Users should avoid looking directly into the optic output of OVA-50. And the use of microscope or magnifier should also be avoided, for the use of such devices can focus a highly intense beam onto the retina, which may result in permanent eye damage Battery: OVA-50 battery type is lithium battery.
Page 3
OVA-50 User’s Manual when the instrument is working Always replace protective dust cap on the detector port when the instrument is not in use. Always avoid looking directly at unconnected end of optic fiber in testing and make the unconnected end pointing at a non-reflective object.
If necessary, you may contact Shineway Technologies, Inc via this email: support@shinewaytech.com. 1.3 Introduction ShinewayTech OVA-50 Optical Variable Attenuator can precisely attenuate input optical signals at 1310/1490/1550/1625nm wavelengths and directly output defined stabilized Shineway Technologies, Inc. - 1 -...
Page 6
Working like a combination of attenuator and power meter, OVA-50 Output Power Control mode enables direct setting of precise optical power level and can automatically stabilize output level without any interference by variations of input power.
2. Basic Operation 2.1 Foreword This part introduces the basic operation on OVA-50. Specific operations of each type instrument are elaborated in Chapter 3 of this manual. Please read this manual carefully for optimal operation. If you encounter any problems during operation, you are welcome to contact the technical staff of our company or representatives.
2.2.1 External Interfaces Optical Input OVA-50 is available with FC connector (Interchangeable SC/ST). USB Power /Data Port OVA-50 can be charged by external power adapter (5V DC, 750mA). 2.2.2 Keypad Operation Power on/off Toggle between calibrated wavelengths Reset to minimum attenuation value Short press to toggle between working modes: Absolute attenuation, Relative attenuation, Program &...
OVA-50 works on lithium battery, please make sure battery is mounted properly before use. When battery is low, low battery indicator will appear on LCD. You can still use OVA-50 as long as its display on LCD is identifiable. Please charge as soon as possible when battery is low to ensure accurate measurement.
1.00 Figure 3.2 3.2 Absolute Attenuation It is default to enter absolute attenuation mode when OVA-50 powers on; or press button to absolute attenuation mode, see Figure 3.2; In absolute attenuation mode, press button to toggle between the steps of 0.05/0.10/1.00/10.00;...
Note: 1. Function shown in Figure 3.3 is mainly applied before testing to set the insertion loss for jumper fiber and the instrument to ensure accurate attenuation calculation. 2. Once the insertion loss value is set, press button and OVA-50 will save the value as default minimum attenuation value. 3.3 Relative Attenuation Press button to relative attenuation value mode, see Figure 3.4;...
OVA-50 User’s Manual button to adjust relative attenuation value; Press button to confirm and return to the initial interface of relative attenuation mode. : 0.05 1550 nm 10.30 SET ATTRef Figure 3.5 3.4 Program Press button to program attenuation mode, see Figure 3.6 :...
OVA-50 User’s Manual 3.5 Output Power Control Press button to output power control mode, see Figure 3.7. 1550 nm - 15.00 OUTPUT PWR Figure 3.7 Press button to enter output power setting interface, see Figure 3.8. 1550 nm - 15.00 SET OUTPUT Figure 3.8...
OVA-50 User’s Manual 4. Calibration 4.1 Optical Interface Cleaning Optical interface must be always kept clean. Always put protective dust cap on when the unit is not in use, and keep the protective dust cap clean. 4.2 Calibration Requirements Calibration of the instrument is recommended every 3 years. Please contact our representatives or customer service centers for proper calibration.
OVA-50 User’s Manual 5. Specifications Model OVA-50 Wavelength Range 1260~1650nm Calibrated Wavelengths 1310/1490/1550/1625nm 9/125μm Singlemode Fiber Type Attenuation Range 2.5 - 60dB Insertion Loss <2.5dB Max Input Power +21dBm Output Power Range +18 ~ -55 dBm Setting Time <3s Display Resolution 0.01dB...
OVA-50 User’s Manual 6 Warranty Information 6.1 Warranty Period All ShinewayTech products are warranted against defective material and workmanship for a period of one (1) year from the date of shipment to the original customer. Any product found to be defective within the warranty period would be repaired or replaced by Shineway Technologies Inc free of charge.
OVA-50 User’s Manual Be sure to pack the instrument with soft cushion like Polyethylene, so as to protect the shell of the instrument. Please use the original hard packing box. If you use other packing material, please ensure at least 3 cm soft material around the instrument.