Summary of Contents for Kane-May KM20REF

  • Page 1 KM20REF REFERENCE THERMOMETER...
  • Page 2: General Information

    GENERAL INFORMATION It is recommended that you read the safety and operation instructions before using this instrument. WARNING TO AVOID DAMAGE OR BURNS, DO NOT MAKE TEMPERATURE MEASUREMENTS IN MICROWAVE OVENS. CAUTION Temperature Measurement Probes This precision instrument has been designed for use with the extensive range of Comark temperature probes.
  • Page 3: Calibration, Certification And Service

    CALIBRATION, CERTIFICATION AND SERVICE 0451 Certifi cation Comark can provide certifi cates of calibration for its whole product range, to suit ISO 9000 and other quality assurance procedures, food hygiene regulations, HACCPs and environmental regulations. Comark certifi cates are produced by independent quality controlled processes which compare product performance against agreed National Standards.
  • Page 4 This certifi cate defi nes the accuracy of the KM20REF at designated points. To maximise the stability of readings from the KM20REF or indeed any other reference thermometer it is recommended that:- a) it is only used as a reference thermometer it is only used in ambient temperatures of +20°C ±5°C...
  • Page 5 If the reading from the food thermometer differs from the traceable reading of the KM20REF by ±0.5°C or more, the food thermometer and probe should be returned to the manufacturer for recalibration.
  • Page 6 EXAMPLE CALIBRATION CHECK RESULTS SHEET DATE: 2nd May 2001 TIME: 0930 AMBIENT TEMPERATURE: 21.3°C KM20REF SERIAL NUMBER: 1234 FOOD THERMOMETER SERIAL NUMBERS: INSTRUMENT: 0675 PROBE: 0231 KM20REF READING: +8.1°C ADJUSTMENT –0.1°C TRACEABLE TEMPERATURE: +8.0°C FOOD THERMOMETER READING: +8.5°C DEVIATION +0.5°C...
  • Page 7 TESTED BY: CHECKED BY: Auto Switch Off The KM20REF has an auto switch off function to conserve battery life. The instrument will automatically switch off after approximately 3 minutes use. Press the ON/OFF button again to continue using the instrument.
  • Page 8: Specification

    SPECIFICATION SCALE °C MEASUREMENT RANGE Pt100 –100°C to +199.9°C RESOLUTION 0.1°C ACCURACY ±0.2°C OPERATING RANGE 0 to +40°C CERTIFICATE OF CALIBRATION DESIGNATED POINTS FOR ACCURACY –18°C, 0°C, +5°C, +8°C, +20°C Tested to EN 61326-1 Criteria B performance...
  • Page 10 Comark Instruments 52 Hurricane Way, Norwich, Norfolk, NR6 6JB United Kingdom Tel: +44 844 815 6599 Fax: +44 844 815 6598 Email: sales@comarkinstruments.com Website: www.comarkinstruments.com Comark Instruments PO Box 500, Beaverton, OR 97077 USA Tel: 503 643 5204 Fax: 503 644 5859 Email: sales@comarkusacom Website: www.comarkusa.com ©...