Table of Contents
loading

Summary of Contents for Extreme Networks 32x

  • Page 1 3 2 x O P T I C A L L E V E L I N S T R U C T I O N M A N U A L...
  • Page 3 SAFETY Read the following safety instructions before ▪ Position the optical level securely on a level attempting to operate this product. surface. Damage or serious injury could result Keep these instructions in a safe place or store if the optical level falls. in the carry case for future reference.
  • Page 4: Item Checklist

    ITEM CHECKLIST Please ensure the following items are included with your optical level. If anything is missing please contact your retailer. Extreme 32x Plumb bob Carry case with attached lens cap (not pictured)
  • Page 5 OPTICAL LEVEL OVERVIEW Carry handle Optical view guide Focusing knob Bubble vial mirror Front lens element Bubble vial Panning screw Panning screw 360° protractor Levelling screw 5/8” x 11 tripod thread Base plate...
  • Page 6 USING YOUR OPTICAL LEVEL SET UP OPERATION ▪ Stand the tripod on a level surface with ▪ Point the instrument towards the desired legs spaced for stability. area. ▪ Thread the tripod yoke through the 58” ▪ Use a bright background or hold a white x 11 thread on the base of the instrument.
  • Page 7 Distance measurement ▪ Read the staff where it is cut by the upper and lower hairs. ▪ Calculate the distance between the upper and lower hairs. ▪ Multiply the figure by 100 to calculate the distance between the instrument and the staff.
  • Page 8 Height measurement ▪ Place the instrument onto a solid tripod ▪ Subtract the measurement of staff B between two staves. from staff A. (Example. 1.924m – 1.712m = 0.212m). This calculation shows that staff B ▪ Level off the optical using the bubble vial. is 0.212m higher than staff A.
  • Page 9: Checking Accuracy

    CHECKING ACCURACY BUBBLE VIAL ▪ Centre the bubble by adjusting the levelling screws until level. ▪ Turn the instrument 90° at a time and check the position of the bubble inside the vial. If the bubble moves out of the centre ring a service and calibration is required.
  • Page 10 COMPENSATOR ▪ Place the instrument onto a solid tripod ▪ Take an accurate measurement reading between two staves spaced 30m to 50m from Staff A. (Example. 1.696m. Subtract the apart. height difference that was calculated in step 3 from the current measurement reading ▪...
  • Page 11 1.712m 1.924m (1.696m) (1.484m) 0.212m...
  • Page 12: Care And Maintenance

    CARE AND MAINTENANCE ▪ This is a precision measuring instrument ▪ The operator should check the accuracy and should always be handled with care and of the instrument before precision levelling transported within the carry case provided. is attempted. Failure to do so may result in inaccurate measurements.
  • Page 13: Warranty

    WARRANTY The Extreme 32x comes with a 10 (ten) year manufacturers warranty. AUSTRALIA NEW ZEALAND “Our goods come with guarantees that more information please visit cannot be excluded under the Australian consumerprotection.govt.nz Consumer Law. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage.
  • Page 14: Customer Support

    CUSTOMER SUPPORT To assist you with any queries or technical questions please contact customer support Australia: 1300 658 338 New Zealand: 0800 367 527...
  • Page 15: Specifications

    SPECIFICATIONS Specifications Extreme 32x Product code 60163 Warranty 10 Years Accuracy ±1mm at 1km (double run) Compensator range 15° Magnification Minimum focus Objective aperture 36mm Weight (kg) 1.80 Dimensions (mm) 210 x 175 x 135...