To reinstall the software for the Villa Sistemi Medicali ROTOGRAPH-D, follow these steps:
1. If a memory corrupted error occurs, try restoring the operating condition by: - Pressing in sequence: small size push button (37), child (34), and TMJ1 (29), as per SET-UP mode. - Wait 5 seconds.
2. If condition is not restored: - Replace the CPU (A2) PCB. - Perform a general system check.
3. After replacing the CPU (A2): - Re-enter all factory or custom parameters lost during the failure. - Use the data recorded on the "Final Test Table" found in the User’s Manual. - Follow the Set-up procedure in paragraph 7.2 and following.
This process effectively reinstalls the software and restores system settings.
This answer is automatically generated
Subscribe to Our Youtube Channel
Related Manuals for Villa Sistemi Medicali ROTOGRAPH-D
Summary of Contents for Villa Sistemi Medicali ROTOGRAPH-D
Page 1
ROTOGRAPH ROTOGRAPH ROTOGRAPH ROTOGRAPH- - - - D D D D (120V version) (120V version) (120V version) (120V version) Service Manual Service Manual Service Manual Service Manual Release 11 November 2013 (Rev. 5)
Page 2
9-4 to 9-8, software correction. from 9-13 to 9-25, Digital Sensor identification label update. 10-2, 10-6, 10-7 Schematics and drawings update. Spare Parts update. (Ref. RDM 6746, RDM 7353, RDM 7575, RDM 7688, RDM 7809, RDM 7857) (Rev. 5) ROTOGRAPH-D (120V)
Page 3
SERVICE MANUAL Revision history THIS PAGE IS INTENTIONALLY LEFT BLANK ROTOGRAPH-D (120V) (Rev. 5)
SERVICE MANUAL Contents Contents INTRODUCTION 1.1 Icons appearing in the manual ............1-1 1.2 How to contact VILLA SISTEMI MEDICALI technical service....1-2 SAFETY INFORMATION 2.1 Warnings ..................... 2-2 2.2 Environmental risk and disposal ............2-3 2.3 Symbols used ..................2-4 DESCRIPTION 3.1 Identification labels ................
Page 5
7.10 Checking (safety) hardware timers............. 7-43 7.11 System reconfiguration after testings ..........7-44 MAINTENANCE SCHEMATICS AND DRAWINGS 9.1 Key to general diagram of the ROTOGRAPH-D........9-2 10. SPARE PARTS 10-1 This publication can only be reproduced, transmitted, transcribed, or translated into any human or computer language with the written consent of Villa Sistemi Medicali S.p.a.
SERVICE MANUAL Introduction How to contact VILLA SISTEMI MEDICALI How to contact VILLA SISTEMI MEDICALI How to contact VILLA SISTEMI MEDICALI How to contact VILLA SISTEMI MEDICALI technical service technical service technical service technical service For any technical queries please contact the following: •...
X-ray generators. Villa Sistemi Medicali shall not be responsible for: • any use of the ROTOGRAPH-D different from that for which it has been designed, • any damage to the equipment, the operator or the patient caused either by incorrect installation and maintenance not compliant with the procedures contained in the relevant user’s and installation...
3 meters away from the ROTOGRAPH-D and be powered via an independent electric line. ROTOGRAPH-D must be switched off during the entire period of use of ESU (Electro Surgery Units) units or similar equipment.
Holding the slider firmly, remove the pin and move the slider towards the top to the metallic stop of the column. At this point, remove all the remaining parts. (Rev. 4) ROTOGRAPH-D (120V)
Connection to neutral conductor Connection to line conductor Protection ground Functional ground OFF ; equipment not connected to power line ON ; equipment connected to power line Warning: read the documentation provided with the unit ROTOGRAPH-D (120V) (Rev. 4)
Page 13
SERVICE MANUAL Description ROTOGRAPH-D WARNING label label ETL certification Digital Sensor label identification label Remote control Tube head identification label identification label Laser WARNING label ROTOGRAPH-D (120V) (Rev. 5)
Page 14
(please refer to Chapter 6 of User’s Manual). ROTOGRAPH-D is controlled by a soft touch console and equipped with an alphanumeric digital display for a clear indication of the working parameters and operative messages.
Page 15
SERVICE MANUAL Description THIS PAGE IS INTENTIONALLY LEFT BLANK ROTOGRAPH-D (120V) (Rev. 5)
TECHNICAL FEATURES TECHNICAL FEATURES TECHNICAL FEATURES TECHNICAL FEATURES General characteristics Equipment ROTOGRAPH-D Manufacturer VILLA SISTEMI MEDICALI Buccinasco (MI) Italy Class Class II according to 21 CFR sub-chapter J. Class I with type B applied parts according to IEC. Degree of protection IP20 120V ±...
Page 17
(*) WARNING: The declared image magnification value is valid after proper software calibration. Tubehead features Type MR05 Manufacturer VILLA SISTEMI MEDICALI Buccinasco (MI) Italy Max peak tube potential 85 kV Nominal power 0.630 kW (85kVp, 10mA) Total filtration ≥ 2.8 mm Al eq. at 85 kV...
Page 18
30% ÷ 75% Operating relative humidity range -4°F ÷ +158°F (-20° ÷ +70°C) Transportation and storage temperature range Maximum transportation and storage < 90% non condensing relative humidity Minimum atmospheric pressure for 630 hPa transportation and storage (Rev. 5) ROTOGRAPH-D (120V)
SERVICE MANUAL Technical features Standards and regulation Standards and regulation Standards and regulation Standards and regulation The ROTOGRAPH-D equipment is manufactured according to the following standards: 21 CFR subchapter J • General safety: IEC 60601-1 IEC 60601-2-7 IEC 60601-2-28 IEC 60601-2-32 UL 2601 •...
VDC 20V, 1V=1mA). The exposure times are determined by using a timer/counter, having an accuracy of 0.1%, measuring the duration of part of the voltage applied to the primary side of the tubehead, during the exposure phase. (Rev. 5) ROTOGRAPH-D (120V)
The room pre-installation requirements for a successful installation are: • a minimum room ceiling height of 8 ft (2,5 mt) and an area to accept ROTOGRAPH-D with its different configuration as showed in Figure 5-1. • to transport the system into the room the doors must have a minimum width of 36"...
AWG 16 (1,5mm²). Grounding of this equipment must meet any standards in force. NOTE: The ROTOGRAPH-D IS PROVIDED with connections for remote control and signalling device to be placed at the entrance of the X-ray room with the following meanings: •...
Page 26
Federal Regulations. It is responsibility of the installers to observe and abide by the rules. NOTE: The electrical connection has to be made on the X15 terminal (refer to Figure 6-8 and the general wiring diagram at chapter 9). (Rev. 4) ROTOGRAPH-D (120V)
200kg - 441 lbs) per bolt applied. Backing and proper support is the full responsibility of the owner. Unpacking Unpacking Unpacking Unpacking ROTOGRAPH-D is shipped packed in two cartons. Dimensions, net weight, gross weight and the packing contents are listed in the following table: Weight Packing dimension Contents...
SERVICE MANUAL Pre-installation Space requirements Space re quirements Space re Space re quirements quirements 1170 (46") Figure 5-1 (Rev. 4) ROTOGRAPH-D (120V)
Page 29
SERVICE MANUAL Pre-installation THIS PAGE IS INTENTIONALLY LEFT BLANK ROTOGRAPH-D (120V) (Rev. 4)
INSTALLATION INSTALLATION INSTALLATION NOTE: ROTOGRAPH-D is dispatched preassembled in groups. The mechanical assembly consists exclusively in assembling these groups together. Calibration of the ball bearing, brakes and eventual tightening torques is factory preset before delivery; any intervention on these parts, besides not being necessary, can cause malfunction.
3. Insert the counterweight stop pin (code 5108102900) into the appropriate hole located at about mid-stand (Figure 6-2). 4. Position the counterweight (68kg - 150 lbs) with the flat surface upwards, and the rear pocket facing backwards. ROTOGRAPH-D (120V) (Rev. 4)
Page 32
(Figure 6-3). Fully insert the motor group on the guide. WARNING: During insertion of the motor group on the guide, pay attention not to place your fingers between track and carriage: they may be seriously pinched. (Rev. 4) ROTOGRAPH-D (120V)
Page 34
6-5). Fix the tubehead with the four screws so that the reference plate is fully in contact with the tubehead fixing plate (Figure 6-6). Wrong mounting will affect on the X-ray beam centering. 12. Connect the wires to the terminal strip making reference to Figure 6-5. (Rev. 4) ROTOGRAPH-D (120V)
Page 35
13. Remove the packing material that is blocking the arm rotation and the counterweight stop pin. 14. Install the two brackets "A" (code 5608100300) and the cable support bracket "B" (supplied inside the relevat cover) (Figure 6-7). ROTOGRAPH-D (120V) (Rev. 4)
Page 36
15. Fix the cables routing it as shown in Figure 6-7. The upper connectors have to enter inside tha chin support through the relevant hole. Fix tha cable to the bracket "B" using the plastic plate "C" (code 5608500600). Figure 6-7 (Rev. 4) ROTOGRAPH-D (120V)
Page 37
X15: Connector for signalling lamp and command protocols Fuse 10A T Fuse 0.5A F PULSANTE RX REMOTO (REMOTE RX PUSHBUTTON) 24Vdc SPIA READY (READY LAMP) SPIA RX (RX LAMP) 24Vdc H U B X 1 5 Figure 6-8 ROTOGRAPH-D (120V) (Rev. 4)
Page 38
SERVICE MANUAL Installation 18. Check the perpendicularity of ROTOGRAPH-D adjusting the 4 levelling screws located under the legs and assemble the support with the appropriate clamp. 19. Secure the unit to the floor by drilling the necessary 2 holes in the floor through the relevant predisposed holes on the legs, and using 2 8mm dowels (outer diameter = 8mm) provided with the unit.
3. Raise the column, position it in the desired spot, taking the total overall dimensions into account, and align it by means of a bubble level. 4. On the wall and floor mark the position of the fastening holes corresponding to the base plate and wall bracket. 6-10 ROTOGRAPH-D (120V) (Rev. 4)
Page 40
SERVICE MANUAL Installation 5. Put the column back on the floor, resting it on the wooden blocks, and remove the wall bracket. Figure 6-10 6-11 (Rev. 4) ROTOGRAPH-D (120V)
Page 41
To fasten the wall bracket to masonry we recommend expanding metal dowels with a screw diam. 8mm while for the base plate it will be sufficient dowels with screw diam.6mm. Where a wooden support is utilised, we recommend a la bolt of 3/8" diameter. 6-12 ROTOGRAPH-D (120V) (Rev. 4)
Page 42
8. Turn on the main switch on the control box and proceed to the calibrations, as described in chapter 7. 6-13 (Rev. 4) ROTOGRAPH-D (120V)
SERVICE MANUAL Installation How to mount the coverings How to mount the coverings How to mount the coverings How to mount the coverings Figure 6-12: Front view 6-14 ROTOGRAPH-D (120V) (Rev. 4)
AND CHECKED DURING THE FINAL TESTING IN THE FACTORY. ANY CHANGES TO THE DEFAULT SETTING MAY ONLY BE CARRIED OUT BY A VILLA SISTEMI MEDICALI TECHNICIAN OR BY EXPRESSELY AUTHORISED THIRD PARTIES. ANY CHANGES MADE BY UNAUTHORISED PEOPLE WILL VOID ANY MANUFACTURER'S RESPONSIBILITY AS TO THE CORRECT OPERATION AND SAFETY OF THIS EQUIPMENT.
Press "increase" (41) and "decrease" (40) during the self-test phase to increase or decrease the brightness of the display. To access the set-up phase, press "s" (39) during the self-test procedure. ROTOGRAPH-D (120V) (Rev. 4)
Non-volatile RAM (NVRAM) memory • press "T" (31) to calibrate the potentiometer used for the detection of the position of the rotating arm. WARNING: Pressing "R" will cause the loss of all factory and/or custom adjusted parameters. (Rev. 4) ROTOGRAPH-D (120V)
Once the desired setting is selected by pressing "s" (39), you can either go to the next functions by pressing "T" (31) or return to the normal working conditions by pressing "R" (32). NOTE: It is recommended to leave this function always ON (S2 ON). ROTOGRAPH-D (120V) (Rev. 4)
Page 50
Press "R" (32) to exit the S5 function. This will not set automatically function S5 to OFF (S5 OFF) and after an exposure the string "T1 020" will be displayed. To access the normal working function again, return to the set-up program and change S5 to OFF. (Rev. 4) ROTOGRAPH-D (120V)
Page 51
(0°÷245°) during all run. To return to 0°, press and keep pressed "R" (32). To exit the set-up function, press "s" (39). The SA value will automatically return to OFF (SA OFF). ROTOGRAPH-D (120V) (Rev. 4)
Page 52
"X-ray emission" button (42) the set-up procedure will terminate storing the set value. NOTE: At the pressure of TMJ1 or TMJ2 button while the SB=OFF value has been stored, the message "TMJ function not allowed" will be visualised on the display. (Rev. 4) ROTOGRAPH-D (120V)
This will automatically return to the normal working condition. NOTE: If you press "R" once by mistake, you can exit the "Non-volatile memory reset" function and return to the normal working condition by pressing any key (except "R"). ROTOGRAPH-D (120V) (Rev. 4)
4. If the values read at previous point "2" differ of more than ±1V from the one read at point "3", act on trimmer R85 of board A2 in order to obtain the matching of the two readings. 5. To return to the normal working condition, press "T" (31). (Rev. 4) ROTOGRAPH-D (120V)
2.4%. 5. In case line voltage regulation does not meet the spec, refer to paragraph 5.1 of this manual and consult with a certified electrician or the local Power Company. 7-10 ROTOGRAPH-D (120V) (Rev. 4)
Once the regulation is complete, tighten the screw on the resistor collar. 5. After the adjustment, set the S1 switch back to its initial position. 7-11 (Rev. 4) ROTOGRAPH-D (120V)
S4 back to OFF. 8. Switch OFF the unit, switch it back ON, activate set-program (see paragraph 7.2), press "s" (39) then "T" (31) several times until reading S4 ON. Press "s" (39) to change S4 to OFF. 7-12 ROTOGRAPH-D (120V) (Rev. 4)
Page 58
ROTOGRAPH-D does not usually require adjustment of the X-ray beam. However, if necessary, follow the instructions provided in the next paragraph to adjust the X-ray beam. WARNING: The followings test implies X-ray emission; please follow all measures envisaged by local safety regulations.
3. Make an exposure by depressing the X-ray emission button (42) and check the geometrical alignment of the visible beam and the slit aperture in the shield assembly. 7-14 ROTOGRAPH-D (120V) (Rev. 4)
4 screws. Check the alignment once more and repeat the procedure if necessary. At the end mount back the covers of the tubehead. Figure 7-3 7-16 ROTOGRAPH-D (120V) (Rev. 4)
Rotate the tubehead in order to align the X-ray beam with the fluorescent screen; screw back the 4 screws "104". Position the reference plate "D" fully in contact with the tubehead fixing plate and screw back the 2 screws "105". Figure 7-4 7-17 (Rev. 4) ROTOGRAPH-D (120V)
A5: X-ray emission start angle IV° exposure (can be changed) A6: X-ray emission end angle IV° exposure (can be changed) T2: Arm rotation time (cannot be changed) T3: Actual X-ray emission time (I°+IV° exp.) (cannot be changed) 7-18 ROTOGRAPH-D (120V) (Rev. 4)
Page 64
A5: X-ray emission start angle III° exposure (can be changed) A6: X-ray emission end angle III° exposure (can be changed) T2: Arm rotation time (cannot be changed) T3: Actual X-ray emission time (II°+III° exp.) (cannot be changed) 7-19 (Rev. 4) ROTOGRAPH-D (120V)
To access the normal working function again, return to the set-up program and change S5 to OFF. NOTE: Every change must be carried out by pressing the "increase" (41) and "decrease" (40) keys. 7-20 ROTOGRAPH-D (120V) (Rev. 4)
This setting determines the starting point of the rotation speed change (ramp up) which can be modified of max ± 3° with respect to the value indicated in the "Final Test table". 7-21 (Rev. 4) ROTOGRAPH-D (120V)
Page 67
Test Table". This function will be used, for instance, in case of a anticipated stop of X-ray emission that will cut the left condile. NOTE: Rotation time setting should be conducted according to paragraph 7.9.3. 7-22 ROTOGRAPH-D (120V) (Rev. 4)
Adult size, but the central area deceleration function is disabled and the activation of such function is not permitted. Also, the value of the string "A2 .." (X-ray emission start angle) can be changed from 14° (014.0) to 17° (017.0). 7-23 (Rev. 4) ROTOGRAPH-D (120V)
6. Press "s" (39) to display the string "A6 .." (X-ray emission end angle, 4th exposure). This setting can be modified of max ±2° with respect to the value indicated in the "Final test table". 7-24 ROTOGRAPH-D (120V) (Rev. 4)
Page 70
7. Press "s" (39) to display the string "A6 .." (X-ray emission end angle, 3rd exposure). This setting can be modified of max ±2° with respect to the value indicated in the "Final test Table". 7-25 (Rev. 4) ROTOGRAPH-D (120V)
2. Make an exposure in Adult mode at 85 kV, acquired by the Quickvision program; the following image will be displayed: 3. Set contrast and brightness level to have good visibility of all centering balls. 7-26 ROTOGRAPH-D (120V) (Rev. 4)
Page 72
Releasing the button, the following image shows up, where it is possible to read the distance in mm, which must be in the range 36 ± 1 mm. 7-27 (Rev. 4) ROTOGRAPH-D (120V)
Page 73
− if the measurement of the left side is greater than the right side, increase the angles − if the measurement of the left side is smaller than the right side, decrease the angles. 7-28 ROTOGRAPH-D (120V) (Rev. 4)
Error messages and Troobleshooting Error messages and Troobleshooting Error messages and Troobleshooting ROTOGRAPH-D is entirely run by a microprocessor that, besides setting the exposure parameters and correcting the main voltage fluctuation, signals by mean of messages on the display the different conditions of the system and the eventual anomalies and procedural mistake.
Page 75
Impossible to regulate See paragraph 7.8.2 the line voltage Memory data corrupted! See paragraph 7.8.3 Call Technical Assistance NO ANSWR See paragraph 7.8.4 Out of Order N° 1 ! Call See paragraph 7.8.5 Technical Assistance 7-30 ROTOGRAPH-D (120V) (Rev. 4)
X10 on Power board (A1) is properly inserted and not damaged; if needed replace it. NOTE: Villa is not responsible of making sure that the supplied line voltage is within the specified range. 7-31 (Rev. 4) ROTOGRAPH-D (120V)
K10 of the Power board used to compensate for line voltage value, change their status; if this does not happen it is possible that the relais network is failed and in this case replace the board. 7-32 ROTOGRAPH-D (120V) (Rev. 4)
Should these checkings have a negative result, replace the hand control or the CPU PCB in order to identify the origin of the failure. After having replaced one of the two components proceed with the system general checking. 7-33 (Rev. 4) ROTOGRAPH-D (120V)
(friction not adjusted, potentiometer gear too close to rotation gear, belt not properly positioned on the tensioning ball bearing, rotation ball bearings too tighten). If it is not the case, proceed with step 6 and following. 7-34 ROTOGRAPH-D (120V) (Rev. 4)
1. With the unit OFF, loosen the clutch of the belt motor. 2. Manually move gently the rotating arm to the 0° position, until reaching the relevant microswitch S2 (see Figure 7-5). 3. Check the value of the new potentiometer to install (5kΩ, 10kΩ or 20kΩ). 7-35 (Rev. 4) ROTOGRAPH-D (120V)
Page 81
16. Press "R" (32) key to bring the arm in starting position. 17. Switch the unit OFF, then back ON, activate the set-up programs (see above Note 2) and press the "s" (39) key. 18. Press "T" (31) key several times until showing SA OFF. 7-36 ROTOGRAPH-D (120V) (Rev. 4)
Page 82
23. Press the "R" (32) key to bring the arm in starting position. 245° Figure 7-5 24. Run the procedure to verify centering and symmetry of the panoramic, making an exposure on the round centering tool as described in paragraph 7.7.4. 7-37 (Rev. 4) ROTOGRAPH-D (120V)
9. This procedure is a trial and error process: repeat steps 6 to 9 until the reading of the potmeter voltage with the arm in 0° position is between 160mV and 180mV. 7-38 ROTOGRAPH-D (120V) (Rev. 4)
Page 84
A2 child = A2 adult +9° − A5 child = A5 adult -9° 17. Run a PAN child on the same centering tool and verify the symmetry of the image; if necessary fine-tune the angles. 7-39 (Rev. 4) ROTOGRAPH-D (120V)
15.3 sec. ± 0.1 sec. If different act on trimmer R39 (Speed 1) located on power board A1 (turning clockwise time becomes shorter) until the correct value is obtained. 7-40 ROTOGRAPH-D (120V) (Rev. 4)
Page 86
7.9.1 (steps 8 ÷ 23) and 7.7. NOTE: In case replacing the Power board, check the output current (mA) following instruction in paragraph 7.4. 7-41 (Rev. 4) ROTOGRAPH-D (120V)
The starting and stop angles of all examination must be stored according to the values present on the "Bollettino di collaudo" sheet, supplied with the equipment, following the procedure described on paragraph 7.7.1, 7.7.2 and 7.7.3; verify also all the configuration parameters. 7-42 ROTOGRAPH-D (120V) (Rev. 4)
When the adjustment is over, proceed as follows: 5. Turn the system OFF. Turn it ON again. Enter in SET-UP and set S4 to OFF. To return to the normal working conditions, press button "R" (32). 7-43 (Rev. 4) ROTOGRAPH-D (120V)
These values have to be written by the technical personnel, together with the date of change. 7-44 ROTOGRAPH-D (120V) (Rev. 4)
Page 90
Skull / Temple clamp control knob Patient positioning handle and slider Motor group unit Support unit of the motor group Slider Brake release button PAN selection button Slot for Memory Card TMJ1 selection button Focus position point TMJ2 selection button (Rev. 4) 7-45 ROTOGRAPH-D (120V)
In addition to the above checks (to be performed by the operator), checking to be performed by the Service Engineer are described in the following table. (Rev. 4) ROTOGRAPH-D (120V)
Page 93
Check the hardware safety timers See paragraph 7.10 Check the tension of the chains and Practical inspection belts and adjust if necessary Clean and lubricate all the carriage Practical inspection guides Check full steel cable integrity Visual inspection ROTOGRAPH-D (120V) (Rev. 4)
6. Circuit diagram Speed Control PCB 7. Layout CPU PCB A2 8. Circuit diagram CPU PCB A2 9. Layout Remote Control PCB 10. Circuit diagram Remote Control PCB 11. Layout Laser Power PCB 12. Circuit diagram Laser Power PCB (Rev. 5) ROTOGRAPH-D (120V)
Page 95
Potentiometer for angular control (assy) Arm return stop microswitch Arm advance stop microswitch Main switch (DPDT type) Brake switch right Cross beam light switch Autotransformer 800 VA Single-phase transformer 124 VA Vertical movement brake Filter 6.5A – 250V ROTOGRAPH-D (120V) (Rev. 5)
Page 96
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D Control Box layout Code 83087627 - Rev. P Page 1 of 1 (Rev. 5) ROTOGRAPH-D (120V)
Page 97
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D General diagram Code 39089348 - Rev. 2 Page 1 of 5 ROTOGRAPH-D (120V) (Rev. 5)
Page 98
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D General diagram Code 39089348 - Rev. 2 Page 2 of 5 (Rev. 5) ROTOGRAPH-D (120V)
Page 99
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D General diagram Code 39089348 - Rev. 2 Page 3 of 5 ROTOGRAPH-D (120V) (Rev. 5)
Page 100
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D General diagram Code 39089348 - Rev. 2 Page 4 of 5 (Rev. 5) ROTOGRAPH-D (120V)
Page 101
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D General diagram Code 39089348 - Rev. 1 Page 5 of 5 ROTOGRAPH-D (120V) (Rev. 5)
Page 102
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D Layout Power Supply PCB A1 Code 58087567 - Rev. 3 Page 1 of 1 (Rev. 5) ROTOGRAPH-D (120V)
Page 104
NOTA: VALORI DI TENSIONE TRA PARANTESI PER ALIMENTAZIONE 120V Fg.01 (Sheet 01) -E4 NOTE: DATA IN BRACKTES ARE FOR 120V ROTOGRAPH-D Circuit diagram Power Supply PCB A1 Code 39089322 - Rev. 2 Page 2 of 3 (Rev. 5) 9-11 ROTOGRAPH-D (120V)
Page 105
X17.12 +12V.RELE' Fg.02 -D-E- X17.14 X17.16 X17.18 MASSA CAMPO X17.20 Fg.01 (Sheet 01) -C1 AI RELE' (TO RELAIS) K16-K17-K18-K19 ROTOGRAPH-D Circuit diagram Power Supply PCB A1 Code 39089322 - Rev. 2 Page 3 of 3 ROTOGRAPH-D (120V) 9-12 (Rev. 5)
Page 106
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D Layout Speed Control PCB Code 58087565 - Rev. 4 Page 1 of 1 (Rev. 5) 9-13 ROTOGRAPH-D (120V)
Page 107
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D Circuit diagram Speed Control PCB Code 39089320 - Rev. 5 Page 1 of 1 ROTOGRAPH-D (120V) 9-14 (Rev. 5)
Page 108
SERVICE MANUAL Schematics and drawings ROTOGRAPH-D Layout CPU PCB A2 Code 58087570 - Rev. 5 Page 1 of 1 (Rev. 5) 9-15 ROTOGRAPH-D (120V)
Page 109
X20B - SPARE 74HC04 MICRO2 DS1232 ABILITAZIONE 74HC00 MICRO DA S10 A S12 74HC04 (SWITCHES S10 TO S12 ENABLE FOR) ROTOGRAPH-D Circuit diagram CPU PCB A2 Code 39089324 - Rev. 4 Page 1 of 6 ROTOGRAPH-D (120V) 9-16 (Rev. 5)
Need help?
Do you have a question about the ROTOGRAPH-D and is the answer not in the manual?
Questions and answers
need to reinstall softwre
To reinstall the software for the Villa Sistemi Medicali ROTOGRAPH-D, follow these steps:
1. If a memory corrupted error occurs, try restoring the operating condition by:
- Pressing in sequence: small size push button (37), child (34), and TMJ1 (29), as per SET-UP mode.
- Wait 5 seconds.
2. If condition is not restored:
- Replace the CPU (A2) PCB.
- Perform a general system check.
3. After replacing the CPU (A2):
- Re-enter all factory or custom parameters lost during the failure.
- Use the data recorded on the "Final Test Table" found in the User’s Manual.
- Follow the Set-up procedure in paragraph 7.2 and following.
This process effectively reinstalls the software and restores system settings.