Detailed Message
INTERMITTENT
BATTERY
I
NTERNAL
E
RROR
146
CODE
Reasons
129
Battery connection opened
five times in one minute.
(
)
Display Buffer is full
NONE
Solutions
Check and clean the
battery contacts. Verify
that correct contacts are
used. Verify that the
battery is properly
inserted. Verify that the
C-code is correct for the
battery.
Reduce the charge rate in
extended C-code.
Increase end-of-
discharge voltage.
For intrinsically safe
batteries, use I/S settings
(see (Intrinsically Safe
Batteries, page 57).
Machine will reset in a
few minutes a resume
service. If message
persists, press the Esc
key for a few minutes, the
machine will reboot and
resume service.
This may occur if
QuickTest
(a processing
TM
intensive program) is
used. Start QuickTest
each station after 10
seconds.