Page 1
Model 7600(M)/7602(M) Wideband Power Am pli fier DC to 1MHz 141V/282V rms 17/34 Watts Op er at ing and Main te nance Man ual...
Page 2
No other war ranty is ex pressed or im plied. Krohn-Hite Cor po ra tion re serves the right to make de sign changes at any time with out in cur ring any ob li - ga tion to in cor po rate these changes in in stru ments pre vi ously pur chased.
Section 1 - General Description Model 7600/7602 1.2 SPEC I FI CA TIONS 1.2.1 OUT PUT (All rms val ues are for sinewave sig nals) For the Model 7602, spec i fi ca tions ap ply to each out put to ground.
Section 1 - General Description Reg u la tion: <0.1%, No load to 1k ohm load, dc to 10kHz, ris ing to 2% at 1MHz. Cou pling: DC. DC Level: Nom i nal zero volts; vs. tem per a ture, 2mV/°C. DC Off set Con trol (no load): 0V to ±200V.
Ac cu racy: ±5% ±1 digit. Rack Mount ing Kit: Part No. RK-37, per mits in stal la tion of the Mod els 7600/7602 into a stan dard 19" rack spac ing. Op tion 003: Rear panel in put and out put con nec tors.
2.1 IN TRO DUC TION This sec tion de scribes the op er a tion of the Model 7600 and 7602. It in cludes the proper ac re quire ments, the rec om mended turn-on pro ce dure and a de tailed ex pla na tion of all the op er at ing con trols.
2.4 OP ER A TION To op er ate the Model 7600 and 7602 Power Am pli fier, pro ceed as fol lows: 1. Make the ap pro pri ate power con nec tions ( re fer to TURN-ON Pro ce dure).
Section 2 - Operation 2.4.1 Front Panel Fig ure 2.1 Model 7600M with Me ter Op tion Pack age Fig ure 2.2 Model 7602M with Me ter Op tion Pack age 2.4.1.1 Power Switch “I” is the ON po si tion, “0” is the OFF po si tion.
Section 2 - Operation 2.4.1.2 Mode The MODE se lec tor switch con sists of a 5-po si tion, ro tary switch that se lects the Mode of op er a tion. The fol low ing are a de scrip tion of each po si tion: OFF: When se lected, there will be no sig nal at the OUT PUT ter mi nals.
Section 2 - Operation 2.4.1.7 Vari able DC Off set Con trol The Vari able DC OFF SET Con trol con sists of a 10-turn po ten ti om e ter that in fi nitely con trols the Out put DC Off set 0V to –200V peak or 0V to +200V peak.
Section 2 - Operation 2.4.2 Rear Panel 2.4.2.1 Op tion 002 Con nec tors BNC con nec tors are pro vided for the In put sig nal, and Bind ing Posts are pro vided for the Out put sig nal. Fig ure 2.1 Rear Panel Panel 2.4.2.2 Chas sis/Float ing Ground Switch The CHAS SIS/FLOAT ING GROUND switch is a 2-po si tion slide switch that dis con nects the am pli fier’s...
3.1 IN TRO DUC TION The fol low ing pro ce dure should be used to ver ify that the Model 7600 and 7602 is op er at ing within spec i - fi ca tions., both for in com ing in spec tion and rou tine ser vic ing. Tests should be made with the cov ers in place, and the pro ce dure given be low should be fol lowed in se quence.
Con nect os cil lo scope chan nel two to the main out put. Set scope gain on chan nel two to 10V/cm, DC cou pled. Set the 7600/7602 for AC cou pling and a gain of 28dB (GAIN dB to 28, gain pot to 0 end stop), MODE to A.
Set the in put squarewave fre quency to 200kHz. Set the in put am pli tude to 2.0V p-p (1.0V peak). Set the 7600/7602 GAIN dB for 14, am pli tude con trol pot to 14 end stop. Set the scope sweep time to 0.5mS/cm.
Con nect the Fluke model 8920A true RMS volt me ter to the in put via a T adapter. Set it for dB rel a tive op er a tion. Zero ref er ence the me ter. Set the 7600/7602 for: MODE –B and AC. GAIN dB to 28 and gain con trol pot to 0 end stop.
3.3.6 Fre quency Re sponse Check Use a flat re sponse sig nal source such as a Krohn-Hite Model 4200 os cil la tor for this test. For the 7602 mea sure the in verted out put for each step.
ME TER SE LECT to PEAK VOLTS. In crease the 7600/7602 gain con trol pot to the point of ob serv able clip ping on the out put. With the Fluke mea sure the rms volt age at which clip ping oc curs.
Con nect the dis tor tion an a lyzer me ter to the main out put. Ad just the gain con trol pot for a 175V read ing on the 7600/7602 me ter. The dis tor tion an a lyzer should read <0.01% dis tor tion.
Page 24
Section 3 - Incoming Acceptance This page in ten tion ally left blank.