Summary of Contents for Millipore Millex-FG
Page 1
User Guide Millex ® Hydrophobic Filter Unit Single use only ■ Sterile ■ Non-pyrogenic ■ Do not use this product as an in-line filter for intravenous administration ■ SLFG M25 LS (50/box)
Page 2
Notice The information in this document is subject to change without notice and should not be construed as a commitment by Millipore Corporation. Millipore Corporation as- sumes no responsibility for any errors that may appear in this document. This manual is believed to be complete and accurate at the time of publication.
Page 3
Introduction The bi-directionally supported Millex-FG filter unit is a sterilizing filter for gases and liquids that are not compatible with standard membrane filters. It will remove microor- ganisms, particles, precipitates, and undissolved powders larger than 0.2 micron (µm). The sterile Millex-FG filter unit is non-pyrogenic and non-toxic. This single-use product consists of a 0.2 µm, hydrophobic Fluoropore™...
Chemical Compatibility The Millex-FG filter unit is compatible with most aqueous solutions. It may be used to filter the agents listed below. This guide has been developed from technical publica- tions, materials suppliers, laboratory tests, and field evaluations, etc., and is believed to be reliable.
Page 5
■ Do not use the same Millex-FG filter unit to filter solutions in both directions. ■ Do not use the Millex-FG filter unit to filter emulsions or suspensions. ■ Do not use the Millex-FG filter unit to filter 5 mg or less of active drug materials unless binding studies have been performed.
Page 6
Directions for the Millex-FG Filter Unit Fill the syringe with the solu- tion to be filtered. NOTE: If you are using the Attach the syringe to the Millex- filter unit with a vacuum or air FG filter unit and remove the...
For more information, contact the Millipore office nearest you. In the U.S., call 1-800- MILLIPORE (1-800-645-5476). Outside the U.S., see your Millipore catalogue for the phone number of the office nearest you or go to our web site at www.millipore.com/ offices for up-to-date worldwide contact information. You can also visit the tech service...
In the event of a breach of the foregoing warranty, Millipore's sole obligation shall be to repair or replace, at its option, the applicable product or part thereof, provided the cus- tomer notifies Millipore promptly of any such breach.