Page 8
Giatec in writing. THE WARRANTY SET FORTH IS EXCLUSIVE AND NO OTHER WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED. GIATEC SCIENTIFIC INC. SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
S A F E T Y I N S T RU C T I O N S This chapter contains important safety instructions that the user must follow to operate and store the Giatec iCOR device. Read the following safety ®...
PERATION WARNING • Power Input: USB only. • Only use the provided USB charging cable for iCOR®, and the provided power adapter for the tablet. • Do not perform measurements at the circuits directly connected to Mains (Live circuit). •...
Disconnect the device from the cable after testing (in case of half- cell measurement). • Do not use chemicals or cleaners containing harsh products such as benzene, toluene, xylene, and acetone. • Do not submerge the iCOR device or the operating tablet ® PERATION NVIRONMENT ATTENTION •...
® the corrosion rate of rebar through a non-invasive approach. This means that the need to connect the measuring unit to the rebar (which is the case for other commercial devices) is eliminated in iCOR . Fig. 1 schematically ®...
CELL OTENTIAL EASUREMENT iCOR® is also equipped with a reference electrode for half-cell corrosion measurement (Fig. 6). The purpose of the half-cell potential measurement is to determine the electrical potential on the surface of concrete which is an indication of corrosion potential of rebar in concrete.
This section describes how to setup and operate the iCOR device for ® measuring the corrosion rate and corrosion potential of steel reinforcement in concrete structures. Before starting a test, please make sure that the iCOR ® device and your data recording unit are fully charged. iCOR ®...
Page 21
User Manual ® Figures 8 to 10 show the components of the Giatec iCOR measurement ® device. Figure 8. Giatec iCOR - Top view ® Notification LED ON/OFF button Handle Figure 9. Giatec iCOR - Side view ® Charging/Half-Cell connection terminal...
Page 22
-0.65 mV per °C for the temperature range between 0 to 50 °C. If the temperature compensation mode is selected in the Giatec iCOR® software, the results are modified and presented for 25°C. The Giatec iCOR® half-cell potential measurement unit is equipped with a maintenance-free electrode.
Figure 11. Giatec iCOR – Bottom view Before performing a test, make sure that the sponges at the tip of the electrodes as well as that on the half-cell electrode are completely moist.
User Manual ® iCOR ® URNING N THE Press and hold the main power key of the iCOR device for 2 seconds. This ® will turn on the device. Push the button for 2 seconds Figure 12. Turning on the iCOR ®...
User Manual ® a fully charged battery. The LED will be an orange color when the iCOR ® unit is being charged. The device will not turn on when connected to the charger. Slow Flashing Mode When the LED is flashing a red color, it represents the “advertising mode”.
User Manual ® Push the charging socket firmly until it clicks. Figure 13. Charging Giatec iCOR ® device. URNING EVICE The device turns off by pressing the ON/OFF button for 5 seconds. If the device remains inactive for 10 minutes, it will automatically power off. To turn the device back ON, press and hold the ON/OFF for 2 seconds.
ABLET OFTWARE The data recording unit for the iCOR device consists of a tablet with a pre- ® installed app. The user would have a chance to install the latest version of the iCOR device as soon as it becomes available online.
Page 28
User Manual ® Step 1: From the home screen on the tablet, tap on the “Giatec iCOR ” icon. ® This will launch the iCOR software. ® Figure 15. Launching the software The main “Control” menu items are located at the toolbar at the top of the screen.
Page 31
Bluetooth communication on the device. Step 7: A list of available device(s) will then appear on the screen (Fig 20). In the search results, select the iCOR . This will connect the device to the ®...
Page 33
When the tablet is connected to the device, the device icon on the toolbar will change to a green color. This indicates that the device and tablet are paired and ready for testing. The battery indicator will also turn on showing the charging level of the battery in the iCOR device. ®...
Page 34
® Figure 22. Device connected Step 8: In order to conduct the measurements on a specific location on the grid, simply move the Giatec iCOR device to the corresponding grid point ® on the concrete surface (Fig. 23(a)). Press and hold the corresponding grid point on the tablet screen (Fig.
Page 36
After the test parameters are all set, press ► to start the measurements. The commands will be transmitted to the iCOR device and the device will ® perform the test. The results will be sent to the tablet after the measurements are taken, analyzed and displayed on the tablet screen.
Page 37
User Manual ® Figure 24. Test setup parameters Figure 25. iCOR device taking a measurement ® The test results will be identified with different color labels to help the user interpret the results. In Tables 4 to 6, definitions of each color codes are provided.
Table 7, 8 and 9. Table 7: Error 1 ERROR 1 This error indicates that the iCOR cannot polarize the rebar. It will show ® either a flat or decreasing voltage measurement curve.
® (tablet) before conducting measurements in the field. Refer to the iCOR Operation Section of this user manual to learn how to charge the device. Electrodes positioning: The location of the corrosion measurement electrodes affects the measurement of corrosion rate.
If the iCOR device does not turn on, check the LED on the device for the ® battery status and charge the device if required. If iCOR still does not work ® properly; review this user manual to make sure you are operating it correctly.
Operating temperature 0 ~ 45 °C Operating humidity 20 ~ 90% Storage temperature 0 ~ 50°C Storage humidity 5 ~ 95% Weight of iCOR Device 1 kg ® Dimensions of iCOR Device 184 mm x 116.5 mm (D x H) ®...
TERMS AND CONDITIONS OF SALE ("SALE AGREEMENT") Please read this Sale Agreement before using the Giatec Scientific Inc. (“Giatec”) product (the “Product”). By finalizing Your purchase order, You, the purchaser of the Product, and, if applicable, any end user (“End User”) on whose behalf You are making this purchase for (You and the End User are hereafter collectively referred to as “Customer”;...
Need help?
Do you have a question about the iCOR and is the answer not in the manual?
Questions and answers