The Model 100 series are simple, accurate cost effective syringe pumps designed to hold glass or plastic syringes, of any make, from 10 microliter to 60 milliliter for the KDS100; 10 ul to 15 ml for the KDS120; and 10 ul to 10 ml for the KDS101.
select This key has three functions: Return to main menu when pressed repeatedly Selects the pulsing or "highlighted" feature of the main menu display. Enter numerical values. run/stop Turns the motor on and off. Acts as a "pause" during a dispense. DISPLAY After the pump is programmed an arrow on the right side of the display indicates the direction of operation.
SYRINGE DIAMETER ENTRY For automatic internal calibration the syringe diameter must be entered. Once entered the diameter is retained in non-volatile memory and need not be entered again unless the syringe used is changed. When a syringe diameter is changed the Volume and Rate settings are set to zero.
RATE can be selected from the main menu or will be displayed automatically after the Volume setting. The display reads: Rate: X.X ul/h or X.X ml/h Rate units are ul/h or ml/h, microliter or milliliter/hour and are selected automatically from the syringe diameter. Use the arrow keys to display the flowrate required.
changed while the pump continues to run. When the changes are entered the pump immediately changes to the new settings. While operating press select to return to the main menu. Select Volume to display the set dispense volume. If no change is required press select For a volume change use the arrow keys and enter with the select key.
LOADING THE SYRINGE KDS100 and KDS101 To facilitate loading , the pusher block (1) can be released from the leadscrew and manually moved along the guide rods to accommodate the syringe. Press the bronze button (2) on the side of the black pusher block (1) to release the block from the leadscrew.
KDS120 Push-Pull pump To facilitate loading , the pusher block can be released from the leadscrew by pressing the bronze button (1) and can be manually moved along the guide rods to accommodate the syringes. First place the withdraw syringe in the withdraw syringe holder, using the spring loaded retaining clamp to retain the syringe barrel in the V of the syringe holder.
NV RAM FAILURE If the settings in the non-volatile memory become corrupted the display will read "NV Ram Failure" and the pump will not operate. To recover from this condition the pump must be powered down and then turned on again. The pump will be re-initialized to the default settings and can now be programmed as normal.
LIMITED WARRANTY KD Scientific Inc. warrants to the first consumer purchaser, for a period of one year from the date of purchase that this unit, when shipped in its original container, will be free from defective workmanship and materials and agree that it will, at its option, either repair or replace the defective unit.