8.4
8.4
8.4
8.4

Error codes

Error codes
Error codes
Error codes
ERROR ZAXIS
ERROR YAXIS
ERROR XAXIS
ERROR HEATER
Legend:
8.5
8.5

Solutions

Solutions
8.5
8.5
Solutions
Solutions
In the event that the machine
malfunctions during the welding
process, the cycle will not be
completed.
Use the keypad to open the cover
and remove the blade and tubing.
Turn off the unit and wait
10 seconds before switching on
again. This should reset the unit.
If the program continues to
malfunction please contact Service
Representative.
16
Time limitation elevating motor (Z-axis)
Z-axis did not reach the programmed position in between 8 sec.
Time limitation slide motor (y-axis)
Y -axis did not reach the programmed position in between 8 sec.
Time limitation eccentric motor (X-axis)
X -axis did not reach the programmed position in between 5 sec.
Time limitation heater
Blade is not able to reach the depyrogenation temperature in between
the programmed 90 sec.
Z axis: corresponds to the stroke
X axis: corresponds to the red holder
Y axis: corresponds to the blue holder