the procedure. When successful, reconnect the cable connector to J4 on the
ADP-232 unit. Reference Figure 17
Figure 17
Reset the Microscan 610 through the ADP-232 J4 connector
4 Reconnect the CPU Interface cable (E7200-61709) to the ADP-232.
Reattach the cable to the ADP-2323 unit to continue with the process.
5 Configure the scanner. Configuring and testing the scanner must be
executed from the AGILENT 5DX User Interface command window. From
the 5DX User Interface window select
The multidrop configuration file should be modified for the Microscan 600
series scanner and the Series 5000 concentrator. Configurable parameters
are located in: C:\5DX\Release Root\CALIB\SCNR.CFG and is show in
Figure 18
Figure 18
Configurable parameters in SCNR.CFG
6 Go back to Master folder. At the Dos prompt, type "cfgscnr 1". This
will program the first BCR read head. When CFGSCNR is executed, the
current configuration settings will be displayed to inform the user how the
scanner is being programmed. You must hear the BCR Read Head beep.
SAVE the configuration. Figure 19 shows the usage and results.
Automatic Barcode Reader
System > Sytem Access > Quit 5DX.
15