Mercury Advance
Trolley

User Manual

D I R E C T H E A L T H C A R E G R O U P . C O M

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Dyna-Form Mercury Advance and is the answer not in the manual?

Questions and answers

Summary of Contents for Direct Healthcare Group Dyna-Form Mercury Advance

  • Page 1: User Manual

    Mercury Advance Trolley User Manual D I R E C T H E A L T H C A R E G R O U P . C O M...
  • Page 2 D Y N A - F O R M  M E R C U R Y A D V A N C E T R O L L E Y D I R E C T H E A L T H C A R E G R O U P . C O M...
  • Page 3: Table Of Contents

    U S E R M A N U A L Mercury Advance Trolley The Dyna-Form Mercury Advance Trolley Mattress has been designed especially for users considered to be that of Very High Risk of pressure ulcer development in potentially lengthy A&E waiting environments. This product can be made to measure in order to fit most trolley designs and is ideal for use in busy A&E departments that have an extensive patient turnover.
  • Page 4: Introduction

    Mattress in an emergency. (Frequently used functions) 2. Quick Reference Guide This is a quick reference guide for the Dyna-Form Mercury Advance Trolley System Product Code MAT/ADV/TROL/SIZE Power Switch Audible Warning Reset The power switch simply switches the mains power to the pump on and off.
  • Page 5 U S E R M A N U A L LED Mode Settings This symbol when illuminated (The green indicator light) is not used to indicate that the equipment is on or ready for use. When a patient requires a true dynamic function or indeed more pressure in the cells, as they may be uncomfortable or feel as though the support surface is too soft or unstable, then please select a “High”...
  • Page 6: Troubleshooting

    D Y N A - F O R M  M E R C U R Y A D V A N C E T R O L L E Y 3. Troubleshooting Warning / Fault Cause Solution The Control 1.
  • Page 7: Installation

    U S E R M A N U A L 4. Installation 4.1 Mattress (This is the applied part type BF) Place the Dyna-Form® Mercury Advance Trolley Mattress directly on to the bed platform ensuring that the Blue multi-stretch waterproof cover is on top and that the umbilical hose is located at the left hand corner at the foot end of the bed.
  • Page 8: Transportation

    5.3 Troubleshooting For assistance (if needed) in setting up, using or maintaining the Mercury Advance System, or to report unexpected operation or events, please contact Direct Healthcare Group on the contact details on the reverse of this manual. 6. Transportation To change the location of the mattress, remove the Umbilical Cord and allow the mattress to return to its Static Mattress form.
  • Page 9: Maintenance Procedures

    Unqualified solution or equivalent disinfectant. personnel attempting to work on Direct Healthcare Group Power 4. Wash Mattress top using hot water (60 degrees C) containing Units risk serious injury to themselves and others and possibly detergent –...
  • Page 10: Technical Specification

    D Y N A - F O R M  M E R C U R Y A D V A N C E T R O L L E Y 9. Technical Specification Declaration – electromagnetic emissions - for all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer’s declaration –...
  • Page 11 U S E R M A N U A L 9. Technical Specification Declaration – electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity The MAT1260001 is intended for use in the electromagnetic environment specified below. The customer or the user of the system should ensure that it is used in such an environment. Immunity test IEC 60601 Compliance level...
  • Page 12 D Y N A - F O R M  M E R C U R Y A D V A N C E T R O L L E Y 9. Technical Specification Declaration – electromagnetic immunity – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration –...
  • Page 13 U S E R M A N U A L 9. Technical Specification Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for ME EQUIPMENT or ME SYSTEM that are not LIFE – SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the MAT1260001 Alternating Control Unit The MAT1260001 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
  • Page 14: Technical Data

    D Y N A - F O R M  M E R C U R Y A D V A N C E T R O L L E Y 10. Technical Data 12. Symbols Guide 9.1 Power Unit (Pump) Mattress Symbols Serial Number ......As per label on rear of pump WASH AT 80˚...
  • Page 15: Detachable/Removable Parts

    Use of detachable parts not listed is not recommended by used in the manufacture of the Mercury Advance System Direct Healthcare Group. comply with the required fire safety regulations. • Do not use sharp objects on or near the mattress system as this will cause damage.
  • Page 16 CF83 1BF T: +44 (0) 845 459 9831 F: +44 (0) 845 459 9832 E: info@directhealthcaregroup.com Asia Pacific Direct Healthcare Group PTY Ltd. PO Box 562, Wembley Western Australia 6913 T: +61 (0) 423 852 810 E: info@directhealthcaregroup.com.au LIT-00008 Issue 4 Date: February 2018 D I R E C T H E A L T H C A R E G R O U P .