Valves are warranted for a period of 180 days. Preston Pressure’s limit of liability shall be to repair or replace the product at their discretion. Transportation shall be pre-paid to the Preston Pressure factory or authorized service facility.
GENERAL INFORMATION Description of the PS-425 Pitot-Static Test Box The PS-425 Air Data, pitot-static test box is a digital display tester used to test aircraft altimeters, airspeed indicators, pitot-static systems, and other vacuum and low-pressure devices. This test box incorporates transducer sensor technology that is largely immune to the failures and costly repairs associated with analog instruments.
Leak test feature: 2 minutes, selectable down to 30 seconds (longer test periods available upon request) VSI (vertical speed indicator) : Increments of 2 when < 100 feet/minute; increments of 25 when > 100 feet/minute. Limited to 9,900 ft/minute Note: The PS-425 does NOT have the precision required to test RVSM altimeter systems.
Front Panel Description The UNITS button is used to toggle between the two primary display modes. Upon powering up, the display indicates altitude in feet and airspeed in knots. Pressing the UNITS button will switch the display to indicate altitude in feet and vertical speed per minute in feet.
CHAPTER 2 THEORY The PS-425 test box consists of one manual vacuum pump, one manual pressure pump, reservoirs for the pressure and vacuum, and a digital display for indicating altitude and airspeed. • Pneumatic Theory Metering valves allow the operator to control the rate of flow of the air from the pressure or vacuum sources, exiting out through the vacuum and pressure ports.
OPERATION • Preliminary Information Prior to using the PS-425 tester, the user should become familiar with the unit by reading this manual and operating the tester independently of any other attached devices. For greater accuracy, the indicator should be powered on for 15 minutes before use.
7. The leak check function will activate and begin to monitor the leak rate. Verify that the PS-425 does not leak exceeding 100 feet per minute. Write down the leak rate for future reference. (Subtract this value from the total leak rate when testing the aircraft, to arrive at the aircraft’s leak rate.)
Note: If the previous pitot system test indicated an excessive leak, do not proceed until that leak is repaired. Damage to aircraft airspeed indicator could result. 1. Turn on the power to the PS-425. 2. Connect the Vacuum Port to the aircraft static port, and the Pressure Port to the aircraft pitot port.
Page 12
3. Close Vacuum Control, Vacuum Vent, Pressure Control, and Pressure Vent valves. Open Crossbleed valve. 4. Actuate the Vacuum Pump until a desired level of vacuum is seen on the Vacuum Reservoir gauge (generally -15 to -20 inHG). Actuate the Pressure Pump until the Pressure Reservoir indicates 15 PSI or more.
The PS-425 is easy to calibrate. There are 5 set points used in the calibration of the altimeter. This enables the altimeter to have a greater degree of accuracy, allowing it to self-correct to the nearest calibrated set point.
Page 14
“Storing” and then re-start. There is no need to re-cycle power. 9. After bringing the PS-425 back down to ground level pressure, wait at least 1 minute for settling. Then, test each calibration point as specified in the chart on page 14. Enter the corrected value at each point.
Airspeed Calibration Procedures 1. Perform the altimeter calibration described in the previous section. The pitot sensor is calibrated in this process, along with the static sensor. 2. When a satisfactory altitude calibration is obtained, go through each test point in the following chart, and enter the corrected value at each point.
1. Check color of beads in air reservoir/dryer. Change if pink. The dryer is a common in- line dryer used in small air compressor systems. A new dryer can be purchased from Preston Pressure, or an internet search for “mini inline desiccant air dryer” may lead to an alternate source.
Page 18
excessively cold. 2. Low battery Unit will not vacuum up to 1. Leak in internal hoses or 1. Troubleshoot plumbing 35,000 feet plumbing connectors to find leak(s) 2. Cylinder pump is weak or Replace defective inoperative cylinder...
CHAPTER 6 SHIPPING AND STORAGE Shipping Instructions • Partially open all valves • Ensure power switch is off • Package the test box inside another box, with at least 2 inches of packing material around the tester • Tester is not likely to be damaged with ground shipping if packaged properly Storage Instructions •...