Checking Of Parameters Relevant To The Device - Siemens SIWAREX WP231 Manual

Hide thumbs Also See for SIWAREX WP231:
Table of Contents

Advertisement

Operation requiring verification
12.2 Verification
12.2.2

Checking of parameters relevant to the device

The following parameters relevant to the device are checked during the verification:
1. Checking of the software ID of the SecureDisplay
The software ID of the DisplayCali.exe function must agree with the requirements in the
SIWAREX SecureDisplay test certificate (EC). The currently valid version can be called
on the SecureDisplay.
– Call up the main menu in the main display (function key with the open-end wrench).
– Select the "Version and variables check" submenu in the "Calibration check"
– Activate the "Display SecureDisplay version" command.
– The version of the SecureDisplay is then shown on the display.
2. Checking of the firmware ID of the SIWAREX WP231 evaluation electronics
The firmware ID of the SIWAREX WP231 evaluation electronics must agree with the
requirements in the EC construction license. The currently valid version can be called on
the SecureDisplay.
– Call up the main menu in the main display (function key with the open-end wrench).
– Select the "Version and variables check" submenu in the "Calibration check"
– Activate the "Display SIWAREX firmware release" command.
– The current SIWAREX firmware release is then shown on the display.
3. Checking of the smallest zoom factor for the SecureDisplay display software
The zoom factor entered in DR3 for the alternative display size must comply with the
minimum readability and font requirements of EN 45501, Chapter 4.2.1.
The main display with the minimum zoom size can be checked for readability.
– Call up the main menu in the main display (function key with the open-end wrench).
– Select the "Version and variables check" submenu in the "Calibration check"
– Activate the "Display smallest SecureDisplay" command.
– The smallest main display is then output and can be checked for readability.
160
Select the "Calibration check" submenu in the main menu.
submenu.
Select the "Calibration check" submenu in the main menu.
submenu.
Select the "Calibration check" submenu in the main menu.
submenu.
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02

Advertisement

Table of Contents
loading

Table of Contents