Setting Recorder Outputs - Martek Mark 25 Instruction Manual

Dissolved oxygen analyzer
Table of Contents

Advertisement

viewed and will remain steady if the
channel viewed is in an alarm state.

3.6 Setting Recorder Outputs

The Mark 25 comes standard with four
4-20 mA analog recorder outputs and
one RS232 digital output. Like the alarm
set points, the four analog recorder
outputs can be assigned to any range of
any parameter on any channel.
Press the <CALIB> until the display
indicates the recorder prompt, rEC.
Next, press the <UP ARROW> key. The
display will prompt with ou 1 and all
parameter LED's will be lit. The <UP
ARROW> key can be used to change
the recorder output to 2, 3, or 4.
To assign a particular analog output,
press the <CALIB> key. The Mark 25
will display three bars. The bars indicate
no analog output assigned. Use the
<UP> or <DOWN ARROW> keys to
select temperature (t) or dissolved
oxygen (do).
NOTE: A decimal point will be present
for do which indicates which one of the
four ranges the channel is currently
using. The range can be changed by
pressing the <RANGE> key.
When the desired parameter is
displayed, the standard 4-20 mA output
range can be changed to any value
desired. In addition, a "balance" is
provided to "fine-tune" each end of the
recorder range.
To change the lower end of the recorder
range, press the <ZERO> key. The
Mark 25 will display 4.00. Use the <UP>
or <DOWN ARROW> key to change the
number to the desired value. When the
desired value is reached, press the
<CALIB> key to set the value in
memory.
At this point, the Mark 25 will display
000. This is the "balance". Use the
<UP> or <DOWN ARROW> key to
change the value in order to make the
lower end of the recorder range agree
with the external recorder. When the
desired balance is reached, press the
<CALIB> key to set the value in
memory.
The same procedure is used to change
and "balance" the upper range of the
recorder output. To exit the menu,
simply press the <RUN> key.
To disable any assigned analog
recorder output, select the particular
output and press the <CALIB> key.
Once the parameter is displayed, use
the <UP> or <DOWN ARROW> keys
until three bars are displayed. Press the
<CALIB> key to disable the output.
Once a recorder output has been
assigned to a particular parameter on a
channel, it cannot be assigned to
another parameter or channel until it
has been disabled.
NOTE: When the auto-ranging function
is OFF, changing a dissolved oxygen
range with an assigned analog output
automatically changes the scaling of the
recorder output to the new range.
When the auto-ranging function is ON,
the assigned analog output will remain
in the range it was assigned to.
When the RS232 digital output is used,
calibration is not necessary. However,
the receiving computer must be set up
to accommodate the Mark 25's digital
output configuration:
Baud Rate - 9600
Parity -
Word length - 7
Stop Bit -
10
Even
1