Statement Manufacturer holds the copyright of this manual, and we are also entitled to deal with this manual as confidential files. This manual is only used for operation, maintenance and service of product, someone else can not publish the manual. This manual contains exclusive information protected by copyright laws and we reserve its copyright.
Page 4
Illustrations are for reference only. They may be not in conformity with your products. Signs in this manual: Warning: Indicates a potential hazard or unsafe practice that, if not avoided, will result in death or serious injury. Caution: Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.
Electronic Sphygmomanometer User’s manual 1、 Intended Use The device is intended to measure blood pressure and pulse rate of adult. The measurement data is available for clinical reference. It can store, manage the measure data, and transmit the measuring result via wireless signal. It is intended to be used in clinic, general ward, physical examination department and at home.
Page 7
Electronic Sphygmomanometer User’s manual pregnant, pre-eclampic patients or people with serious arrhythmia. Consult your physician before use on children; Don’t use the electronic sphygmomanometer in the environment containing inflammable anesthetic gas; Consult your physician for any of the following situations: a) The application of the cuff over a wound or inflamed area;...
Page 8
Electronic Sphygmomanometer User’s manual adapter must be provided or designated by manufacturer (see optional item list). If use another power adapter, it should be verified, otherwise electric shock may occur and cause patient death; The effective wireless transmission distance is 5m in a barrier-free environment.
After smoking; f) After drinking alcohol, coffee and black tea; g) After the movement; h) Room temperature changes suddenly i) Within an hour after a meal; j) Mood be agitated when caused by tension, uneasy. Appearance 4、 WBP100:...
Page 10
Electronic Sphygmomanometer User’s manual WBP201/WBP202: 1、 Cuff Display screen 2、 Display systolic blood pressure, diastolic blood pressure, pulse rate and so on. 3、 Power button;Start/Stop button — Long press (press and hold it more than 2s) this button to turn off the device;...
Electronic Sphygmomanometer User’s manual 5、 Select users button — In measurement sate: Short press this button to select users; — In setting state: Short press this button, the time value would increase one. 6、 Historical data review button — Not in setting state: Short press this button to review historical data. Long press it to enter setting mode.
Electronic Sphygmomanometer User’s manual Error code Definition Possible Reason Measurement range exceeds module specification of 255 Range exceeded mmHg maximum. a. Too many retries due to interference of motion artifact; Excessive motion b. Signal is too noisy during step-down to detect pressure pulse, e.g.
Electronic Sphygmomanometer User’s manual 7、 Historical data review function This device can record the blood pressure, pulse rate and time automatically to facilitate the management of blood pressure. Press the historical data review button “ ”, the latest measurement result would be displayed on the screen. Press the button “...
Page 14
Electronic Sphygmomanometer User’s manual Loosen the calibre of the curly cuff slightly. Place a bare arm through the cuff directly according to the instruction icon “ ”, “ ”. Position the cuff 1-2cm above the elbow joint and ensure the mark “Φ” on the cuff shall lie above artery.
Page 15
Electronic Sphygmomanometer User’s manual Note: Incorrect measuring posture showed as follows might cause an incorrect measurement: a) Bent over(lean forward); b) Sitting cross-legged; c) Bend the body because of the low table or sitting on sofa. (3) Measuring ...
Electronic Sphygmomanometer User’s manual : Caution You can stop the measurement process at any time by pressing “Start/Stop button”. Warning: Prolonged and frequent NIBP measurements are associated with ischemia and neuropathy in the limb wearing the cuff. Please examine the extremities of the limb frequently for normal color, warmth and sensitivity.
Accessory and Specifications 11.1 Optional item Items Mode Power adapter (optional) LXCP12-05(Output DC5V,1.0A) 11.2 Technical Specifications Production model WBP100、WBP201、WBP202 Size 130 mm(L) x 55 mm(W) x 30 mm(H) <350g Weight Measure specifications Measurement way Automatic oscillometry Measurement range Systolic: 30 mmHg~255 mmHg(4.0kPa~34.0kPa);...
Page 18
Electronic Sphygmomanometer User’s manual Software overpressure 295± 5 mmHg protection Normally, it is 20s to 45s (depending on HR and moving interference typically) Measurement time Maximal measurement time: 120s LED indicating lamp 1 (Blue) Bluetooth Not connected: The lamp flashes blue; indicating Connected: The lamp lights blue without flashing;...
Electronic Sphygmomanometer User’s manual 13、 Equipment Symbols Symbol Symbol Note Type BF applied part, without defibrillation protected. Refer to this user’s manual. Symbol for the marking of electrical and electronics devices according to Directive 2002/96/EC Batch code Serial number IP22 Enclosure degree of ingress protection.
Electronic Sphygmomanometer User’s manual 14.2 Classification of blood pressure The World Health Organization (WHO) has created the following blood pressure classify table for assessing high blood pressure (without regard age or gender). Please note that other factors (e.g. diabetes, obesity, smoking, etc.) also need to be considered.
Appendix A Guidance and Manufacturer’s Declaration of EMC Note: This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment . ...
Page 23
Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should assure that it is used in such an environment. Electromagnetic environment Immunity test IEC 60601 test level Compliance level - guidance...
Page 24
Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Immunity IEC 60601 Compliance Electromagnetic environment - guidance test...
Page 25
Recommended separation distances between portable and mobile RF communications equipment and the device . The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.