Introduction The Druck UPS-II Loop Calibrator can supply power (source mode) and produce readings (measure mode) to perform field calibrations on 2-wire devices. Operation For different applications place switches in position as indicated below: MODE READ EXT. INT. Milliamp. source •...
Page 4
3 Batteries size AA Replacement Unscrew the centre screw (1). Remove the cover (2). Make sure the polarity of the new batteries (3) is correct. K0434 Issue No. 2 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 5
UPS-II Loop Calibrator Technical Notes Current reading Switch positions Loop power Function Ext. Read Transmitter Calibration Switch positions Loop power Function Read Int. K0434 Issue No. 2 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 6
Transmitter Simulation Switch positions Loop power Function Ext. Valve Stroking Switch positions Loop power Function Int. K0434 Issue No. 2 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
FLOW Adjust mA source to any value within range - check reading example VALVE FLOW Enter Input calibration complete new offset values stored continue with output calibration. K0434 Issue No. 2 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 9
FLOW Check reading. VALVE Enter FLOW Calibration complete new output values stored Completion Switch off and disconnect the calibration equipment. Update the calibration records for the instrument. K0434 Issue No. 2 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...