Operating Manual Ultrasonic Thickness Gauge DM4E, DM4 Models DM4 DL with On-board data Logger Manual Part No. 021-002-073 REV. S Copyright 2002 Agfa NDT Inc.
Areas covered by the sound instrument is calibrated is the sound beam. velocity of the material being tested. Operators must be familiar with the use Actual sound velocities in materials of ultrasonic couplants. Testing skills Page 1 DM4E, DM4, DM4 DL Operating Manual...
Page 5
This is particularly important when the test piece is being ultrasonically measured for the first time or in any case where the history of thickness of the test specimen is unknown. Page 2 DM4E, DM4, DM4 DL Operating Manual...
However, should service become necessary, Agfa NDT Inc. Krautkramer DM4E, DM4, and DM4 DL test instruments has established a number of Factory Trained Service Centers. are conditionally guaranteed to be free from defects in...
Automatic probe zero for fast calibration • Automatically recognizes DIALOG Intelligent Transducers. This manual describes the features and operation of the DM4E, DM4, and DM4 DL (with on-board data logger) Ultrasonic Thick- • Automatic V-path error correction compensates for non- ness Gauges.
Page 9
The result is the thickness of the material. The figure at right illustrates the pulse-echo principle of ultrasonic thickness measurement. Dual Element Thickness Measurement Page 6 DM4E, DM4, DM4 DL Operating Manual...
Note: To charge the unit of measure and display resolution refer to Section 2.14. 2.2 Displays Symbols Thickness measurement and standard 1-point calibration mode. Dual Multi “Through Coating (paint)” SPEC measurement mode. Page 7 DM4E, DM4, DM4 DL Operating Manual...
NOTE: The scrolling keys are also used to change the resolution and unit of measure by pressing both keys simultaneously in THK mode (CAL indicator must be off) Page 8 DM4E, DM4, DM4 DL Operating Manual...
(Section 2.5) or 2-point calibration (Section 2.6). will resume at the same speed. Figure 2-4 shows how the display should appear with a 0.200 inch calibration standard. NOTE: DM4E, DM4 and DM4 DL feature automatic probe zero for standard probes and automatic probe recognition for Press again to end the calibration procedure.
) lights and the reading is stable. You may keep the probe coupled or uncouple and wipe the excess couplant from the probe face. to adjust the displayed value to match the thickness of the thin calibration standard. Page 10 DM4E, DM4, DM4 DL Operating Manual...
NOTE: This information is provided for the convenience of the user and Agfa NDT Inc. assumes no responsibility for inaccura- cies. Actual velocities depend on exact composition, temperature, and processing of each material. Page 11 DM4E, DM4, DM4 DL Operating Manual DM4E, DM4, DM4 DL Operating Manual...
Automatic gain High gain Figure 2-13 Medium gain Low gain 2.9 Minimum Capture Measurement to select a new setting. To enable the minimum capture mode, press until MIN illuminates. Page 12 DM4E, DM4, DM4 DL Operating Manual...
Use to set the limit (0.020 to 20.00"; Figure 2-18 0.5 to 500 mm). Press again to return to display. Press to return to THK mode. Page 13 DM4E, DM4, DM4 DL Operating Manual...
3 seconds.. When the instrument shuts off, the resolution setting is saved and restored when the unit is turned on. Page 14 DM4E, DM4, DM4 DL Operating Manual...
. The location where the reading has been stored is displayed briefly (Figure 2-20). After a displayed reading has been sent, the SEND key is disabled until a new reading is taken. Zero values cannot be recorded. Figure 2-21 Page 15 DM4E, DM4, DM4 DL Operating Manual...
This means that the acoustic zero point is dynamically adjusted during the measurement process, compensating for wear and “growth” during measurements. This mode may offer advantages for certain applications such as high temperature measurements. Page 16 DM4E, DM4, DM4 DL Operating Manual...
PC (Section 4.5). Individual readings and velocity can be configured files, 1 through 806. sent to the Krautkramer DR1 Data Recorder. Refer to Chapter 4 for information on connecting external devices. NOTE: DIALOG Intelligent Transducer I.D., LO limit, HI limit, and dIF value are saved with each file.
Section 3.5. File Full: displays when SEND is attempted to a file with no empty locations. No data is recorded. To select another file, refer to Section 3.3. Figure 3-5 Page 18 DM4E, DM4, DM4 DL Operating Manual...
Press and hold for 3 seconds. DM4 DL briefly displays “C.nnn” (Figure 3-7). “nnn” is the cleared file number. Then “- F.nnn” is displayed (empty file, Figure 3-2 on page 31). DM4E, DM4, DM4 DL Operating Manual Page 19...
To change the baud rate setting, press until SPEC illumi- - Printer cable from Agfa NDT Inc. If necessary, follow the nates. procedure in Section 4.2 to set baud rate to match that of the printer. Page 20 DM4E, DM4, DM4 DL Operating Manual...
Section 4.4. to select another file to report. To abort the transmission, press while readings are being printed. The DM4 DL displays Abor. Figure 4-4 DM4E, DM4, DM4 DL Operating Manual Page 21...
Abor, and return to the cur- rently selected file number. In this case, the entire transmission must be repeated by following the procedure in Section 4.4, if printing data, or Section 4.5 Page 22 DM4E, DM4, DM4 DL Operating Manual...
<1000°F . 12.7mm .75 to 250 mm <538°C NOTE: HT400 actual measurement range depends upon surface condition and couplant. HT400 temperature cycling required with surface temperatures above 400°F per Krautkramer instruction card. Page 24 DM4E, DM4, DM4 DL Operating Manual...
Model condition. The problem can usually be corrected by rotating HT400. A special couplant, Krautkramer ZGM, and duty the probe so that the crosstalk barrier is at a right angle to the cycle are also required for high temperature use. The grooves, or using the Dual Multi measurement mode.
• 30 to 120 seconds for temperatures from 550°F to 1,000°F flaw detection instrument. Before taking another reading, carefully remove the remaining couplant and residue from the probe. DM4E, DM4, DM4 DL Operating Manual Page 27...
To verify that a successful reset has occurred, switch to VEL (velocity) mode and check the displayed velocity value. The factory default value of .2313 in/µs should be displayed. If it is not, repeat the procedure. Page 34 DM4E, DM4, DM4 DL Operating Manual...
Need help?
Do you have a question about the DM4E and is the answer not in the manual?
Questions and answers