Do you have a question about the HV-25 and is the answer not in the manual?

Questions and answers

Jess Pormiel
February 3, 2025

Hello. I have an Autoclave model HV-50L that beeps whenever a cycle is running or after it has completed. We have looked to the manual and tried to do suggested troubleshooting, but it continues to beep

1 comments:
Mr. Anderson
February 10, 2025

The beeping sound in the HIRAYAMA HV-50L autoclave could be caused by the following:

1. Error Detection Alarm – If a malfunction occurs, the error detection circuit is triggered, displaying an error number on the digital display and sounding an electronic alarm. Possible causes include:
- Lack of water in the system.
- Clogged piping due to a waste disposal bag or other obstruction.

2. Warming Cycle Notification – During the warming cycle (Modes 1 and 4), the electronic alarm beeps when the temperature in the working chamber falls to the set incubation temperature.

To stop the alarm:
- For an error, check the error number, press the START/STOP switch, and resolve the issue.
- For the warming cycle, wait for the cycle to complete or stop it manually by pressing the START/STOP switch.

This answer is automatically generated