Signet 5700 pH/ORP Monitor - ORP Instructions English *3-5700.090-1* CAUTION! • Refer to instruction manual for more details. 3-5700.090-1 Rev. K 3/06 English/ORP • Remove power to unit before wiring input and output connections. • Follow instructions carefully to avoid personal injury.
To isolate output and prevent ground loop problems: 4-20 1. Use monitor device with isolated inputs, or 5700 2. Use separate DC supply for 5700 and monitor device, or Terminals 3. Power 5700 with12 - 24 VAC step down transformer 4. Relay Connections...
Press and release to access EASY CAL Procedure G. EASY CAL: press the right arrow Buffer calibration key to access the EASY CAL ↑A buffer calibration procedure (sec. See EASY CAL Procedure (Section 6.2) 6.2). Signet 5700 pH/ORP Monitor - ORP Instructions...
Page 4
Differences in excess of 50 mV may indicate a need to service the electrode. Refer to electrode manual for maintenance recommendations. Theoretical mV values pH @ 25 °C +296 +237 +177 +118 - 59 - 118 - 177 - 237 - 296 Signet 5700 pH/ORP Monitor - ORP Instructions...
Page 5
Quick Reference Calibration Procedures: 1-Point Calibration (optional): 2-Point Quick Calibration (recommended): 1. Set solution standard (step B) 1. Set solution standard (step B) 2. Set solution slope (step C) Signet 5700 pH/ORP Monitor - ORP Instructions...
+1000 96 mm 91 mm 88 mm (3.6 in.) (3.8 in.) (3.5 in.) Side View ENTER Signet Monitor 76 mm (3 in.) Front View 96 mm (3.8 in.) 88 mm (3.5 in.) (continued) Signet 5700 pH/ORP Monitor - ORP Instructions...
Last calibration 00 - 00 - 00 - date 39 - 39 - 99 EASY CAL Two pH buffers saturated procedure with Quinhydrone required: pH 4= +87 mV, pH 7= +264 mV, (sec. 6.2) Signet 5700 pH/ORP Monitor - ORP Instructions...
Page 9
-1000 mV -1000 - +2000 mV 20 mA setpoint -1000 - +1000 mV +2000 mV 4 mA adjust 3.0 to 5.0 mA 4.00 mA 20 mA adjust 19 to 21 mA 20.00 mA Signet 5700 pH/ORP Monitor - ORP Instructions...
Page 10
• Power instrument and verify 0 mV on display. If 0 mV is not displayed, calibrate Standard input for 0 mV (sec. 6.3B) • If instrument displays correctly after calibration, instrument is ok. If error condition persists, instrument requires factory service. (continued) Signet 5700 pH/ORP Monitor - ORP Instructions...
Page 11
Press any key to reload factory defaults then reprogram sys- ERROR saving setup menu entry tem setup parameters. Press any Key 11. Maintenance Clean the instrument case and front panel with a soft cloth and a mild soap solution. Notes: Signet 5700 pH/ORP Monitor - ORP Instructions...