Page 3
KPDM-1 The present users guide (hereinafter UG) is a document according to GOST 2.601 and is applied to the micromanometer with static and dynamic pressure receiver KPDM-1 (hereinafter – the device). UG contains information certifying manufacturer's guarantees, values of basic parameters...
KPDM-1 1 APPLICATION 1.1 The device is applied for receipt and measurement of dynamic and static pressure (p) in air and gas flues, speed determination of air and gas flows (v). It can be completed with a set of accessory primary converters for temperature measurement (t) of both gases and other environ- ments.
KPDM-1 2 DESCRIBTION AND OPERATION 2.1 Specification -400 ÷ +400 Range of measurement of pressure differential, mm w.g. 0.4 ÷ 70.0 Range of speed determination of air and gas flows, m/s 0 ÷ +300 Range of temperature measurement (according to certificate), º...
Page 6
2.3.1 The device functional principle is similar to the principle of the pneumometric tubes of Pito-Prandlia type. 2.3.2 KPDM-1 is a portable digital device with autonomous power supply. 2.3.3 PPC consists of a signal sensor (pneumometric tube of Pito-Prandlia type with a di-...
Page 7
KPDM-1 Primary converter for temperature meas- urement of gas and liquid environments Display and control unit Digital display Sensing element (sensor) of a prima- ry converter Button "I" Button "P" Pneumojumper Primary pneumoconverter Toggle switch of the device switch- ing-on...
KPDM-1 3 PRELIMINARY STARTING PROCEDURE 3.1 General provisions. 3.1.1 Liquid penetration into the PPC sensor holes is not allowed. 3.1.2 When installing (replacing) the sensor the measures should be applied to ensure the connection hermiticity of the sensor and the PPC body frame pneumochannels.
Page 9
KPDM-1 4.1.4 When pushing and holding "I" button the measurements are performed in series with an interval 1.5 seconds, when releasing the button the digital information corresponding to the av- erage value of air (gas) differential pressure for the time of holding the button appears on a display and will be kept there until the next pushing of "I"...
5 INSTRUCTIONS ON VERIFICATION 5.1 Verification of the devices is performed in accordance with the verification technique "MP 07-221-2009. Micromanometer with static and dynamic pressure receiver KPDM-1. Verifica- tion technique" 5.2. Calibration interval - 1 year.
KPDM-1 6 MAINTENANCE AND REPAIR 6.1 If during the device operating a sign of battery discharge in the lower left corner of a display is constantly indicated it is necessary: a) to change the battery, to do this loosen the screws on a back cover of the display and control unit and remove the device cover (or remove the cover of a separate bay for battery), re- move the old battery and insert a new one.
KPDM-1 8 TRANSPORTATION 8.1 Transportation of the devices can be performed by any means of transport in packaged form with mandatory protection from atmospheric precipitation and strong strokes. 9 MANUFACTURER'S GUARANTEES 9.1 The manufacturer guarantees conformity of a product to the requirements of the opera- tional documentation under keeping by the consumer operational conditions, storage and transpor- tation.
KPDM-1 12 ACCEPTANCE AND PACKING CERTIFICATE 12.1 Micromanometer with static and dynamic pressure receiver KPDM-1 serial number _______ produces and accepted in accordance with TC 4311-004-25057366-2008, the existing design documentation and recognized valid for use and storage. Date of issue _____________________________________________________...
KPDM-1 ATTACHEMENT 1 (obligatory) CERTIFICATE ON PERIODIC VERIFICATION Micromanometer with static and dynamic pressure receiver KPDM-1 serial number ________ is verified in accordance with the verification technique MP 07-221-2009. Verification results are recorded in a table. Date Signature Verification mark stamping or...