GE Druck DPI 820 User Manual

Dual input thermometer

Advertisement

Quick Links

DPI 820
Dual Input Thermometer
User manual
K386 Issue 1
99 Washington Street
Melrose, MA 02176
Fax 781-665-0780
TestEquipmentDepot.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DPI 820 and is the answer not in the manual?

Questions and answers

Summary of Contents for GE Druck DPI 820

  • Page 1 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com DPI 820 Dual Input Thermometer User manual K386 Issue 1...
  • Page 2 Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 K386 Issue 1 FAX 781.665.0780 - TestEquipmentDepot.com...
  • Page 3: Table Of Contents

    Table of Contents Introduction The DPI 820 Dual Input Thermometer is part of the GE Druck Introduction ................1 DPI 800 series of hand held instruments. Safety ..................1 The DPI 800 series uses Intelligent Digital Output Sensor (IDOS) Safety - Marks and symbols on the instrument ....1 technology to give instant plug and play functionality with a To start ..................
  • Page 4: To Start

    To start To start - Items on the display To start - Location of items Item Description Item Description UMM - Pressure only. Task indication for the leak test. On or off button. Refer to: Select Task (Table 2) Left-hand soft-key. Selects the function above it on the display (Item 20).
  • Page 5: To Start - Set Up The Basic Operation

    To start - Set up the basic operation To start - Select a task (Measure) Use the Set Up menu to set up the basic operation of the When the instrument is set up (Table 1), use the Select Task instrument.
  • Page 6: To Start - Set Up The Settings

    To start - Set up the settings To start - Edit functions When the task is set up (Table 2), use the Settings menu to Example 1) Set up a label for x:y Scaling = % adjust the input operation Settings Edit Edit...
  • Page 7: Operation - Measure Temperature

    Operation - Measure temperature To measure temperature: 1. Connect the applicable thermocouples (Figure 1/2) and, if necessary, adjust the Set Up (Table 1). 2. Select a Temperature input task from Select Task (Table 2) and, if necessary, adjust the Settings (Table 3). If the display shows <<<<...
  • Page 8: Operation - Use The Snapshot Functions

    When the UMM is connected, make the necessary pressure Operation - Use the Snapshot functions selections. Use Snapshot to record up to 1000 displays then examine the IDOS - Measure the pressure results on the display, or transmit all the data to a PC or printer To measure the pressure: (Figure 5).
  • Page 9: Maintenance

    All the other configuration options stay in memory. specified data format. Maintenance – Recalibrate the unit Data format Note: GE Druck can provide a calibration service that is Baud rate: 19200 Data bits: 8 traceable to international standards. Parity: None...
  • Page 10: Specification Data

    Specification data All accuracy statements are for one year. Specification - General Languages English [Default] Operating temperature 14 … 122°F (-10 … 50°C) Storage temperature -4 … 140°F (-20 … 60°C) Humidity 0 to 90% without condensation (Def Stan 66-31, 8.6 cat III) Shock/Vibration BS EN 61010:2001;...