MAINTENANCE NOTICE
NDI 450
N O R M A L A B F r a n c e
1/118

Advertisement

Table of Contents
loading

Summary of Contents for NORMALAB NDI 450

  • Page 1 NORMALAB France 2012 NORMALAB France MAINTENANCE NOTICE NDI 450 N O R M A L A B F r a n c e 1/118...
  • Page 2: Table Of Contents

    NORMALAB France 2012 1 CONTENTS 1 CONTENTS 2 STRUCTURE OF THE MACHINE 2.1 M ECHANICAL STRUCTURE 2.1.1 H EATING BLOCK 2.1.2 S UBSET COMPARTMENT 2.1.3 S UBSET CONDENSER 2.1.4 T HE VENTILATION 2.2 E NVIRONNEMENT OF THE MACHINE 2.3 E LECTRONIC STRUCTURE 2.3.1 E...
  • Page 3 NORMALAB France 2012 3.4 C ONDENSER 3.4.1 R EPLACEMENT CONDENSER PROBE 3.5 C OOLING CONDENSER SYSTEM 3.5.1 R EPLACEMENT OR DRAIN OF THE COOLANT LIQUID CIRCUIT 3.5.2 R EPLACEMENT OF PUMP 3.5.3 R EPLACEMENT OF FILTER OR FAN 3.5.4 R EPLACEMENT OF FAN 3.5.5 R...
  • Page 4 NORMALAB France 2012 4.7 C ALIBRATION OF TEST TUBES 4.8 C ALIBRATION OF THE TACTILE PAD 4.9 M ONITORING EXPIRY OF CALIBRATION DEADLINES 5 DENOMINATION AND FUNCTIONS OF THE ELECTRONIC BOARDS 5.1 P ROCESS 5.2 MMI CPU 5.3 CL1 BOARD 5.4 CL2...
  • Page 5 NORMALAB France 2012 9.7 I MPORTANT RECOMMENDATIONS 9.8 A NDI450 DDITIONAL REQUIREMENTS TO UPDATE YOUR INSTRUMENT 9.9 P REPARE YOUR FLASH DRIVE FOR UPDATING VIDEO EXAMPLE 9.10 I NSTALL A NEW VERSION USING THE PDATE OPTION VIDEO EXAMPLE 9.11 I...
  • Page 6: Structure Of The Machine

    NORMALAB France 2012 2 Structure of the machine 2.1 Mechanical structure 2.1.1 Heating block 2.1.2 Subset compartment N O R M A L A B F r a n c e 6/118...
  • Page 7: Subset Condenser

    NORMALAB France 2012 2.1.3 Subset condenser 2.1.4 The ventilation N O R M A L A B F r a n c e 7/118...
  • Page 8: Environnement Of The Machine

    NORMALAB France 2012 2.2 Environnement of the machine The machine must be installed on the stable bench , can supported the weight of the apparatus ( see characteristics) It is necessary to respect a minimum environment volume to ensure proper operation and longevity Allow minimum space : 25cm behind and 40 cm on the sides , 100 cm above to ensure adequate ventilation space.
  • Page 9: Electronic Structure

    NORMALAB France 2012 2.3 Electronic structure 2.3.1 Electronic structure 2.3.2 Location of the electronic boards CL2 board and Stepper motor board board N O R M A L A B F r a n c e 9/118...
  • Page 10 NORMALAB France 2012 CL56F board CL56F board CL56C Inverter pour board backlight CL56F MMI CPU board board Inverter pour backlight Touch screen’s driver Hard drive disk N O R M A L A B F r a n c e...
  • Page 11: The External Connections

    NORMALAB France 2012 Power supply Power supply for for CPU stepper motor CPU Process CL4 Board CL1 Board 2.3.3 The external connections 2.3.3.1 Keyboard, mouse, printer, monitor, network Devices type Port Video monitor VGA port on the back side Keyboard...
  • Page 12: Instruction Sheets

    NORMALAB France 2012 3 Instruction sheets 3.1 Access to parts of the machine 3.1.1 Opening the panels N O R M A L A B F r a n c e 12/118...
  • Page 13: Removing The Front Panel

    NORMALAB France 2012 3.1.2 Removing the front panel Tools 1 hex head key 2 mm Remove the side and rear panels (see 1 hex head key 2.5 mm above procedure 3.1.1) 1 hex head key 3 mm Remove the compartment door : remove the 4...
  • Page 14 NORMALAB France 2012 Remove the holding screw on the left side Remove the 5 screw of the front panel Remove this holding screw Remove the 2 black knobs fixing the electronic drawer Pull slightly the panel forward Unplug the ribbon of front panel...
  • Page 15: Access To The Electronic Drawer

    NORMALAB France 2012 Unplug the network cable ( RJ45) The front panel is free , you can remove it 3.1.3 Access to the electronic drawer Remove the 2 screws of the upper left panel and lift the panel on the back...
  • Page 16: Level Follower

    NORMALAB France 2012 3.2 Level follower 3.2.1 Replacement of the 1rst drop optical sensor Remove the holding screw Pull out the optical sensor from the slide Remove the Jack connector Plug the connector on the new optical sensor Locking Holding...
  • Page 17: Replacement Of The Follower Optical Sensor

    NORMALAB France 2012 3.2.2 Replacement of the follower optical sensor With the maintenance functions ,to do raise the follower sensor of 5 cm to release it from the follower base Proceed the same way as first drop sensor to remove...
  • Page 18: Replacement Of The Limit Switches Of Follower

    NORMALAB France 2012 3.2.3 Replacement of the limit switches of follower Tools 1 hex head key 4 mm 1 tubular key 5.5 mm 1 flat blade screwdriver 3.5m Remove the right panel (see the above procedure 3.1.1) Position of the low and high sensors...
  • Page 19: Replacement Of Drip Deflector Detector

    NORMALAB France 2012 3.2.4 Replacement of drip deflector detector Tools 1 hex head key 3 mm 1 wrench 7mm Remove the front panel (see above procedure 3.1.1) Remove the roof of the receiver Remove the flange around the condenser compartment(2 hex screws)and extract it...
  • Page 20 NORMALAB France 2012 Remove the 2 maintaining scews Remove the compartment probe Lift the probe above condenser Release the wires and turn the holder around the condenser ( lift slightly the condenser tube) Vis de maintien capteur Capteur Place the holder in front of condenser tube...
  • Page 21 NORMALAB France 2012 Once the screws to release the spring, the sensor can Remove completely the screw and close the be extracted (mark the direction of the sensor) hole to maintain the spring Replace the medium in the slide, put the spring, push...
  • Page 22: Replacement Of Step By Step Motor

    NORMALAB France 2012 3.2.5 Replacement of step by step motor Tools 1 hex head key 2.5 mm 1 hex head key 2 mm 1 hex head key 1.5 mm 1 hex head key 4 mm Remove the follower optical sensor (see above procedure 3.2.2)
  • Page 23 NORMALAB France 2012 Return the guide assembly Loosen the motor coupling Remove the motor At the reassembling , replace correctly the insulation N O R M A L A B F r a n c e 23/118...
  • Page 24: Replacement Of The Motor Board

    NORMALAB France 2012 3.2.6 Replacement of the motor board Tools 1 hex head key 2 mm 1 tube wrench 5.5 mm Remove the CL2 board with the 4 screws Remove the connector of the motor board Remove the motor board...
  • Page 25: Replacement Of Detector Or Magnet Of Compartment Door

    NORMALAB France 2012 3.2.7 Replacement of detector or magnet of compartment door Tools 1 hex head key 2 mm Remove the front panel of the apparatus (see above procedure 3.1.2) Remove the sensor Identify the position of the sensor Red wire on 7 of J1...
  • Page 26: Heating Block

    NORMALAB France 2012 3.3 Heating block 3.3.1 Replacement of the heating element Tools 1 cutting pliers Apparatus power off , and cold heating element Pull verticaly the heating assembly Carefully pry the heating element Access to the connectors of heating element and...
  • Page 27: Replacement Of Turbine

    NORMALAB France 2012 Plug the new element ( heating and Replace by clipping the element thermocouple) 3.3.2 Replacement of turbine Tools 1 tube wrench 7 mm 1 soldering iron welding Remove the 3 fixing screws Follow the wires in the raceway to remake the...
  • Page 28: Replacement Of Fire Detection Thermocouple

    NORMALAB France 2012 3.3.3 Replacement of fire detection thermocouple Tools Cutting pliers Remove the front panel (see above procedure 3.1.2) Position of thermocouple Cut the collar , unplug the thermocouple N O R M A L A B F r a n c e...
  • Page 29: Replacement Of Vapor Probe's Detector

    NORMALAB France 2012 3.3.4 Replacement of vapor probe’s detector Remove the left upper panel (see above procedure3.1.2) Tools 1 hex head key 2.5 mm Position of the sensor Remove the circuit breaker holder Remove the fixing screws of sensor board and...
  • Page 30: Replacement The Mechanical System Of Sample Probe Detection

    NORMALAB France 2012 3.3.5 Replacement the mechanical system of sample probe detection Remove the left upper panel (see above procedure 3.1.1) Tools 1 hex head key 2.5 mm Remove the fixing screw Take care at spring 3.3.6 Replacement of heating block door’s detector Remove the front panel (see above procedure 3.1.2)
  • Page 31: Condenser

    NORMALAB France 2012 3.4 Condenser 3.4.1 Replacement condenser probe Tools 1 hex head key 3 mm 1 wrench 7 mm Remove the front panel (see above procedure 3.1.2) Remove the roof of the compartment (2 screws) Loosen the 2 screws of the flange and deposit it...
  • Page 32: Cooling Condenser System

    NORMALAB France 2012 3.5 Cooling condenser system 3.5.1 Replacement or drain of the coolant liquid circuit Remove the right and rear panels (see above procedure 3.1.1) Open the cap of the expansion tank Open the valve to drain Note: The liquid volume to fill circuit is approximately 250 mL 3.5.2 Replacement of pump...
  • Page 33 NORMALAB France 2012 Position of the pump Remove the 2 fixing screws Or remove the 4 nuts and lift the assembly and And remove the 2 fixing screws of the pump rotate it Remove insulation fitting and disconnect them Remove the frame fixations...
  • Page 34 NORMALAB France 2012 Remove the ribbons of CL3 board Release the power supply frame on the left Unplug the wires of the board White wire in 1 and black wire on 2 of J10 At the reassembling ,properly position the frame...
  • Page 35: Eplacement Of Filter Or Fan

    NORMALAB France 2012 3.5.3 Replacement of filter or fan Remove the side and rear panels (see above procedure 3.1.1) Push down with blade of the screwdriver To unclip the filter holder Clean or replace the filter Replace the filter holder with filter...
  • Page 36: Eplacement Of Fan

    NORMALAB France 2012 3.5.4 Replacement of fan Remove the left panel (see above procedure 3.1.1) Remove the filter holder (see above procedure 3.5.3) Tools 1 screwdriver 3 mm 1 tube key 7 mm Remove the 4 fixing nuts Remove the power supply wires 3.5.5 Replacement of fan speed thermostat...
  • Page 37: Replacement Of Peltier Subset

    NORMALAB France 2012 3.5.6 Replacement of Peltier subset Remove the front panel (see above procedure 3.1.2) Drain the liquid circuit (see above procedure 3.5.1) Tools 1 wrench 10 mm 1 hex head key 3 mm 1 hex head key 2.5 mm...
  • Page 38 NORMALAB France 2012 Remove the level detector ( 3 screws) Remove the tank Raise the assembly to release the 4 studs Release the assembly to unplug the pilot wires Unplug the pilot wires : orange , red , green and...
  • Page 39 NORMALAB France 2012 Unplug the power supply wires Extract the assembly N O R M A L A B F r a n c e 39/118...
  • Page 40: Djustment Or Replacement Of The Liquid Level Detector

    NORMALAB France 2012 3.5.7 Adjustment or replacement of the liquid level detector Tools 1 hex head key 2.5 mm 1 screwdriver 1.5 mm Remove the side and rear panels (see above procedure 3.1.1) Remove the 3 fixing screws of sensor...
  • Page 41: Replacement Of Subset Peltier Power Supply

    NORMALAB France 2012 3.5.8 Replacement of subset Peltier power supply Remove the right and rear panel (see above procedure 3.1.1) Tools 1 cross tip screwdriver Unplug the power supply wires Unplug the main supply wires Remove the 2 screws Remove the power supply...
  • Page 42: Compartment Cooling System

    NORMALAB France 2012 3.6 Compartment cooling system 3.6.1 Replacement of the filters or fans Remove the right and rear panel (see above procedure 3.1.1) Tools 1 screwdriver 3.5 mm Position of filters and fans Pull the filter holder to unclip...
  • Page 43: Replacement Of The Peltier Subset Of Receiver Compartment

    NORMALAB France 2012 3.6.2 Replacement of the Peltier subset of receiver compartment Tools 1 screwdriver 2.5 mm 1 iron solder 1 cutting plier 1 wrench 7 mm 1 wrench 10 mm Remove the side and rear panels (see above procedure 3.1.1) Remove the front panel (see above procedure 3.1.2)
  • Page 44 NORMALAB France 2012 Unplug the 2 ribbons of CL3 board Remove the roof of the compartment Extract the roof Remove the flange of condenser Remove the condenser fixing in the heating Remove the first drop optical sensor block part of condenser ( 3 screws)
  • Page 45 NORMALAB France 2012 Remove the centering condenser in Remove the 4 fixing screws of subset compartment Release the centering condenser Extract the subset by forward rotation N O R M A L A B F r a n c e...
  • Page 46: Eplacement Of The Compartment Probe

    NORMALAB France 2012 3.6.3 Replacement of the compartment probe Tools 1 hex head key 2.5 mm Remove the side and rear panels (see above procedure 3.1.1) Remove the front panel (see above procedure 3.1.2) Remove the first drop sensor ( without Unplug the sensor modifying the adjustment) see procedure 3.2.1)
  • Page 47: Replacement Of Subset Peltier Power Supply

    NORMALAB France 2012 3.6.4 Replacement of subset Peltier power supply See above procedure 3.5.8 3.7 Chassis 3.7.1 Replacement of the heating transformer Tools 1 screwdriver 3.5 mm 1 hex head key 5.5 mm 1 hex head key 5 mm 1 wrench 12 mm Remove the side and rear panels (see above procedure 3.1.1)
  • Page 48: Eplacement Of The Heating Fuse

    NORMALAB France 2012 Coupling transformer 3.7.2 Replacement of the heating fuse Remove the side and rear panel (see above procedure 3.1.1) N O R M A L A B F r a n c e 48/118...
  • Page 49: Eplacement Of The Circuit Breaker

    NORMALAB France 2012 3.7.3 Replacement of the circuit breaker Tools 1 screwdriver 2.5 mm 1 iron solder 1 cutting pliers 1 wrench 7 mm 1 wrench 10 mm Remove the left upper panel (see above procedure 3.1.3) Remove the circuit breaker holder...
  • Page 50: Replacement Of The Cl3 Board

    NORMALAB France 2012 3.7.4 Replacement of the CL3 board Remove the side and rear panels (see above procedure 3.1.1) Tools 1 hex head key 5.5 mm 1 tube key 5.5 mm 1 wrench 12 mm Remove the fixing supply frame...
  • Page 51: Eplacement Of The Ambient Probe

    NORMALAB France 2012 3.7.5 Replacement of the ambient probe Remove the side and rear panels (see above procedure 3.1.1) Tools 1 hex head key 2.5 mm 1 tube key 5.5 mm The probe is fixed with 2 collars Unplug the wires on 10-11 and 12-13 of J3 The check of calibration may be necessary 3.8 Electronic drawer...
  • Page 52: He Barometric Sensor

    NORMALAB France 2012 3.8.2 The barometric sensor Remove the left upper panel (see above procedure 3.1.3) Tools 1 hex head key 4 mm 1 tube key 5.5 mm 1 screwdriver 3.5 mm Remove the connections and disconnect the The barometric sensor is welded on SMD , see...
  • Page 53: Eplacement Process Of The Microprocessor Board

    NORMALAB France 2012 3.8.3 Replacement process of the microprocessor board Tools 1 hex head key 2.5 mm 1 tube key 5.5 mm Remove the left upper panel (see above procedure 3.1.3 The CPU is mounted in mezzanine on the CL1...
  • Page 54: Front Panel

    NORMALAB France 2012 3.9 Front panel 3.9.1 Replacement of MMI microprocessor board Remove the side and rear panels (see above procedure 3.1.1) Remove the front panel (see above procedure 3.1.2) Tools 1 screwdriver 2.5 mm 1 tube key 5.5 mm Remove the connections and remove the CPU 3.9.2 Replacement of the hard drive disk...
  • Page 55: Replacement Of The Lcd Display

    NORMALAB France 2012 3.9.3 Replacement of the LCD display Remove the side and rear panels (see above procedure 3.1.1) Remove the front panel (see above procedure 3.1.2) Tools 1 hex head key 2.5 mm Remove the 2 fixing screws Release the cable...
  • Page 56: Eplacement Of Inverter Backlight

    NORMALAB France 2012 3.9.4 Replacement of inverter backlight Remove the side and rear panels (see above procedure 3.1.1) Remove the front panel (see above procedure 3.1.2) Tools 1 screwdriver 2.5 mm 1 tube key 5.5 mm Remove the connection and remove the inverter 3.9.5 Replace the touch screen controller...
  • Page 57: Calibration And Maintenance Functions

    NORMALAB France 2012 4 CALIBRATION AND MAINTENANCE FUNCTIONS 4.1 Calibration functions Enter the password. NB: on delivery of the instrument, the password is 0000 There are 3 options for calibrating the instrument: Calibrating the complete measuring system (probe and electronics) Simulating the probe using a decade resistor or a calibrator (this method allows calibrating exclusively the electronics, the probe error is thus not taken into account).
  • Page 58 NORMALAB France 2012 Calibration of the complete system is the most accurate system since it incorporates the probe error in the calibration to offset it. As a result, the probe must be mobile and two calibration baths stabilised at precisely known temperatures are required. This method is reserved for the sample probe (the other probes internal to the instrument are not mobile and their calibration does not require such accuracy).
  • Page 59: Options And Equipment Required

    NORMALAB France 2012 4.2 Options and equipment required Sensor to be calibrated simulation bath point Bath + extension lead Life book with Simulator + cable reference 60525 + Equipment necessary previous values reference XXX and the sensor to be calibrated...
  • Page 60: Xample Of Creation Of A New Calibration

    NORMALAB France 2012 For each calibration, it is possible to: Creation: Create a new calibration (case of a new probe). Modification: Modify a calibration (in the event of a deviation or erroneous measurements). Delete: Delete a calibration (if the probe is destroyed or calibration is false).
  • Page 61 NORMALAB France 2012 Depending on the equipment available, there are two possibilities: one or two baths stabilised at temperature are available and it is possible to carry out calibration with the probe. One Pt100 probe simulator (decade resistor or calibrator): the probe will be replaced by the simulator.
  • Page 62 NORMALAB France 2012 If preselections are not appropriate, reply ‘yes’ and use the keyboard on screen to enter the new point values. End each input session with a green tick. NB: to make a calibration with sufficient accuracy, the system imposes a difference of 80°C between the two...
  • Page 63 NORMALAB France 2012 NB: If the signal is too unstable, the instrument will not achieve stability and remains in this function. It is possible to force it by pressing the green tick (but the calibration will not be accurate given the instability).
  • Page 64: Ommissioning A New Sample Temperature Probe

    NORMALAB France 2012 4.3.2 Commissioning a new sample temperature probe When commissioning a new sample temperature probe, it is necessary to verify the position of the ring which detects correct positioning of the probe on the flask.  Place the centering stopper on the probe and place the whole assembly on the neck of the flask (1).
  • Page 65: Temperature Measurements By Thermocouple

    NORMALAB France 2012 4.5 Temperature measurements by thermocouple These measurements refer to: The dry point detection thermocouple The heating element temperature measuring thermocouple The calibration of these sensors can only be performed in simulation mode using the calibration reference 60523.
  • Page 66 NORMALAB France 2012 NB: If it is not possible to carry out calibration of the barometric sensor for lack of measuring equipment, it is possible to set the instrument in manual configuration to correct the atmospheric pressure. As a result,...
  • Page 67: Calibration Of Test Tubes

    NORMALAB France 2012 4.7 Calibration of test tubes In order to measure the exact volume, calibration of test tubes used on the instrument is recommended. As for the sample temperature probes, it is possible to store several calibrations (maximum 1000).
  • Page 68: Calibration Of The Tactile Pad

    NORMALAB France 2012 In order to ensure the same type of meniscus during calibration and measuring, it is recommended carrying out calibration with the same type of liquid as the samples to be analysed. Meniscus Procedure: Equipment necessary: The test tube or test tubes to be calibrated.
  • Page 69: Denomination And Functions Of The Electronic Boards

    NORMALAB France 2012 Its purpose is to monitor each magnitude to be calibrated in the instrument and expiry of the calibration period. It also allows, if the deadline is exceeded, prohibiting starting a test if the option is selected. To choose this option, select the icon on each screen of a magnitude to be measured (after completing the field “modified by”...
  • Page 70: Cl2 Board

    NORMALAB France 2012 PT 100 probes o Thermocouples o Drip deflector sensor o Liquid detector of expansion tank o Door sensor of heating block o Door sensor of compartment o Limit switches of level follower o Optical sensor ( first drop and follower)
  • Page 71: Stepper Board

    NORMALAB France 2012 5.8 Stepper board: Drive the stepper motor power 6 Schematic electrical and electronic 6.1 Interconnection scheme N O R M A L A B F r a n c e 71/118...
  • Page 72 NORMALAB France 2012 N O R M A L A B F r a n c e 72/118...
  • Page 73: Spares Part List

    NORMALAB France 2012 6.2 Spares part list 6.2.1 Probes and thermocouples 6.2.2 First drop and follower optical sensors N O R M A L A B F r a n c e 73/118...
  • Page 74: Scheme Of Electronic Boards

    NORMALAB France 2012 6.3 Scheme of electronic boards 6.3.1 CL1 board N O R M A L A B F r a n c e 74/118...
  • Page 75 NORMALAB France 2012 N O R M A L A B F r a n c e 75/118...
  • Page 76: Cl2 Board

    NORMALAB France 2012 6.3.2 CL2 board N O R M A L A B F r a n c e 76/118...
  • Page 77: Cl3 Board

    NORMALAB France 2012 6.3.3 CL3 Board N O R M A L A B F r a n c e 77/118...
  • Page 78: Cl4 Board

    NORMALAB France 2012 6.3.4 CL4 board N O R M A L A B F r a n c e 78/118...
  • Page 79: Cl56 Board

    NORMALAB France 2012 6.3.5 CL56 board N O R M A L A B F r a n c e 79/118...
  • Page 80: Stepper Motor Board

    NORMALAB France 2012 6.3.6 Stepper motor board 6.3.7 Detector probe and door closed detector sensor N O R M A L A B F r a n c e 80/118...
  • Page 81: Front Panel

    NORMALAB France 2012 6.3.8 Front panel N O R M A L A B F r a n c e 81/118...
  • Page 82 NORMALAB France 2012 N O R M A L A B F r a n c e 82/118...
  • Page 83: Spares Part List

    NORMALAB France 2012 7 Spares part list 26111 100 mL glass test tube with base 60516 100 mL anti-mist glass test tube with base 60512 Diaphragm 60514 Bag of pumice stone 60513 Condenser cleaning cable 12609EC 5mL graduated cylinder for residue...
  • Page 84 NORMALAB France 2012 40081 Grip for centering plug for 200 mL flask Stopper for connecting condenser tube and 60515 flask 60510 PT 100 sample probe without position ring 60552 Position ring for Pt 100 sample probe 60509 Thermocouple for dry point detection...
  • Page 85 NORMALAB France 2012 60505 Heating block plate with hole diameter 25 mm Kit of 10x 'O' ring seals with centering stopper 40040 for 125 mL flask Kit of 10 'O' ring seals with centering stopper 40042 for 200 mL 60520...
  • Page 86 NORMALAB France 2012 Extinguishing system nitrogen connection 60528 tube 60553 Resistor box to simulate Pt100 probes 60554 1st drop detection fork 60555 Level follower fork 40012 Coolant liquid (2 liter bottle) 60556 CL1interface card 60557 UC IHM card 60546 CL2 card...
  • Page 87 NORMALAB France 2012 60559 Electronic cards power supply 60560 Back-light card Peltier element S/E regulation control card 60561 (compartment or condenser) 60562 Sample probe presence sensor card 60563 Step motor card 60564 Heating block high position sensor card 60550 230 V color inkjet printer...
  • Page 88 NORMALAB France 2012 60569 Compartment Peltier S/A fan (without filters) 60570 Dust filter for fan Peltier S/A Sensor detection of level in fluid circuit coolant 60571 liquid 60572 Drip deflector detection sensor 60573 Compartment door closed detection sensor 60574 Sensor detecting follower high or low position...
  • Page 89 NORMALAB France 2012 High test tube retention system with detection 60581 of flow finger 60582 12 V DC relay 60583 230 V AC relay 60584 24 V DC relay 60585 125 V AC relay 60586 Main circuit breaker 60587 800 VA 25 V transformer...
  • Page 90 NORMALAB France 2012 60593 LCD color screen Condenser temperature regulation sub- 60594 assembly Test tube compartment temperature 60595 regulation sub-assembly Peltier element S/A power supply 60596 (compartment or condenser) 60597 Hard disk with Windows license 60598 Mains on/off switch 60599...
  • Page 91 NORMALAB France 2012 60605 Latch for locking heating block door 60606 Circulation pump for coolant circuit 60607 Coolant circuit drain valve Receiver compartment door hinge with spring 60608 (pair) 60609 Latch for locking test tube compartment door 60610 Receiver compartment door...
  • Page 92 NORMALAB France 2012 60617 Translator for test tube compartment follower 60529 4 port 2.0 USB hub Nitrogen electro valve kit for test tube 60530 none compartment 60531 Extinguishing system nitrogen bottle 60618 30 A 230 V power relay 60619 30 A 115 V power relay...
  • Page 93: Troubles Shooting

    NORMALAB France 2012 8 Troubles shooting Degraded Description Solution or check to perform function The circuit breaker does Verify that the mains voltage is in not remain engaged. concordance with that displayed on the instrument. The circuit breaker is Check that the transformer of the flask tripped when a test is heating circuit.
  • Page 94 NORMALAB France 2012 start or takes a of compartment or power supplies are correctly positioned. Verify that the switch for each internal power supply is at ‘on’. Verify that the fan filters are not clogged by dust. Verify that the temperature exchanges of...
  • Page 95 NORMALAB France 2012 exempt from vibration (which can generate surface waves on the liquid and interfere with the follower). Verify the calibration of the plate thermocouple. Verify the mechanical position of the plate thermocouple. Product file ill-adapted. Flow rate value not Poor choice of standard.
  • Page 96 NORMALAB France 2012 Verify that the modification of final heating is not too great in relation to the declared Final boiling point No detection or detection threshold. detection (FBP) divergent value Verify declared protection threshold. Verify the calibration of the test tube and its declaration.
  • Page 97: Annexe

    What's new.pdf 9.5 How to get a new version The recommended (and easiest) way to get NDI450 new software is to download on Normalab Website. Download and save the appropriate files for your instrument from the Normalab NDI450 Download Center: http://www.normalab.com/software/NDI450/Software/Update.zip...
  • Page 98: How To Know The Software Version

    Be sure your new version is compatible with your instrument. Please contact us for confirmation. An incorrect installation can involve a dysfunction of the unit, company NORMALAB France SAS could not be responsible for incorrect installation of software or firmware done by your side on this equipment.
  • Page 99 NORMALAB France 2012 Open the drive letter assigned to the USB flash drive by looking at the list of drives in window explorer. Create a new folder named “Update” (if not exist). Extract and copy the NDI450 new version into the Update folder.
  • Page 100: Install A New Mmi Version Using The Update Option (Video Example)

    NORMALAB France 2012 9.10 Install a new MMI version using the Update option (video example) After NDI450 software has started, check the version of the MMI software before starting an update. Plug-in the USB flash drive to your instrument. Software should detect automatically your USB flash drive.
  • Page 101 NORMALAB France 2012 Select the new MMI software “MMI_X_ XX” and click “Install update”. N O R M A L A B F r a n c e 101/118...
  • Page 102 NORMALAB France 2012 When you see "Welcome to the setup wizard", check the new version and click "next". If you abort with “Cancel” button, the NDI450 software does not restart and it will be necessary to reboot your instrument with the main ON/OFF switch.
  • Page 103: Install A New Firmware Version Using The Update Option (Video Example)

    NORMALAB France 2012 At the end of the wizard, check that "Start the program" is checked. If not checked, the NDI450 software does not restart and it will be necessary to reboot your instrument with the main ON/OFF switch. Click “Done” to close the wizard.
  • Page 104 NORMALAB France 2012 Plug-in the USB flash drive to your instrument. Software should detect automatically your USB flash drive. On the “Software setup” page, click “Software update”. On the “Software update” page, click “Load the updates list”. N O R M A L A B F r a n c e...
  • Page 105 NORMALAB France 2012 Select the new firmware “FIR_Y_ YY” and click “Install update”. The new firmware will be installed automatically. N O R M A L A B F r a n c e 105/118...
  • Page 106: Error Messages

    NORMALAB France 2012 After the update, microprocessor card will restart. Thank you for your patience (around 70 seconds). When the firmware is restarted, click “Ok”. Check the new version and click "Ok". The NDI450 software will restart automatically. 9.12 Error messages Cannot update firmware, configuration file not found: The settings file is not found.
  • Page 107: Annexe

    NORMALAB France 2012 unit. If this cable is not present, please shutdown the unit and connect the Ethernet cable between the "Network" and "Update". Cannot update firmware, network communication problem (0x02): The update file cannot be copied to the microprocessor card. Please contact technical services.
  • Page 108 NORMALAB France 2012 N O R M A L A B F r a n c e 108/118...
  • Page 109 NORMALAB France 2012 N O R M A L A B F r a n c e 109/118...
  • Page 110 NORMALAB France 2012 N O R M A L A B F r a n c e 110/118...
  • Page 111: Alarm List

    NORMALAB France 2012 11 Alarm list N O R M A L A B F r a n c e 111/118...
  • Page 112 NORMALAB France 2012 Alarm Action level 1 Action level 2 Action level 3 Drip-deflector drip deflector must be correctly missing : check presence of drip deflector adjusted contact technical services Receiver chamber door open: check if door is correctly closed...
  • Page 113 NORMALAB France 2012 Alarm Action level 1 Action level 2 Action level 3 problem : Dry point probe problem : check thermocouple connection check thermocouple contact technical services Flask heater area check if the heating block door is correctly locked in...
  • Page 114 NORMALAB France 2012 Alarm Action level 1 Action level 2 Action level 3 First drip of condensate false condensation may affect the photo detection : Condenser not cleaned up or water condensation detectors contact technical services Level follower fork problem :...
  • Page 115 NORMALAB France 2012 Alarm Action level 1 Action level 2 Action level 3 Receiver chamber control condenser temperature out of range (<-5°C or >70°C problem : ) check condenser temperature check probe positioning contact technical services Power supply problem :...
  • Page 116 NORMALAB France 2012 Alarm Action level 1 Action level 2 Action level 3 Check instrument (after reported see instructions and follow the fire) if real fire, check distillation flask (crack or warping) check heating block and tray checkup procedure N O R M A L A B F r a n c e...
  • Page 117: Notes

    NORMALAB France 2012 12 Notes : N O R M A L A B F r a n c e 117/118...
  • Page 118 NORMALAB France 2012 N O R M A L A B F r a n c e 118/118...