery disease. Consequently the determination of serum
HDL-cholesterol is a useful tool for identifying high risk patients.
HDL-cholesterol results may also be indicative of various lipid disorders
such as diabetes mellitus and other liver and renal diseases.
HDL cholesterol is measured using a two step process.
Accelerator + CO + POD + DSBmT
Step 1:
HDL, LDL, VLDL, Chylomicrons
Non Reactive LDL, VLDL, Chylomicrons
Step 2:
HDL Specific Detergent
HDL-Cholesterol
CO + CE
HDL-Cholesterol + O
Cholest-4-ene-3-one + H
2
2 H
O
+ 4-Aminoantipyrine + DSBmT
2
2
Non HDL-esterified and free cholesterol are consumed by cholesterol
oxidase, peroxidase and DSBmT in step 1 yielding colorless products.
HDL cholesterol is unaffected. Reagent 2 contains cholesterol esterase,
a chromogenic coupler, and a detergent capable of selectively solubiliz-
ing the HDL cholesterol. In step 2, this reagent is added and the HDL
cholesterol reacts to produce a chromogen that absorbs at 578 nm.
The change in absorbance is proportional to the concentration of HDL
cholesterol in the sample.
This reaction scheme may be referred to as the Accelerator Selective
Detergent methodology.
REAGENTS
COMPOSITION
HDL Cholesterol Reagent 1 contains < 1,000 U/L cholesterol oxidase (E.
coli), < 1,300 ppg U/L peroxidase (horseradish), < 1 mmol/L disodium N, N-bis
(4-sulfobutyl)-m-toluidine, < 1 mmol/L accelerator, < 0.06% preser-
vative, < 3,000 U/L ascorbate oxidase (Curcubita), buffer, and other
ingredients. HDL Cholesterol Reagent 2 contains < 1,500 U/L choles-
terol esterase (Pseudomonas sp.), < 1 mmol/L 4-aminoantipyrine,
< 2% detergent, < 0.06% preservative, buffer, and other ingredients.
WARNINGS AND PRECAUTIONS
- This reagent is for professional in vitro diagnostic use only.
- Take normal precautions and adhere to good laboratory practice.
- Use clean or single use laboratory equipment only to avoid contam-
inations.
- Dispose of contents in accordance with all local, state and federal
regulations.
- For more information, Safety Data Sheet (SDS) is available on
request for professional user.
ELITech Clinical Systems SAS
(10/2015)
Zone Industrielle
61500 SEES
France
PREPARATION
Both HDL Cholesterol Reagent 1 and HDL Cholesterol Reagent 2 are
ready for use as packaged.
STORAGE AND STABILITY
Store these reagents at 2 to 8 °C. Do not freeze. Unopened reagents are
stable to the expiration dates on the bottle labels.
Open reagents are stable for 28 days onboard the Envoy 500 Chemistry
System.
SPECIMENS
SERUM AND PLASMA COLLECTION AND STORAGE
1
- Fasting serum is the preferred specimen. Fasting heparinized plasma is
Low HDL-cholesterol
also acceptable. Do not analyze whole blood. Collect specimens by veni-
puncture according to accepted clinical protocol. Patients should maintain
their usual diet for at least two weeks before blood collection. Blood should
be drawn after a 12 hour fast and after the subject has been sitting quietly
2
for at least 5 minutes. Separate the serum or plasma sample from the cells
within three hours of collection.
- Venipuncture should be performed prior to the administration of drugs. Of
particular note, venipuncture performed during an acetaminophen overdose
situation, when N-acetyl-p-benzoquinone imine (NAPQI) an atypical meta-
bolic breakdown product of acetaminophen, may be present, may lead to
erroneously low HDL Cholesterol results. Venipuncture performed during or
immediately after administration of N-acetylcysteine (NAC), a drug used to
HDL Disrupted
treat acetaminophen overdose, or Metamizole may lead to erroneously low
HDL cholesterol results.
O
- For best results, HDL should be analyzed on the day of collection. HDL
2
2
POD
Cholesterol in serum and plasma is stable for up to two days at 2 to 8 °C,
Colored End Product
one month at -20 °C, or 2 years at -70 °C. Once thawed, the specimen may
3
not be refrozen.
COMPATIBLE ADDITIVES
Acceptable chemical preservatives are lithium and sodium heparin. Do not
use anticoagulants containing citrate or any other chemical additives.
PROCEDURE
MATERIALS PROVIDED
The Envoy 500 HDL Cholesterol Reagent Kit includes the following compo-
nents:
4 x 30.4 mL boats of Envoy 500 HDL Cholesterol Reagent 1
4 x 11.4 mL bottles of Envoy 500 HDL Cholesterol Reagent 2
MATERIALS REQUIRED BUT NOT PROVIDED
Envoy 500 HDL Calibrator (product no. 55119)
Envoy 500 Serum Controls (product no. 55131)
Normal saline (0.85% saline)
ASSAY PROCEDURE
Program the instrument using the application parameters and programming
instructions provided at the end of this Instructions For Use.
REAGENT INSTALLATION AND USE
The Envoy 500 HDL Cholesterol Reagent is ready to use as packaged.
Snap the small reagent bottle onto the reagent boat if it has become
dislodged during shipping. Mix the reagents by gently inverting the
assembled boat several times.
Record the installation date on the label and insert the assembled boat
into the designated position on the reagent tray.
Let the reagent equilibrate on the instrument for at least 30 minutes
before use.
For Technical questions, Please call or contact
(855) 354-8324 - www.elitechgroup.com
27 Wellington Road
Lincoln, Rhode Island 02865 - U.S.A.
3
.../...
(10/2015)
FTEVY-HDLL-v5