Remove the Packing List and verify that you have received all equipment, including the following: Orbit Controls Model OC 502-t Handheld Calibrator. Operator's Manual OC 502-t. If you have any questions about the shipment, please call the Orbit Controls Customer Service Department. NOTE When you receive the shipment, inspect the container and equipment for signs of damage.
INDEX CALIBRATOR-MULTIMETER OC 502-t page 5 FUNCTION OUTPUTS and INPUTS L C D DISPLAY KEYBOARD SPECIFICATIONS CALIBRATOR – FUNCTION SELECTION Setting of the output current 0 - 22mA in steps or ramps Direct setting of the output current between 0 and 22mA Setting of the voltage output 0-14V in steps or ramps Setting of mV Outputs 0–560mV and 0–28mV in steps or ramps Thermocouple output in steps or ramps...
9 hour uninterrupted operation. Back ground light can be switched on. FUNCTIONS OC502-t can be set for following functions by using the keyboard and the digital display: Tara of the analogue outputs, the filter selection, Range selection 0-28mV or 0-560mV background light.
Calibrator Current Source 0 ... 22 mA Current Sink 0 ... 22 mA Voltage Source 0 ... 14 V mV Source 0 … 28.000 mV and 0 … 560.00 mV DIN Thermometers J, K, N, R, S, T, B, E The calibrator functions are settable with three sliding switches: ON-OFF Power ON and OFF...
KEYBOARD SPECIFICATIONS CALIBRATOR Outputs + OUT - Current Source Range 0 ... 22mA, max. Load 750 Ohm Accuracy ± (0.05% from Value + 0.1% from Range) Option: 0 … 110mA, see page 16: OC502-V2-100. Current Sink Range 0 ... 22mA @ max. 24VDC Accuracy ±...
CALIBRATOR - FUNCTION SELECTION The calibrator functions are selectable with three sliding switches. The keyboard permits setting of Ramps, Steps or direct Value entry. The Multimeter is active all the time and can be simultaneously used with the calibrator. The function is described in §...
Direct setting of the output voltage value between 0 and 14 V, 0-560mV, 0-28mV and T/C Sliding switch set to OUT-V. The key Num activates the numeric selection. Enter the required value with the keyboard and press Enter. To erase the entry, use the keys ← and → . Selection of the display function The key MENU selects following operation modes –...
MULTIMETER - FUNCTION SELECTION The multimeter is active all the time and can be simultaneously used with the calibrator. The calibrator output and the multimeter inputs have common GND. The upper part of the display shows the values and parameters of the Multimeter: AUTO Autoranging ±...
8.1 CALIBRATION of CALIBRATOR OUTPUTS 8.1.1 Current Output 4-20mA Outputs + OUT - with connected mA-Meter. Sliding switch OUT-I. Press 1. The display shows: With the keys Up or Down appears the calibration constant 0. To change it, enter 11 with Up or Down. The whole constant can be set with the keyboard.
8.1.3 Voltage Ranges 0-14V, 0-560mV, 0-28mV Output Terminals + OUT – with external V-Meter connected. Sliding Switch in Position OUT-V. Set the output voltage 2V with Up or Down and store with Enter. Select 10V Range with Menu, set the calibration Constant with Up or Down and store with Enter.
8.2.4 Voltage input 200VDC Inputs GND and IN-V supplied from external voltage calibrator 200VDC. Press 5. The display shows: Short the inputs. Press Enter to store the zero input. Press Menu. The display changes to: Apply 200VDC from external calibrator. Press Enter to store the 200V full range input.
TARE The Tare is active in the Multimeter function and can be used for DC-V and DC-I measurements to set the display to zero. Attention has to be paid when the Tare is set. The entire measuring range has to be considered in order not to overrange the meter.
√ Rechargeable Battery √ Software calibration Orbit Controls Model OC 502-V2-100 is a Calibrator-Multimeter for generation of 0/4 - 110mA in Source and Sink mode and Voltages between 0 to 11V. External voltages ± 2V, ± 20V and ± 200VDC (firm ranges or Autoranging) and currents up to ±100mA can be measured...
1.0000 V 10.000 V 100.000 V 50.000 mA 2.0000 V 20.000 V 200.000 V 100.000 mA Technician ........ORBIT CONTROLS AG Zürcherstrasse 137 QC: ....CH-8952 Schlieren/ZH Tel: +41 44 730 2753 Fax: +41 44 730 2783 Date: ……………………………..info@orbitcontrols.ch...
We will be pleased to answer all your questions not only to this instrument but also to all our calibrators and measuring equipment. Please call our customer service or write to us info@orbitcontrols.ch Thank you. Orbit Controls AG Customer Service Zürcherstrasse 137 Tel: +41 44 730 2753 8952 Schlieren-Zürich...