Charging Station - Instructions For Use READ THE HYBRESIS CONTROLLER AND PATCH INSTRUCTIONS FOR USE FOR ADDITIONAL IMPORTANT INFORMATION. REF: 199586...
GLOSSARY OF SYMBOLS This device may contain one or all of the following symbols: Refer to Instruction Manual/Booklet Council Directive 2002/96/EC concerning Waste Electrical and Electronic Equipment (WEEE). Indicates a requirement not to dispose of WEEE as municipal waste. Contact your local distributor for information regarding disposal of the unit and accessories.
Page 3
Warnings • Keep out of the reach of children. • Electronic monitoring equipment (such as ECG monitors and alarms) may not operate properly when this equipment is in use. • Electrodes should be removed before using equipment that generates high frequency or high energy electromagnetic radiation such as welding equipment, surgical electrocautery units, shortwave and microwave therapy.
DESCRIPTION The Hybresis Charging Station is designed to charge and store Hybresis Controllers. OPERATION The Hybresis Charging Station charges up to four Hybresis Controllers, maintains them in a fully charged state and provides a convenient place for storage. General -...
Page 5
Charging Station Lights - The two windows on each of the charging bays indicate the following: Light Status Corrective Action Green Charge Complete; None Required The Controller is fully charged and ready for use. Yellow Charging; None Required The Controller is charging normally.
Charging Station Specifications Voltage Requirement 100 – 240 VAC @ 50 – 60 Hz Current Requirement 0.125 A maximum Environmental Storage Temperature 5° - 40° C (41° – 104° F) Conditions DO NOT expose to temperatures above 50° C (122° F) Operating Temperature 15°...
Page 7
Guidance and manufacturer’s declaration – electromagnetic emissions The Hybresis Charging Station is intended for use in the electromagnetic environment specified below. The customer or the user of the Hybresis Charging Station should assure that it is used in such an environment. Emission tests...
Page 8
Guidance and manufacturer’s declaration – electromagnetic immunity The Hybresis Charging Station is intended for use in the electromagnetic environment specified below. The customer or the user of the Hybresis Charging Station should assure that it is used in such an environment. Immunity test...
Page 9
Guidance and manufacturer’s declaration – electromagnetic immunity The Hybresis Charging Station is intended for use in the electromagnetic environment specified below. The customer or the user of the Hybresis Charging Station should assure that it is used in such an environment. Immunity test...
Page 10
Recommended separation distances between portable and mobile RF communications equipment and the Hybresis Charging Station The Hybresis Charging Station is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Hybresis Charging Station can help prevent...
Service For clinical questions, contact the Professional Services Department at 1-800-328-2536 ext. 8506. If any component of the Hybresis System is not functioning properly or requires servicing, contact the Empi Repair Department at 1-800-862-2343. When returning any products, please include your name, address, phone number and a description of the problem.
Hybresis Patch Limited Warranty I. Warning While, in the opinion of Empi (“Empi”), the use of the Hybresis System (the “Product”) has met with some success, Empi makes no warranties to the purchaser as to the effectiveness of the product.
Page 13
II. Warranty A. Empi warrants to the initial Purchaser (“Purchaser”) (and to no other person) that the Product (with the exclusion of accessories) and the component parts thereof, distributed or manufactured by Empi, shall be free from defects in the workmanship and materials for three years from the initial date of purchase from Empi (the “Warranty Period”).
Page 14
Notice and return shipment shall be sent to Empi at 47492 SD Hwy. 22, Clear Lake, South Dakota 57226, USA, or to an Empi Authorized Service Center. To locate the appropriate service center outside of North America, or to request shipment approval, contact Empi directly.
Page 15
LIMITED TO, ANY IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE; PROVIDED, HOWEVER, THAT NOTWITHSTANDING THE FOREGOING SENTENCE, IN THE EVENT AN IMPLIED WARRANTY IS DETERMINED TO EXIST, THE PERIOD FOR PERFORMANCE BY EMPI THEREUNDER SHALL BE LIMITED TO THE LIFETIME OF THE INITIAL PURCHASER.