Powering Up The Monitor; Power-Up Sequence; Monitor Power-Up Sequence - Scott Protege SG User Manual

Single gas monitoring device
Table of Contents

Advertisement

24
CHAPTER 4: OPERATION
Powering Up the
Monitor

Power-Up Sequence

Table 13 Monitor Power-Up Sequence
RESULTS
The monitor starts and runs through a self-test.
During the self-test ensure the following:
ON
LCD display "
".
Monitor emits one audible beep
LED lights for (1) one second
Monitor vibrates for (1) second
087-0065 REV B
To activate the detector, press and hold the left button.
Only power up the detector in an non-hazardous environment of 20.9% oxygen when
auto-zero function is enabled.
The LCD shows the power-up sequence while the detector initializes. The sensor
installed in the detector defines which gas type is shown on the LCD.
Warning: If the monitor does not operate in this fashion, do not use. Tag it out of
service. Failure to do so may result in serious injury or death.
Warning: Never operate the device in gas monitoring mode with the calibration
adapter attached. This can cause false readings of detected gases. Failure to do so
can lead to death or injury.
The displayed value (for normal mode) appearing on the monitor's LCD, may be
changed using the Portable Gas Detection (PGD) Application Software.
To power on the monitor, press and hold down the left button for three (3) seconds.
After pressing the Left button, the countdown screen appears, 3-2-1. If releasing the
left button, the monitor will return to power off mode. Beep, LED flash and vibration
on
Table 13
shows up in the power-up sequence.
DESCRIPTION
LCD display "On" and the Firmware version.
Then the current
month/date, year
and real time will
appear in sequence.
:
SCOTT SAFETY