Advertisement

Model 1000+ Printer User Manual
AMBCO 1000+P
AUDIOMETER
AMBCO ELECTRONICS
15052 REDHILL AVE SUITE #D • TUSTIN, CA 92780 • (714) 259-7930 • FAX (714) 259-1688
WWW.AMBCO.COM
10-1004, Rev. A
DCO 17–008, 11–13–17

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 1000+P and is the answer not in the manual?

Questions and answers

Summary of Contents for Ambco 1000+P

  • Page 1 Model 1000+ Printer User Manual AMBCO 1000+P AUDIOMETER AMBCO ELECTRONICS 15052 REDHILL AVE SUITE #D • TUSTIN, CA 92780 • (714) 259-7930 • FAX (714) 259-1688 WWW.AMBCO.COM 10-1004, Rev. A DCO 17–008, 11–13–17...
  • Page 2: Table Of Contents

    Table of Contents 1 About the 1000+P Audiometer 2 Features and Specifications of the Model 1000+P 3 Setting up the AMBCO Model 1000+P with the Printer 4 Getting Started – Patient Instructions 5 Threshold Testing – Establishing Hearing Level at Selected Frequencies...
  • Page 3: About The 1000+P Audiometer

    Model 1000+ Printer – User Manual ABOUT THE 1000+P AUDIOMETER The AMBCO Model 1000+P is a microprocessor controlled pure tone air conduction audiometer with automated screening and automated threshold test feature. FEATURES AND SPECIFICATIONS OF THE MODEL 1000+P Meets performance requirements of ANSI S3.6-1996 OPERATING TEMPERATURE: 15°...
  • Page 4: Getting Started - Patient Instructions

    Model 1000+ Printer – User Manual connected to the three jacks located on the rear jack panel. Insert the blue plug into the blue jack, the red plug into the red jack and the black plug into the black jack. 3.3.
  • Page 5: No Response

    Model 1000+ Printer – User Manual 5.10. This indicates the patient’s hearing at 500 Hz in the right ear is 20 dB. The patient responded to 20 dB, but not to 15 dB. 5.11. Record result per Section 7 – Recording Results. 5.12.
  • Page 6: Manual Test Print-Out

    Model 1000+ Printer – User Manual Results 500Hz 20dB 10dB 1000Hz 25dB 10dB 2000Hz 30dB 15dB 3000Hz 20dB 10dB 4000Hz 25dB 10dB 6000Hz 20dB 10dB 8000Hz 30dB 30dB MANUAL TEST PRINT-OUT 8.1. For printed manual test results, screening, threshold testing press and hold ‘EAR’ twice (1-2 seconds).
  • Page 7: 10 Instructing The Patient On How To Take The Test Correctly

    Model 1000+ Printer – User Manual 10. INSTRUCTING THE PATIENT ON HOW TO TAKE THE TEST CORRECTLY THIS IS THE SINGLE MOST IMPORTANT STEP !!! 10.1. Instruct the patient using the following phrases, “you are about to take a hearing test, it is not a contest so do not guess, respond only if you are sure you hear a tone (or set of tones if ‘PULSED’...
  • Page 8: 12 Automatic Threshold Test

    Model 1000+ Printer – User Manual 11.5. No response to the familiarization tone causes the alarm to sound, pressing any key stops the alarm. Then press ‘EAR’ for a print out of the error message, TEST FAILED AT FAMILIARIZATION TONE. 11.6.
  • Page 9 Model 1000+ Printer – User Manual 14.2. Exit AUTO-TEST after FAIL by pressing AUTO-TEST twice. 15. INTERPRETING RESULTS The hearing test results are interpreted by qualified persons such as physicians, audiologists, or nurses. Numerous factors affect hearing, such as age and background noise. This list below may be used as a general guideline to compare hearing test results.