1 Introduction 1.1 Scope and Purpose The 206A Olfactometer is an instrument designed to deliver a series of odor concentrations in a precisely controlled fashion. Careful fluid design of the instrument ensures that odors released have a well-defined “square” profile with minimal delay between activation of the valve and the arrival of the odor at the subject location.
Olfactometer Instruction Manual Model: 206A 2 Quick Start Guide This quick start guide will provide you with basic information that will allow you to get your new 206A Olfactometer up and running as quickly as possible. 2.1 Instrument Setup The 206A Olfactometer is shipped in 1 or more boxes depending on how many extra vial modules you have purchased.
Page 8
Olfactometer Instruction Manual Model: 206A The following is a list of several items required in order to use the 206A Olfactometer. These items are not included with the 206A. Required Items Description Clean/Dry Air Source A source of clean and dry air is required by the 206A. This air can be supplied by various means including: high pressure gas bottle, lab air supply or a compressor.
USB port on the PC. 4. Pipe the air supply to the 206A Olfactometer. (For the purposes of this quick start guide we will assume that you are using a high pressure air cylinder as your air source and therefore there is no requirement for drying, filtering or removing odors from the air source.)
Vial table above. f) Follow the same procedure to enter all odorants required by the study. Note: the 206A will auto detect the number of vials connected to the olfactometer and then grey out all vial numbers not present.
Mixing valve, and Final valve. Additional odor vial sections, each consisting of four vials and valves, can be added to the base 206A to increase the total vial count from 4 to 8 to 12.
Page 14
206A control program. The odor concentration released to the animal can be calculated as follows.
A miniPID fast response photo-ionization sensor was used to measure the output odor pulse shape and timing produced by the 206A olfactometer. Figure 2 shows a typical trace, in this figure the green line shows the odor concentration and the red line shows when the Final valve was turned on and the off.
3.3 Instrument Setup 3.3.1 Unpacking the Instrument The 206A Olfactometer is shipped in 1 or more boxes depending on how many extra Odor Vial Subassemblies you have purchased. Unpack all boxes included with the shipment and check the contents against the list provided below. If there are any issues with what you received, please...
Page 17
Accessories for 206B including: 4 vials with septa. present with 206B) Table 3-1: Items included with a 206A Olfactometer The following is a list of several items required in order to use the 206A Olfactometer. These items are not included with the 206A. Required Items...
USB port on the PC. 4. Pipe the air supply to the 206A Olfactometer. (For the purposes of this document we will assume that you are using a high pressure air cylinder as the air source and therefore there is no requirement for drying, filtering or removing odors from the air source.)
3.4.1 Insertion of Teflon hoses through Vial Septum An accessory kit is supplied with the 206A which includes a 16GA x 1.5” hypodermic needle that can be used to assist in inserting the 1/16” OD Teflon flow entry and exit tubes through the septum of the vial.
4 Control Software 4.1 Software Setup The 206A Olfactometer is supplied with a Software Install CD for Windows. Inserting this CD into a PC should launch the installer. Follow the on-screen wizard to complete installation of the control software. Following successful completion of the install wizard, the user should find an icon on the desktop to launch the control software.
Olfactometer Instruction Manual Model: 206A f) Follow the same procedure to enter all odorants required by the study. Note: the 206A will auto detect the number of vials connected to the olfactometer and grey out all vial numbers not present.
Model: 206A 4.3.2.3 Valve Check Menu Item Selecting this menu option will sequentially open and then close all vial valves attached on the 206A. This menu is only used to quickly check correct operation of the olfactometer. 4.3.2.4 Sequencer Menu Item Selecting this menu option will open the Sequencer window.
At the top of the window is a list box which describes the current setup of each vial. Note: the 206A will auto detect the number of vials connected to the olfactometer and grey out all vial numbers not present.
The software auto detects the number of vials connected to it and then updates the diagram to show all vials in the system. Figure 8 shows the System Status of a 206A with 4 vials while Figure 9 shows the same window with a 206A and a 206B installed for a total of 8 vials.
Model: 206A Figure 8: Advanced Status and Manual Control Window with 206A – 4 Vials Figure 9: Advanced Status and Manual Control Window with 206A and 206B – 8 Vials 4.7 Sequencer Window The Sequencer Window allows the user to create a sequence of challenges to be executed by the olfactometer with pre-defined timing.
Our performance guarantee states: if for any reason a new product does not meet your research needs then you can return it to Aurora Scientific for exchange or a full refund. The performance guarantee only applies to new products and must be exercised within 60 days of receipt of the instrument.
Products manufactured by Aurora Scientific Inc. are guaranteed to the original purchaser for a period of three (3) years. Under this warranty, the liability of Aurora Scientific is limited to servicing, adjusting and replacing any defective parts that are of Aurora Scientific manufacture. Aurora Scientific is not liable to the customer for consequential or other damages, labour losses or expenses in connection with or by reason of the use or inability to use the products manufactured by Aurora Scientific.
SHALL BE VALID UNLESS MADE IN WRITING AND SIGNED BY AN EXECUTIVE OFFICER OF AURORA SCIENTIFIC INC. 6.5 Returning Products to Aurora Scientific for Repair There are a few simple steps that must be completed before returning your product to Aurora Scientific. Obtain a Return Material Authorization number (RMA#). Go to our website or contact our technical support department to obtain a RMA #.
Weight [kg]: Dimensions [cm]: Mass Flow Control Module: 17L x 16W x 21H 4 CH Odor Module (206B): 17L x 16W x 21H Complete 206A: 38L x 16W x 21H 206A Software Recommended PC Specifications Processor: Quad-core Intel or AMD...
8 Appendix – Serial Communication Commands To communicate with a 206A, you must connect it to your PC via a USB port. The 206A will connect to the PC and appear as a serial COM port. The PC operating system will normally assign the connection to either COM3 or COM4.
Olfactometer Instruction Manual Model: 206A 8.1.2 Vial The vial command is used to set the state of a pair of valves on the 206A connected to each vial. The vials are numbered as follows: Vial Number Location/Function 1–3 MFC board...
Olfactometer Instruction Manual Model: 206A 8.1.4 findModules The findModules command is used to determine how many vial modules are connected to the 206A. Command findModules <I2C ADDR> e.g. findModules 1 Description: Inquire number of modules attached to olfactometer at address 1.