® Universal Tester Operating Instructions Model 18029 WARNING Read these instructions, particularly the Warnings and Cautions, before using the WanderGuard Universal Tester. Failure to do ® so could result in injury or death to a person in your care. 0163-390...
Page 2
Trademark owner. ARIAL, AUTISM ALLY, AUTISM ALLY design, the CATCH-ALL design, SENIOR TECHNOLOGIES, TABS, TIM, WANDERGUARD, and the WANDERGUARD design are some of the registered and unregistered trademarks owned by Stanley Logistics, Inc. and/or its Affiliates ("Stanley Trademarks"). Stanley...
Page 3
WARNING Test bracelets daily as detailed in these instructions, in the bracelet instructions, and the WanderGuard ® Door Monitor User Manual. Record the results in the resident’s records. It is important to test the bracelet before putting it into use and daily thereafter.
Page 4
Features of the WanderGuard ® Universal Tester The Universal Tester is a hand-held, battery-powered device used to test ® ® WanderGuard door monitors and WanderGuard bracelets. When using the tester to test bracelets, the tester will detect if a bracelet...
Page 5
The tester has an ambient light sensor built-in. When the tester is used in low light conditions, it will automatically back-light the LCD display. Using the RECALL button will display the result of the last bracelet test. AMBIENT LIGHT SENSOR BRACELET STATUS DOOR STATUS LED BRACELET...
Page 6
BEFORE the bracelet becomes inoperable. Bracelets must be tested daily. IMPORTANT: Failure to replace the bracelet “on time” could result in injury to or the death of a person in your care. To test a WanderGuard ® bracelet with ID capability (36000 series): •...
Page 7
• If the bracelet is still operational, the LED will flash green four times and the ID number of the bracelet will be displayed. LED blinks GREEN 4 times • If the bracelet has reached the End-of-Service phase, the LED will be a steady red.
Page 8
® To test a WanderGuard bracelet WITHOUT ID capability (32000 Series): • Hold the Universal Tester within one foot of the bracelet. • Press the BRACELET button once. The LED above the button will light briefly. Press button The display will...
Page 9
• If the bracelet is not detected, the LED will be a steady red and display “- - -”. LED is steady If the Universal Tester does not detect the bracelet: • Move the tester closer to the bracelet. • Hold the tester at a different orientation in relation to the bracelet.
Page 10
Testing Door Monitors Using the Door Test Function The Universal Tester will test all WanderGuard ® door monitors except the FirstQ series. Test door monitors at least weekly on each shift with all surrounding power devices turned on to maximize possible electromagnetic interference.
Page 11
2. Move to X inches in front of the middle of the door, or middle of doorway if testing a double door system, and hold the tester at waist level. 3. Press the DOOR button. The alarm on the WanderGuard ® door monitor will sound as long as the door button is held.
Page 12
® To test a WanderGuard door coverage: 1. While holding the tester in a vertical position (screen facing you), place the tester in one of the positions illustrated below - you may have to kneel to keep the tester in the vertical position.