Center due to the unique calibration process; test equipment utilized and properly trained SPY technicians. ® We DO NOT recommend using any other calibration service other than a SPY ® Authorized Service Center; using an outside service with technicians not trained to work on SPY brand equipment could damage the unit and if the enclosure is ®...
670 overview The NEW SPY Model 670 Wet Sponge Holiday Detector operates on a single 9 volt alkaline battery. Unit output voltage is 67.5 volts DC. This voltage makes the detector highly sensitive and accurate. Compact and light weight with no bulky battery packs allowing for non destructive inspection of paints, epoxy paints and coatings 1-20 Mils.
STEP 3. CONNECTING THE GROUND CABLE. The ground cable attach connecting accessories point is on the backside of the 670, untwist the black knob to expose the hole in the ground connector, thread the ground wire through the hole and then tighten the CAUTION! DO NOT TOUCH Make sure the switch is in the OFF position black ground connector knob tightly with your hand only.
STEP 2: The battery compartment is located on the left side of the unit, Attach the strap clips tabs on each side of the 670 detector. Adjust strap using the notch on the bottom of the battery compartment lift up and for comfort.
(test media). This will allow for a lower The test is performed by turning the 670 on and by pressing the 80 Kohm surface tension, which will help the test media to penetrate the thicker button, the signal will become active causing the horn and LED to be coatings.
STEP 6: Place the electrode sponge on the coated surface and begin troubleshooting guide passing the sponge across the surface (See Diagram 2 Below). Several passes should be made over the test area to ensure that moisture has PROBLEM POSSIBLE SOLUTION CORRECTION METHOD penetrated any possible holiday.
Page 9
troubleshooting guide troubleshooting guide PROBLEM POSSIBLE SOLUTION CORRECTION METHOD PROBLEM POSSIBLE SOLUTION CORRECTION METHOD HOLIDAY DETECTOR Check Detector Ground With the detector OFF Check Battery When the battery is DETECTOR Connection remove the ground HORN low the horn will sound WILL NOT cable, clean banana plug CONSTANT...
& repair technical specifications If maintained the Model 670 Wet Sponge Holiday Detector will offer you years of dependable operation. Regular care and maintenance of the unit is a requirement of good quality preventative maintenance practices. MODEL WARNING! This detector does not contain any user-serviceable Voltage Type parts;...
Electrode types The Model 670 uses cellulose sponge electrodes that vary in form and Pipeline Inspection Company, hereafter referred to as (SPY ® ) warrants that SPY ‚ Model 670 Series Holiday Detectors shall, under normal use ® size and fit a wide range of coating inspection projects. If we don’t carry and service, be free from defects in material and workmanship.
The Model 670 has a wide variety of accessories to assist you with your coating inspection project. MODEL 673 ® GROUND CABLE SHOULDER WET SPONGE (25 FT.) STRAP DETECTOR ITEM #11927 ITEM #10187 16 OZ. BOTTLE OF ROLLER SPONGE Please call our office at (713) 681-5837 or email us at sales@picltd.com...