The electrical connection is provided through a screened 8-pin FCC 68 connector. Unpacking and Checking Unpack the PEG100 immediately after delivery, even if it is to be installed at a later date. Examine the packaging for any external damage. Completely remove all packaging materials.
Design and Function Based on a supply voltage of 24 V the PEG100 generates the internal supply vol- tages required for operation of the integrated Penning measurement system. The PEG100 supplies a logarithmic representation of the vacuum pressure by way of a voltage signal which ranges from 0.66 V to 10 V.
6 Troubleshooting 7 Wearing Parts and Original Spare Parts 9 Returning the Product Disposal Appendix Output Voltage vs. Pressure Examples for Connecting EU Declaration of Conformity For cross-references within this document, the symbol (→ XY) is used. tina14e1-a (2017-12) PEG100.om...
Communicate the safety instructions to all other users. 1.4 Liability and Warranty INFICON assumes no liability and the warranty is rendered null and void if the end- user or third parties • disregard the information in this document •...
U <7 V (for this refer also to figure 3 and section 3.2.2) Status Output Ready to measure High level (13.5 to 35 V, max. 50 mA) Error (not ignited, HV off) For this refer also to section 3.2.4. tina14e1-a (2017-12) PEG100.om...
Page 7
10 °C to 50 °C Max. rel. humidity of the ambient air (on 30 days per year, non-condensing) 70 % or 95 % usable measurement range 10 to 10 mbar usable measurement range 10 to 10 mbar tina14e1-a (2017-12) PEG100.om...
Flange up installation is not permissible because under such circumstances con- densate may collect in the PEG100. This will either adversely affect the measure- ments, or the sensor itself may possibly be damaged.
Page 9
Two examples of how to connect the PEG100 are given in Annex B. Fig. 3 Block diagram tina14e1-a (2017-12) PEG100.om...
5×10 mbar. 3.2.3 Measurement Signal The PEG100 supplies a defined output signal ranging from 0.66 V to 10 V at Pin 3 Output with reference to Pin 5 which is signal ground. For this also refer to Table 1 in Annex A.
For the purpose of identifying the connected type of gauge and the pressure range, the PEG100 is equipped with an identification resistor (R = 100 kΩ) between Pin 4 and Pin 2. This resistor may be sensed by connected operating or control units so that these can automatically adapt.
PE G100. You mu ust always sw witch off the su pply voltage the PEG100 b before starting g any work on the gauge (ev ven when moving the gau uge from the c...
70 °C. Before baking out, you must make sure that an ultra sealing disc (PEG100, DN 25 KF) or a copper seal (PEG100, DN 40 CF) is used as the flange seal. The PEG100 is equipped with all-metal sensing cells so that any outgassing caused by polymer seals is entirely avoided.
5.2.2 Disa assembly o of the The sen nsor consists o of the vacuum housing, the anode ring (7 7/4) with ignitio on aid Sens (7/2) an d the cathode e plate (7/5). S See Fig. 7. How to d disassemble: 1) Use a pair of twee...
When the electronics assembly has been placed as required, the black magnet housing is fully surrounded by the housing of the PEG100. 3) Retighten the cross head screws at the rear of the gauge.
9 Returning the Product WARNING WARNING: forwarding contaminated products Products returned to INFICON for service or repair should, if possible, be free of harmful substances (e.g. radioactive, toxic, caustic or micro- biological). Otherwise, the type of contamination must be declared.
Examples for Connecting Example for connecting the PEG100: Switching the high tension via an external switch or contact Example for connecting the PEG100: Switching the high tension automatically through a Pirani Standard Gauge with switching point tina14e1-a (2017-12) PEG100.om...
EU Declaration of Conformity We, INFICON, hereby declare, that the equipment mentioned below complies with the following directives: • 2014/30/EU, OJ L 96/79, 29.3.2014 (EMC directive; Directive on electromagnetic compatibility) • 2011/65/EU, OJ L 174/88, 1.7.2011 (RoHS directive; Directive on the restriction of the use of certain hazardous substances in electrical...