BIRD TPM1 Operation Manual

Transmitter power monitor/meter tpm series

Advertisement

Quick Links

Transmitter Power Monitor/Meter
Model TPM series
&
3140A Power Meter Panel
Operation Manual
©Copyright 2016
by Bird Technologies, Inc
Instruction Book Part Number 920-TPM Rev. C
Microsoft and Windows are registered trademarks
of the Microsoft Corporation

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TPM1 and is the answer not in the manual?

Questions and answers

Summary of Contents for BIRD TPM1

  • Page 1 Transmitter Power Monitor/Meter Model TPM series & 3140A Power Meter Panel Operation Manual ©Copyright 2016 by Bird Technologies, Inc Instruction Book Part Number 920-TPM Rev. C Microsoft and Windows are registered trademarks of the Microsoft Corporation...
  • Page 2: Safety Precautions

    Safety Precautions The following are general safety precautions that are not necessarily related to  any specific part or procedure, and do not necessarily appear elsewhere in this  publication. These precautions must be thoroughly understood and apply to all  phases of operation and maintenance. WARNING Keep Away From Live Circuits Operating Personnel must at all times observe general safety precautions. Do  not replace components or make adjustments to the inside of the test  equipment with the high voltage supply turned on. To avoid casualties,  always remove power. WARNING Shock Hazard Do not attempt to remove the RF transmission line while RF power is present. WARNING Do Not Service Or Adjust Alone Under no circumstances should any person reach into an enclosure for the  purpose of service or adjustment of equipment except in the presence of  someone who is capable of rendering aid. WARNING Safety Earth Ground An uninterruptible earth safety ground must be supplied from the main  power source to test instruments. Grounding one conductor of a two  conductor power cable is not sufficient protection. Serious injury or death  can occur if this grounding is not properly supplied. WARNING Resuscitation Personnel working with or near high voltages should be familiar with modern  methods of resuscitation. WARNING Remove Power Observe general safety precautions. Do not open the instrument with the  power on.
  • Page 3: Safety Symbols

    Safety Precautions Safety Symbols WARNING Warning notes call attention to a procedure, which if not correctly  performed, could result in personal injury. CAUTION Caution notes call attention to a procedure, which if not correctly performed,  could result in damage to the instrument. Note: Calls attention to supplemental information. Warning Statements The following safety warnings appear in the text where there is danger to  operating and maintenance personnel, and are repeated here for emphasis. WARNING Disconnect the unit from the RF power source and the ac line before any  disassembly. The potential for electrical shock exists. See page 17...
  • Page 4: Caution Statements

    Transmitter Power Monitor/Meter Caution Statements The following equipment cautions appear in the text and are repeated here for  emphasis. CAUTION Pin 4 must be connected to power supply ground, other grounds can be left  open if desired. See page 3 CAUTION DB9 are not serial connections. Serial connections to them could damage the  equipment. See page 6 CAUTION A Bird Calibration Kit is recommended for this procedure. Performing a  calibration with anything other than a Bird Calibration Kit voids Bird’s  accuracy guarantee. Performing a field calibration will void any remaining warranty on this  product. Bird recommends waiting until the terms of the warranty have  passed before performing a field calibration. See page 7 CAUTION Do not use harsh or abrasive detergents for cleaning. See page 17...
  • Page 5: Safety Statements

    Safety Precautions Safety Statements USAGE ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION. EL USO DE ESTE INSTRUMENTO DE MANERA NO ESPECIFICADA POR EL FABRICANTE, PUEDE ANULAR LA PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO.
  • Page 6 Transmitter Power Monitor/Meter SERVICE SERVICING INSTRUCTIONS ARE FOR USE BY SERVICE - TRAINED PERSONNEL ONLY. TO AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS QUALIFIED TO DO SO. SERVICIO LAS INSTRUCCIONES DE SERVICIO SON PARA USO EXCLUSIVO DEL PERSONAL DE SERVICIO CAPACITADO. PARA EVITAR EL PELIGRO DE DESCARGAS ELÉCTRICAS, NO REALICE NINGÚN SERVICIO A MENOS QUE ESTÉ...
  • Page 7 Safety Precautions RF VOLTAGE MAY BE PRESENT IN RF ELEMENT SOCKET - KEEP ELEMENT IN SOCKET DURING OPERATION. DE LA TENSION H.F. PEAT ÊTRE PRÉSENTE DANS LA PRISE DE L'ÉLÉMENT H.F. - CONSERVER L'ÉLÉMENT DANS LA PRISE LORS DE L'EMPLOI. HF-SPANNUNG KANN IN DER HF-ELEMENT-BUCHSE ANSTEHEN - ELEMENT WÄHREND DES BETRIEBS EINGESTÖPSELT LASSEN.
  • Page 8: About This Manual

    Transmitter Power Monitor/Meter About This Manual This manual covers the operating and maintenance instructions for the  following models: All TPM series 3140A4 3140A8 Changes to this Manual We have made every effort to ensure this manual is accurate. If you discover  any errors, or if you have suggestions for improving this manual, please send  your comments to our Solon, Ohio factory. This manual may be periodically  updated. When inquiring about updates to this manual refer to the part number  and revision on the title page. Terminology There are some unique terms used throughout this literature. They are defined here clarify any misunderstanding. TPM — Refer to the in‐line Transmitter Power Monitor. Display or Panel Meter — Refers to either the 3140A4 or 3140A8 meter.
  • Page 9: Table Of Contents

    Table of Contents Safety Precautions ......... i Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...
  • Page 10: Chapter 1 Introduction

    Chapter 1 Introduction TPM7 3140 Panel TPM1 Calibrated coupling ports and accurate power measurement combined in the  same unit provides for easy use and accurate readings. The TPM is the first inline  power monitor that can be calibrated in‐line and on site minimizing downtime  and optimizing on‐air time. In addition, a simplified interface allows for a high  level of customization and integration. Purpose and Function The TPM is used to measure power signals in broadcast systems. It is used  directly in line with the system being tested. It’s output provides a linear DC  voltage output from 0 to 4 volts allowing for a wide variety of interface options.  It’s in‐line calibration capability allows for greater accuracy in a single  application (with an accurate power reference) and its integrated non‐ directional coupler allows for signal analysis in minimal space requirements. It is  available in 7/8”, 1 5/8”, and 3 1/8”, 50 ohm lines for FM, VHF, and UHF broadcast  frequencies. When combined with a Bird model 3140A4 or 3140A8 Transmit Power Meter  (TPM) display, a complete adjustment system is achieved. The display supplies a  visual indication of Forward and Reflected power being sampled by the inline  TPM. Multiple TPM’s can be selected (only one per channel setting) to give the  operator an overall condition of the transmitter system.
  • Page 11: Chapter 2 Installing & Operating

    Chapter 2 Installing & Operating Instructions - TPM Figure 2 Calibration Cover FWD Port Test To Source Port Port TPM1 Shown Communication Connector Operation LED Calibration Cover To Load Note: Figure shown with terminations assembled. Note: The label on the top of the unit identifies the connectors  and ports. They are defined as follows: Connect to the source of the RF signal (transmitter side) SOURCE Connect to the load of the RF signal (antenna/dummy load) LOAD The forward directional coupler port. Terminate this port in  FORWARD 50 ohms at all times.
  • Page 12: Installing The Tpm

    Transmitter Power Monitor/Meter Installing the TPM The TPM has two RF connectors (Source and Load) that connect to the  transmission line, three test port connectors that are type N female connectors  (Forward, Reflected, and Test), and one nine‐pin sub‐miniature communication  connector (DB‐9 Male). 1. Do the following:  Connect the RF connector labeled Source to the transmitter side of the  transmission line.   Connect the RF connector labeled Load to the antenna or load side of  the transmission line.   Use an appropriate coupling kit to secure the TPM in place. 2. Connect the DB‐9 connector to a single channel on Model 3140A power  meter panel, or to an alternate appropriate power supply and voltmeters  that will indicate the forward and reflected power.  Note: The chart below Figure 2 identifies the signals on each pin  of the DB‐9 connector. Figure 4 DB-9 Connector Pin 4 - Power Supply Ground CAUTION Pin 4 must be connected to power supply ground, other  grounds can be left open if desired. Description Forward Voltage Output, 0 to 5 VDC, 1.0 kohm impedance Ground Reflected Voltage Output, 0 to 5 VDC, 1.0 kohm impedance Power Supply Ground...
  • Page 13: Normal Operation

    Installing & Operating Instructions ‐ TPM Normal Operation Note: Coupler ports must be terminated to ensure accurate power  measurements. Terminate both forward and reflected coupler  ports in 50 ohms. Failure to terminate these ports will result in  inaccurate power measurements. 1. Install the TPM to the transmission line. See  "Installing the TPM" on page  2. Apply DC input power to the TPM. 3. Apply RF power to the TPM. 4. Use a voltmeter to measure the forward voltage and reflected voltage.  Note: These voltages will be between 0 V (no power) and 4.0 V  (full scale power). The forward and reflected voltages are linear. Note: The full scale reflected power is 1/10 the value of the full  scale forward power.
  • Page 14: Chapter 3 Installing & Operating

    Chapter 3 Installing & Operating Instructions - 3140A Panel The 3140A is the standard display and power supply for the TPM. It comes in  both 4 Channel TPM and 8 Channel TPM models. It fits in a standard 2U rack and  requires a single 15V power supply (supplied with panel).  Figure 5 3140 Panel Meter Zero Calibrate Adjustment Switch Meter Adjustment Meter Front View RFL Meter Switch Notation Area Channel Selector Power Note: 3140A8 shown. Button Zero Meter Meter DB9 Connections Calibrate Adjustment...
  • Page 15: Installing The Panel

    Installing & Operating Instructions ‐ 3140A Panel CAUTION DB9 are not serial connections. Serial connections to them  could damage the equipment. FWD/RFL Meter Provides visual indication of measured power for selected TPM. Meter Adj Pot Provides full scale meter calibration. Selects either one of four (3140A4) or one of eight (3140A8)  Channel Selector connected TPM’s. Power Button Applies DC input power to panel meter. Zero Calibrate  Grounds pin 9 to Zero calibrate the selected TPM. Switch Interface connection to TPM’s. Provides power and Zero adjust to  DB9 Connection connected TPM’s. Receives TPM output to drive meters. Installing the Panel 1. Place panel in an empty slot on the rack and attach with mounting screws  (optional). 2. Connect the panel to a power supply. 3. Attach a TPM to one of the available channels on the back of the panel. 4. Label what TPM is attached to which channel in the switch notation area on  the front of the panel. Note: User should note each unit’s full scale power and location in  the notation area. Use a water soluble marker to avoid permanent  entries. 5. Repeat steps 3 through 5 for other TPM’s, if necessary. 6. Turn the panel on. Note: All TPMs are powered, but only one TPM output is dis‐...
  • Page 16: Chapter 4 Calibration

    Chapter 4 Calibration CAUTION A Bird Calibration Kit is recommended for this procedure. Performing a  calibration with anything other than a Bird Calibration Kit voids Bird’s  accuracy guarantee. Performing a field calibration will void any remaining warranty on this  product. Bird recommends waiting until the terms of the warranty have  passed before performing a field calibration. Every calibration procedure will finish by providing a table of voltages that  correspond to power levels at the calibration frequency (f ). It will be up to the  user to use those voltages to generate an equation or fill in a look up table on  whatever they choose to use as an output. Zero Calibrating Note: This procedure can be performed any time the RF is off and  the TPM is on. There is no need to cycle power again on the TPM.  1. Do one of the following:  Locate pin 9 on the TPM’s DB9 connector and short it to ground.  If connected to a display panel, press the zero calibration switch. Refer  to Figure 4 for Zero Calibration Switch location. Note: Zero calibrate will set both the FWD and RFL at the same  time. Figure 7 Pin 9 Pin 9...
  • Page 17: Passive Calibration

    Calibration Passive Calibration Calibrate the TPM on the transmitter it is monitoring. Also a reference power  meter is needed to measure the output of the test port on the TPM (model  7006A250). This calibration can be performed while transmitting on air (forward  only) or into a dummy load. Coupling calibration data for the TPM is also  needed. By using several different power levels, an equation or lookup table can be  generated for the voltage output of the TPM as a function of power level at the  frequency of calibration. 1. Install the TPM to the transmission line. See  "Installing the TPM" on page  2. Remove the load from the Fwd testport.  3. Connect the calibration standard to the Fwd testport. 4. Apply RF power. 5. Record the transmitter frequency displayed and find the listing closest to it  on the TPM’s coupling calibration data card. Note: Make note of the calibration factor. 6. Enter the calibration factor into the calibration standard’s offset function. 7. Measure the power on the calibration standard and voltage on the TPM  output simultaneously. 8. Record the voltage verses power. 9. Turn the unit around and repeat steps 1 to 7 for Reflected calibration.  Note: Reflected full scale is 1/10th of forward full scale.  Note: Calibrating the reflected port is often optional. Check with  the FCC or other ruling body to determine if it is necessary or not.  Figure 8 Calibration Forward Calibration Set‐Up Connector...
  • Page 18: Active Calibration

    Transmitter Power Monitor/Meter Active Calibration The active method will recalibrate the TPM completely. This procedure will void  the calibration and accuracy specifications for the unit. The accuracy will be  based on the accuracy of the reference power meter used for this procedure. In  addition to the equipment needed for the active calibration, a Bird 6A340‐ADJ  Potentiometer Adjustment Tool to adjust the potentiometers on the TPM is also  needed. Set the transmitter’s RF output to a desired power level using the reference  meter on the test port and the test port calibration data. Then, using a Bird  6A340‐ADJ Potentiometer Adjustment Tool, turn the potentiometers until the  desired voltage output is reached. Extra data points can be taken at different  power levels to confirm the linearity and generate a lookup table to drive the  rest of the power levels. The benefit of this calibration method is that the customer can set the output  voltage they want to drive their external meter. With the other calibration  methods, the customer simply uses TPM output voltages as set at the factory. 1. Install the TPM to the transmission line. See  "Installing the TPM" on page  2. Remove the load from the Fwd testport.  3. Connect the calibration standard to the Fwd testport. 4. Apply RF power. 5. Record the transmitter frequency displayed and find the listing closest to it  on the TPM’s calibration data card. Note: Make note of the calibration factor. 6. Enter the calibration factor into the calibration standard’s offset function. 7. Break the seal over the Fwd Cal hole.  8. Adjust the potentiometer manually until the voltage is such that the  calibration standard reads full scale power = 4.00V.  Example: If the power is exactly 50% of full scale, adjust the potentiometer to get ...
  • Page 19 Calibration Figure 9 Calibration Data Card Rfl Coupling Offset at 195 MHz Transmitter Frequency Fwd Coupling Offset at 195 MHz...
  • Page 20: Chapter 5 Performance Specifications

    Chapter 5 Performance Specifications Transmitter Power Meter (TPM) Model Naming Table Line Size and Interface Frequency Power Style Band* Range Sub-Band 7xy = 7/8” with QC connectors,  where x is the input connector  and y is the output connector.  Possible values of x and y are: A =N(F) B = N(M) C = LC(F) D = 7/8” Flange H = 7‐16  V = Very Low J = 7‐16 DIN(M) L = 54 ‐ 88 MHz A = 470 ‐ 554 MHz DIN(F) L = Low F = 88 ‐ 108 MHz B = 554 ‐ 638 MHz K = UHF(F) L = UHF(M) M = Medium H = 174 ‐ 216 MHz C = 638 ‐ 722 MHz...
  • Page 21 10.0 High 25.0 Very High 50.0 RF Specs Reflected full scale power is 1/10 of forward full scale  FWD / RFL Ratio power. Peak/Average Ratio 10 dB min* Main Line Insertion Loss  0.2 dB max, 0.08 dB typical  Main Line VSWR  1.1 max  Testport VSWR  1.5 max, 1.3 typical  ± 5% of reading (2s) after zeroing sensor, if calibrated by  Bird. This does not include user mismatch uncertainty and  Accuracy  user voltmeter uncertainty. If not calibrated by Bird, add  the uncertainty of the calibration method and reference  standard. Derate accuracy by 1% if not at 25 ± 10 °C.  Dynamic Range  16 dB min  Directivity, Rfl  26 dB min, 30 dB typical  Measurement type  In‐line, true‐average power  Unit can be calibrated by user by adjusting pot to get 4V  Calibration  output for a full scale RF input.  Recommended Cal Interval  1 year  Note: Peak to Average ratio is specified based on the full scale  power of the specific TPM model. For example, a TPM with a full  scale forward power rating of 1 kW will provide accurate readings ...
  • Page 22 Transmitter Power Monitor/Meter Interface Specs Main line Diameters and center conductor setbacks per Bird drawings.  Connectors One DB9(M), three N(F)  DB9 connector pinout  Analog output for forward RF power. 0 – 5.0 Vdc proportional to RF  Pin 1 = “FWD” power, 1 kohm  impedance. full scale RF input = 4.0V output.  Analog output for reflected RF power. 0 – 5.0 Vdc proportional to RF  Pin 3 = “RFL” power, 1 kohm  impedance. full scale RF input = 4.0V output.  Pin 5 = “Power” Power supply input. +11 to +18 Vdc, < 0.1 A.  Short this pin to ground to zero out any input offsets.  Pin 9 = “Zeroline” Normally floating. Unit can be zeroed whenever the unit is powered  up and there is no RF power being applied.  Pins 2,4,6–8 = GND Ports are labeled Forward, Test, and Reflected. Forward and  reflected testports include attached loads. These testports must be  terminated with the load, or some other good termination such as a  N(F) Testports power sensor, during normal operation to prevent measurement  errors. Loads can removed from the unit if desired without affecting  calibration.  Forward Testport output is a sample of the forward RF signal.  Test Testport output is a non‐directional sample of the RF.  Reflected Testport output is a sample of the reflected RF signal.  Environmental Specs Temperature, Operating 0 to +50 °C (32 to 122 °F) ...
  • Page 23: Dimensions

    Performance Specifications Dimensions Figure 10 VPM7 Figure 11 Unflanged 1-5/8” TPM Figure 12 Flanged 1-5/8” TPM...
  • Page 24 Transmitter Power Monitor/Meter Figure 13 Unflanged 3-1/8” TPM Figure 14 Flanged 3-1/8” TPM...
  • Page 25: 3140A Meter Panel

    Performance Specifications 3140A Meter Panel General Specs Provide power to multiple TPM units. Display Fwd and Rfl  Function readings from one of the connected TPMs, chosen with a  switch.  Models 3140A4  Supports up to 4 TPMs  3140A8 Supports up to 8 TPMs  Two backlit analog meters. One displays forward power as  measured on the selected TPM, the other displays reflected  Display  power. Which TPM’s readings are displayed on the meters is  controlled by a switch on the front panel.  Units are provided with Bird 5A2436 power supply.  5A2436 requires 115/230 Vac @ 50/60 Hz, < 1.0 A  Power requirements  Panel requires 11 to 16 Vdc, < 1.0 A. Jack connector is  compatible with 5A2436.  TPM Connectors  DB9(F), either 4 or 8 connectors on the back panel  DB9 Pinout    Same as TPM DB9 pinout     Panel Meters can be calibrated independently of the TPMs.  Calibration There is one user‐adjustable pot for each meter.  Environmental Specs Temperature, Operating 0 to +50 °C (32 to 122 °F)  Temperature, Storage –20 to +80 °C (–4 to +176 °F) Altitude, Max 3,000 m (10,000 ft) above sea level ...
  • Page 26: Chapter 6 Maintenance

    Chapter 6 Maintenance Inspection and Cleaning This unit requires only simple and routine maintenance. WARNING Disconnect the unit from the RF power source and the ac  line before any disassembly. The potential for electrical  shock exists. CAUTION Do not use harsh or abrasive detergents for cleaning. 1. Wipe off dust and dirt regularly. Use a soft, clean cloth dampened with mild  detergent. Check connectors, connector pins, and cables for damage. If needed, clean the  connectors using a self‐drying contact cleaner that leaves no residue. ROHS Part Toxic or hazardous Substances and Elements Name Lead Mercury Cadmium Hexavalent Polybro- Polybro- (Pb) (Hg) (Cd) Chromium minated minated (Cr(VI)) biphenyls diphenyl (PBB)
  • Page 27 Maintenance Line Section Max Power vs. Frequency 10000 1000 6 1/8" 6-1/8 4-1/16" & 4-1/2" 3-1/8" 1-5/8" 7/8" 1000 10000 Frequency (MHz)
  • Page 28: Troubleshooting

    Transmitter Power Monitor/Meter Troubleshooting The TPM has no operator serviceable parts. Any required service must be  performed at an authorized service facility. The table below contains troubleshooting information for problems which can  occur during normal operation. This manual cannot list all malfunctions that  may occur or their corrective actions. If a problem is not listed or is not  corrected by the listed actions, notify a qualified service center. Transmitter Power Monitor Problem Possible Cause Correction Check power source,  No DC power  and cable Power LED does not  Return the unit to an  illuminate Defective LED authorized service  center Dirty connectors Clean connectors Defective connectors Replace connectors High VSWR  Shorted or open  Have the line serviced. transmission line 3140 Panel Problem Possible Cause Correction Set AC Power Switch to  Unit is not turned on  Display does not light Unit is not plugged in Connect ac power cord...
  • Page 29: Parts List

    Maintenance Parts List Customer Replacement Part Consists of Description 7006B112 Digital PC Board RPK7006B112 ‐ Replacement Digital  Board  920‐RPK7006B112 Instruction Sheet 7006A109 Metal Lid 7006A124 Label RPK7006A1000 ‐ Replacement Lid 5A2833‐LPC020CTP Light Pipe 5A2833‐RTN150 Rubber Ring 920‐RPK7000A1000 Instruction Sheet 2‐T‐MN‐2 ‐ Replacement 2‐T‐MN‐2 Terminations Terminations 5A2211‐2 ‐ Replacement  5A2211‐2 Data Calibration Data 5A2264‐09‐MF‐25 Cable, 9‐pin Male/ 5A2264‐09‐MF‐25 Replacement Cable Female, 25 ft 6A340‐ADJ ‐ Calibration  6A340‐ADJ Calibration Tool Adjustment Tool...
  • Page 30: Customer Service

    Transmitter Power Monitor/Meter Customer Service Any maintenance or service procedure beyond the scope of those in this chapter  should be referred to a qualified service center. If the unit needs to be returned for any reason, request an Return Material  Authorization (RMA) through the Bird Technologies website. All instruments  returned must be shipped prepaid and to the attention of the RMA number. Bird Service Center 30303 Aurora Road Cleveland (Solon), Ohio 44139‐2794 Fax: (440) 248‐5426 E‐mail: bsc@birdrf.com For the location of the Sales Office nearest you, visit our Web site at: http://www.birdrf.com...
  • Page 31: Limited Warranty

    Limited Warranty All products manufactured by Seller are warranted to be free from defects in  material and workmanship for a period of one (1) year, unless otherwise  specified, from date of shipment and to conform to applicable specifications,  drawings, blueprints and/or samples. Seller’s sole obligation under these  warranties shall be to issue credit, repair or replace any item or part thereof  which is proved to be other than as warranted; no allowance shall be made for  any labor charges of Buyer for replacement of parts, adjustment or repairs, or  any other work, unless such charges are authorized in advance by Seller. If Seller’s products are claimed to be defective in material or workmanship or  not to conform to specifications, drawings, blueprints and/or samples, Seller  shall, upon prompt notice thereof, either examine the products where they are  located or issue shipping instructions for return to Seller (transportation‐ charges prepaid by Buyer). In the event any of our products are proved to be  other than as warranted, transportation costs (cheapest way) to and from  Seller’s plant, will be borne by Seller and reimbursement or credit will be made  for amounts so expended by Buyer. Every such claim for breach of these  warranties shall be deemed to be waived by Buyer unless made in writing within  ten (10) days from the date of discovery of the defect. The above warranties shall not extend to any products or parts thereof which  have been subjected to any misuse or neglect, damaged by accident, rendered  defective by reason of improper installation or by the performance of repairs or  alterations outside of our plant, and shall not apply to any goods or parts  thereof furnished by Buyer or acquired from others at Buyer’s request and/or to  Buyer’s specifications. Routine (regularly required) calibration is not covered  under this limited warranty. In addition, Seller’s warranties do not extend to the  failure of tubes, transistors, fuses and batteries, or to other equipment and  parts manufactured by others except to the extent of the original  manufacturer’s warranty to Seller. The obligations under the foregoing warranties are limited to the precise terms  thereof. These warranties provide exclusive remedies, expressly in lieu of all other  remedies including claims for special or consequential damages. SELLER NEITHER  MAKES NOR ASSUMES ANY OTHER WARRANTY WHATSOEVER, WHETHER  EXPRESS, STATUTORY, OR IMPLIED, INCLUDING WARRANTIES OF  MERCHANTABILITY AND FITNESS, AND NO PERSON IS AUTHORIZED TO ASSUME  FOR SELLER ANY OBLIGATION OR LIABILITY NOT STRICTLY IN ACCORDANCE WITH  THE FOREGOING.

This manual is also suitable for:

Tpm73140a panelTpm seriesTpm33140

Table of Contents