Electronic Sphygmomanometer
User's Manual
Guangdong Biolight Meditech Co., Ltd.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the WBP100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Biolight Meditech WBP100

  • Page 1 Electronic Sphygmomanometer User’s Manual Guangdong Biolight Meditech Co., Ltd.
  • Page 2 Contents of this manual are subject to change without prior notice.  Document No.: J/WBPCE-A-008  Revision number: V1.0  Release time: Sept. 2015 Copyright © 2015 Guangdong Biolight Meditech Co., Ltd. All rights reserved. CE mark EC Representative Name: Shanghai International Holding Corp.
  • Page 3: About This Manual

    Statement Manufacturer holds the copyright of this manual, and we are also entitled to deal with this manual as confidential files. This manual is only used for operation, maintenance and service of product, someone else can not publish the manual. This manual contains exclusive information protected by copyright laws and we reserve its copyright.
  • Page 4 Illustrations are for reference only. They may be not in conformity with your products. Signs in this manual: Warning: Indicates a potential hazard or unsafe practice that, if  not avoided, will result in death or serious injury. Caution: Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.
  • Page 5: Table Of Contents

    Electronic Sphygmomanometer User’s manual Contents 1、 Intended Use ......................... 1 2、 Application ..........................1 3、 Safety Information ........................ 1 4、 Appearance ........................... 4 5、 Display ..........................6 6、 Setting ........................... 7 7、 Historical data review function ..................... 8 8、 Download And Install The Software ..................8 9、...
  • Page 6: Intended Use

    Electronic Sphygmomanometer User’s manual 1、 Intended Use The device is intended to measure blood pressure and pulse rate of adult. The measurement data is available for clinical reference. It can store, manage the measure data, and transmit the measuring result via wireless signal. It is intended to be used in clinic, general ward, physical examination department and at home.
  • Page 7 Electronic Sphygmomanometer User’s manual pregnant, pre-eclampic patients or people with serious arrhythmia. Consult your physician before use on children;  Don’t use the electronic sphygmomanometer in the environment containing inflammable anesthetic gas;  Consult your physician for any of the following situations: a) The application of the cuff over a wound or inflamed area;...
  • Page 8 Electronic Sphygmomanometer User’s manual adapter must be provided or designated by manufacturer (see optional item list). If use another power adapter, it should be verified, otherwise electric shock may occur and cause patient death;  The effective wireless transmission distance is 5m in a barrier-free environment.
  • Page 9: Appearance

    After smoking; f) After drinking alcohol, coffee and black tea; g) After the movement; h) Room temperature changes suddenly i) Within an hour after a meal; j) Mood be agitated when caused by tension, uneasy. Appearance 4、  WBP100:...
  • Page 10 Electronic Sphygmomanometer User’s manual  WBP201/WBP202: 1、 Cuff Display screen 2、 Display systolic blood pressure, diastolic blood pressure, pulse rate and so on. 3、 Power button;Start/Stop button — Long press (press and hold it more than 2s) this button to turn off the device;...
  • Page 11: Display

    Electronic Sphygmomanometer User’s manual 5、 Select users button — In measurement sate: Short press this button to select users; — In setting state: Short press this button, the time value would increase one. 6、 Historical data review button — Not in setting state: Short press this button to review historical data. Long press it to enter setting mode.
  • Page 12: Setting

    Electronic Sphygmomanometer User’s manual Error code Definition Possible Reason Measurement range exceeds module specification of 255 Range exceeded mmHg maximum. a. Too many retries due to interference of motion artifact; Excessive motion b. Signal is too noisy during step-down to detect pressure pulse, e.g.
  • Page 13: Historical Data Review Function

    Electronic Sphygmomanometer User’s manual 7、 Historical data review function This device can record the blood pressure, pulse rate and time automatically to facilitate the management of blood pressure. Press the historical data review button “ ”, the latest measurement result would be displayed on the screen. Press the button “...
  • Page 14 Electronic Sphygmomanometer User’s manual  Loosen the calibre of the curly cuff slightly. Place a bare arm through the cuff directly according to the instruction icon “ ”, “ ”. Position the cuff 1-2cm above the elbow joint and ensure the mark “Φ” on the cuff shall lie above artery.
  • Page 15 Electronic Sphygmomanometer User’s manual  Note:  Incorrect measuring posture showed as follows might cause an incorrect  measurement: a) Bent over(lean forward); b) Sitting cross-legged; c) Bend the body because of the low table or sitting on sofa. (3) Measuring ...
  • Page 16: Maintenance And Cleaning

    Electronic Sphygmomanometer User’s manual : Caution  You can stop the measurement process at any time by pressing “Start/Stop button”.  Warning:  Prolonged and frequent NIBP measurements are associated with ischemia and neuropathy in the limb wearing the cuff. Please examine the extremities of the limb frequently for normal color, warmth and sensitivity.
  • Page 17: Accessory And Specifications

    Accessory and Specifications 11.1 Optional item Items Mode Power adapter (optional) LXCP12-05(Output DC5V,1.0A) 11.2 Technical Specifications Production model WBP100、WBP201、WBP202 Size 130 mm(L) x 55 mm(W) x 30 mm(H) <350g Weight Measure specifications Measurement way Automatic oscillometry Measurement range Systolic: 30 mmHg~255 mmHg(4.0kPa~34.0kPa);...
  • Page 18 Electronic Sphygmomanometer User’s manual Software overpressure 295± 5 mmHg protection Normally, it is 20s to 45s (depending on HR and moving interference typically) Measurement time Maximal measurement time: 120s LED indicating lamp 1 (Blue) Bluetooth Not connected: The lamp flashes blue; indicating Connected: The lamp lights blue without flashing;...
  • Page 19: Troubleshooting

    Electronic Sphygmomanometer User’s manual Large cuff 32 cm~42cm(Optional) Medium cuff 22 cm~32cm(With) Small cuff 17 cm~22cm(Optional) Adapter Input 100~240V, 50/60Hz 0.3Amax Output DC 5V, 1A Environment specifications Operating: 5℃~+40℃; Temperature Transportation and storage: -20℃~+55℃. Operating: 25%~95%(non condensing); Humidity Transportation and storage: 25%~95% (non condensing). Operating: 700hPa~1060hPa;...
  • Page 20: Equipment Symbols

    Electronic Sphygmomanometer User’s manual 13、 Equipment Symbols Symbol Symbol Note Type BF applied part, without defibrillation protected. Refer to this user’s manual. Symbol for the marking of electrical and electronics devices according to Directive 2002/96/EC Batch code Serial number IP22 Enclosure degree of ingress protection.
  • Page 21: Classification Of Blood Pressure

    Electronic Sphygmomanometer User’s manual 14.2 Classification of blood pressure The World Health Organization (WHO) has created the following blood pressure classify table for assessing high blood pressure (without regard age or gender). Please note that other factors (e.g. diabetes, obesity, smoking, etc.) also need to be considered.
  • Page 22: Appendix A Guidance And Manufacturer's Declaration Of Emc

    Appendix A Guidance and Manufacturer’s Declaration of EMC  Note:   This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment . ...
  • Page 23 Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should assure that it is used in such an environment. Electromagnetic environment Immunity test IEC 60601 test level Compliance level - guidance...
  • Page 24 Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Immunity IEC 60601 Compliance Electromagnetic environment - guidance test...
  • Page 25 Recommended separation distances between portable and mobile RF communications equipment and the device . The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.