Download Print this page

Centromed Mattress Variator Instruction Manual

Frame, control unit and a hose

Advertisement

Quick Links

Page 2
Page 3
Page 4
Page 10
Page 11
Mattress Variator
Instruction Manual
Page 1 The Mattress Variator Manual
Contents
Fitting Instructions
Using your Ultimate Pillow Lift
Repair Instructions
Technical Information
Contact Information

Advertisement

loading

Summary of Contents for Centromed Mattress Variator

  • Page 1 Contents Page 2 Fitting Instructions Page 3 Using your Ultimate Pillow Lift Page 4 Repair Instructions Page 10 Technical Information Page 11 Contact Information Mattress Variator Instruction Manual Page 1 The Mattress Variator Manual...
  • Page 2 The frames can be fitted to almost any bed by means of straps and buckles supplied with the frame. See diagrams for details. To fit the Mattress Variator to the bed it will be necessary to remove the mattress itself. Fit the main frame to the base of the bed at the head end. The frame should be placed with its base against the base of the bed (the base of the frame has the air connections on it).
  • Page 3 The Mattress Variator is a fully portable device for people who experience problems getting in and out of bed or getting to a sitting position. The Mattress Variator is easily fitted to beds, and can support a patient in variable positions for added comfort...
  • Page 4: Safety Warning

    REPAIRS Note: If this equipment was supplied by Centromed, you may contact us on the details given towards the back of this manual. If the equipment was supplied to you by another company, please contact this supplying company with regards to repairs.
  • Page 5 N. To check that the wiring is in place it will be necessary to remove the lid of the equipment. CAUTION this equipment is mains powered and should be isolated from the mains before it is opened. To remove the Page 5 The Mattress Variator Manual...
  • Page 6 P. If there is a fault with the wiring it should be repaired to IEC60601-1 if you are unsure of the correct way to do this return the unit to Centromed for repair. Refer to section N for description of wiring circuit.
  • Page 7 Page 7 The Mattress Variator Manual...
  • Page 8 FRAME REPAIR FLOWCHART EXPLANATIONS A. Test the frame with a spring gauge available from Centromed. If the frame inflates / deflates OK and can hold 20 lb. on the spring gauge for 10 minutes it should work properly in the field. Reasonable inflation time is 10 seconds, and deflation time of 20 to 30 seconds.
  • Page 9 Valve plate lower deflate valve frame Replace plate seal Does it leak, or inflate & Check deflate Water test seal slowly? bung aligned Bung Replace bung Water test frame Replace Frame Replace bellows frame Page 9 The Mattress Variator Manual...
  • Page 10: Technical Information

    Mattress Variator and / or the other equipment, also try to connect them on different mains circuits. RECYCLING There are no toxic components inside your Mattress Variator, so it can be disposed of quite safely in the normal manner. Alternatively, it can be returned to Centromed for recycling. The life expectancy of this equipment is about five years.
  • Page 11: Contact Information

    Have your order delivered or have a factory trial* Make use of our educational posters, videos & presentations Hire products at HIRE HIRE affordable prices Maintenance, repair & training courses available * Only available on certain products Page 11 The Mattress Variator Manual...