Temperature Meter 430-020 Contents Section Page Safety Introduction General Instrument Specification Installation Unpacking Connection Display and Controls Temperature Adjustment Manual Temperature Adjustment Automatic Temperature Correction (ATC) Calibration and Measurement Care and Maintenance Battery Replacement Spare Parts Service and Support...
The fault condition should immediately be reported to the appropriate servicing authority. Introduction 2.1 General The 430-020 is a general purpose hand-held pH/mV and temperature meter offering 2 point calibration and automatic buffer recognition. This model simultaneously displays either temperature compensated pH readings or electrode potential and temperature.
Temperature Meter 430-020 2.2 Instrument Specification Model 430-020 Type pH/mV and temperature Waterproof Yes (IP67) pH Range -2 to 16pH pH Resolution 0.01pH pH Accuracy +/-0.02pH mV Range -1000 to 1000mV mV Resolution mV Accuracy +/- 1mV 0 to 100°C Temperature Range -39.9 to 149.9°C...
Temperature Meter 430-020 Installation 3.1 Unpacking Remove the 430-020 from the packaging and ensure the following items are included: Model 430-020 pH and temperature meter Epoxy bodied combination pH electrode (430-020/10) Temperature probe (430-020/12) pH 4, 7 and 10 buffer sachets...
Page 6
Temperature Meter 430-020 Fig. 1 - Display and Controls Display showing pH or electrode potential and temperature reading BNC connector for pH electrode Connector for temperature probe UP arrow Key for adjusting values DOWN arrow key for adjusting values...
±1000mV. Temperature Adjustment Model 430-020 is supplied with a temperature probe enabling automatic temperature compensation to be used. If the probe is disconnected then manual temperature compensation can be used instead.
Before performing a measurement the instrument must be calibrated. It is recommended that the instrument is calibrated daily to achieve consistent and accurate results. Model 430-020 has a 2 point calibration process. The first calibration point MUST be buffer pH 7.00. The second calibration point depends on the acidity of the solution which is to be measured.
Page 9
Temperature Meter 430-020 ‘FAIL’ will be shown if 7.00 pH buffer isn’t used as the first calibration point. If the electrode and buffer solution’s combined error is greater than 0.5 pH or 30mV then ‘Error I/P’ will be displayed and the calibration aborted. If the electrode and buffer solution’s combined error is greater than 30mV then...
Tris buffers, high temperatures and strongly acidic solutions a more suitable pH/reference electrode pair may be required. Details or advice are supplied on request at ele@eleint.co.uk. If the electrode readings are slow or erratic, place the electrode into cleaning solution or 7.00 pH solution for at least half an hour to one hour before...
Temperature Meter 430-020 DO NOT touch the sensitive glass pH membrane or reference junction. Excess droplets of solution may be removed by gently blotting with filter paper or tissue. DO NOT rub the electrode as this may induce an electrostatic charge.
10 working days. 8.2 Technical Support ELE International have a dedicated Technical Support team who are on hand to help with any applications advice and questions you may have about our products and how to use them. If you require any technical or application...
Temperature Meter 430-020 Declaration of Conformity pH/mV Temperature Meter, Model 430-020 This product complies with the requirements of the EU Directives listed below: 2004/108/EC EMC Directive. 2011/65/EC RoHS Directive. 9907X0088 Issue 2...