DAVI & CIA MA08-02/B User Manual

DAVI & CIA MA08-02/B User Manual

Medical scale

Advertisement

USER'S MANUAL
MA08-02/B
15/09/2011
www.davicia.com

Advertisement

Table of Contents
loading

Summary of Contents for DAVI & CIA MA08-02/B

  • Page 1 USER’S MANUAL MA08-02/B 15/09/2011 www.davicia.com...
  • Page 2: Table Of Contents

    USER’S MANUAL MA08-02/B INDEX Page 1. 1. 1. 1. DESCRIP DESCRIPTIO TION N N N DESCRIP DESCRIP 2. 2. 2. 2. INSTAL INSTAL INSTALLATION INSTAL LATION LATION LATION 2.1. Unpacking 2.2. Scale assembly 2.3. Location and setting 2.4. Startup 3. 3. 3. 3. OPERATION...
  • Page 3 USER’S MANUAL MA08-02/B 1. DESCRIPTION (1) Height meter sensor (2) Height meter arm (3) LCD display (4) Weighing platform (5) Adjustable feet (6) Height meter arm fixing (7) Connector panel (8) Power supply connection (9) ON/OFF switch (10) Auxiliar connection...
  • Page 4: Unpacking

    USER’S MANUAL MA08-02/B 2. INSTALLATION MEDICAL SCALE scale comes factory configured according to the voltage of each country and is ready for operation immediately after unpacking. Connect the device to the power supply via the power supply cable. It is necessary that the outlet is provided with grounding.
  • Page 5: Scale Assembly

    USER’S MANUAL MA08-02/B The attached box (06) contains: User manual. Power source with mains lead. 4 Allen screws. Allen key. Metal cover. 4 metal cover screws. 2 plastic plugs. Cable grip strip. 2 rolls of paper. (For model with printer only) 2.
  • Page 6 USER’S MANUAL MA08-02/B...
  • Page 7 USER’S MANUAL MA08-02/B If the equipment is not not equipped with the height measure arm: If the equipment is not equipped with the The console has to be properly held in the rear height measurement arm, move the display to position.
  • Page 8: Location And Setting

    USER’S MANUAL MA08-02/B If the equipment is is is is equipped with the height measurement arm: Remove the height measurement arm from the packaging Contains:4 Allen screws (2 for each arm), 1 Allen key. Connect the connector to the CPU board as shown in the diagram.
  • Page 9: Startup

    USER’S MANUAL MA08-02/B 2.2. STARTUP 2.2. STARTUP 2.2. STARTUP 2.2. STARTUP Connect the equipment to the power supply. Check that no objects are placed on the platform of the equipment. Actuate the switch on the bottom. At this point the equipment will perform the process of self-test.
  • Page 10: Keyboard Functions

    USER’S MANUAL MA08-02/B 3. OPERATION After the equipment startup appears the home screen. This means that the scale has done the calibration This means that the scale has done the calibration This means that the scale has done the calibration This means that the scale has done the calibration successfully and is ready for use.
  • Page 11: Taring

    USER’S MANUAL MA08-02/B IMPORTANT: During the measurements it’s recommended to maintain an upright posture and not moving. Failure to follow directions, the result of the measurement may be altered. At the end of the measurements appear on the screen the values obtained.
  • Page 12: Printed Results

    USER’S MANUAL MA08-02/B the patient's height and then confirm by pressing <OK>. The equipment will automatically display the appropriate BMI. IMPORTANT: BMI is a valid reference for adults only. Evaluating the Body Mass Index Classification of the BMI according to the WHO: <...
  • Page 13: Programming

    USER’S MANUAL MA08-02/B 4. SETUP The pre-configured settings can be modified and functions activated via the menu in order to adapt them to suit specific needs. To access the setup menu of the equipment, follow the procedure below: Switching off the equipment.
  • Page 14: Bmi Programming

    USER’S MANUAL MA08-02/B 4. 4. 4. 4.1 1 1 1 .1 .1 .1 .1..BMI programming BMI programming BMI programming BMI programming This option turns the BMI function on and off. When the BMI function is on, the scale automatically calculates the BMI.
  • Page 15: Printout Mode

    USER’S MANUAL MA08-02/B 4 4 4 4 .1 .1 .1 .1.5 .5 .5 .5..Printout mode Printout mode Printout mode Printout mode This option allows you to select the printout mode: automatic or manual. Select the PRINTOUT...
  • Page 16: Coin Operated

    USER’S MANUAL MA08-02/B Deletes character, back Up arrow key: raises the value. 4. 4. 4. 4.1 1 1 1 .7 .7 .7 .7..Coin operated Coin operated Coin operated Coin operated This option allows you to programme the number of coins to indicate the price of the service if the equipment has a coin selector.
  • Page 17: Adjust Height

    USER’S MANUAL MA08-02/B STANDARD MASS PRESS <OK> Now place a standard mass on the platform. We recommend using a mass of at least 20 kg to ensure a correct adjustment. Then press <OK>. STANDARD MASS 000000 g Use the keypad to enter the value (in grams) of the standard mass you have placed on the platform. For example, 20 kg.
  • Page 18 USER’S MANUAL MA08-02/B 1,85 m 6’0.7” The equipment is now automatically measuring height. Place an object of known height and check that the measurement given by the equipment is correct. Press <OK> to exit. 4. 4. 4. 4.3 3 3 3 .3 .3 .3 .3..Test...
  • Page 19: Paper Roll Change

    USER’S MANUAL MA08-02/B 5. MAINTENANCE AND CLEANING The only maintenance that can be performed by unqualified personnel is the change of paper roll. Do not perform repairs on your own. In case of breakdown contact the Technical Support Service. If the power cable is damaged, it must be replaced by the Technical Support Service or qualified personnel in order to avoid dangerous malfunctioning.
  • Page 20: Equipment Disposal

    USER’S MANUAL MA08-02/B 5. 5. 5. 5.2 2 2 2 . EQUIPMENT DISPOSAL . EQUIPMENT DISPOSAL . EQUIPMENT DISPOSAL . EQUIPMENT DISPOSAL All electrical and electronic products should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities.
  • Page 21 USER’S MANUAL MA08-02/B 6. LISTADO DE ERRORES Error Indicator Error Indicator Error Indicator Error Indicator Cause Cause Cause Cause Correction Correction Correction Correction WEIGHT ERROR Error weight reading Failure in load cell CODE 01 Failure in A/D Failure in load cell wires...
  • Page 22 USER’S MANUAL MA08-02/B 7. TECHNICAL SPECIFICATIONS Model: MEDICAL SCALE Manufacturer: DAVI & CIA C/Murcia 35 nave G E08830 Sant Boi de Llobregat Barcelona – SPAIN Display: LCD alphanumeric 2 x 16 characters Weight (with / without height meter): 33 kg / 36 kg (aprox.)
  • Page 23 USER’S MANUAL MA08-02/B INMUNIDAD ELECTROMAGNÉTICA / EM Immunity. -EN 61000-4-2 (1995) / A1 (1998) / A2 (2001): Descarga electrostática / ESD; -EN 61000-4-3 (2006): Campo radiado EM de RF / EM radiated field of RF; -EN 61000-4-4 (2004); Ráfagas de transitorios rápidos / EFT burst;...
  • Page 24 USER’S MANUAL MA08-02/B 8. WARRANTY CERTIFICATE We provide a 1 1 1 1 - - - - year year guarantee from the date of purchase. The guarantee covers manufacturing defects or year year defects occurring as a result of normal use and operation.
  • Page 25 USER’S MANUAL MA08-02/B 9. CE DECLARATION OF CONFORMITY FABRICANTE PRODUCTO: DAVI & CIA S.L. PRODUCT MANUFACTURER: DIRECCIÓN: C/ Murcia, 35, Nave G ADDRESS Polígono Industrial Can Calderón E08830 Sant Boi de Llobregat BARCELONA - ESPAÑA DECLARAN BAJO SU RESPONSABILIDAD QUE EL PRODUCTO:...
  • Page 26 USER’S MANUAL MA08-02/B This manual and other support documents are available on the web www.davicia.com www.davicia.com www.davicia.com www.davicia.com C/ Murcia, 35 Nave G Pol.industrial Can Calderón. Tel. +34 93 654 32 32 info@davicia.com E 08830 Sant Boi de Llobregat. Barcelona. Spain Fax.