Advertisement

USER MANUAL

ZSP-150-F
ELECTRONIC BALANCE
December 2016
ARIZONA INSTRUMENT LLC
3375 N Delaware Street | Chandler, AZ 85225 USA
800.528.7411 | 602.470.1414 | f 602.281.1745
www.azic.com
Email:
General –
azi@azic.com
International –
intl@azic.com
Customer Service –
support@azic.com
ARIZONA INSTRUMENT LLC
3375 N Delaware St | Chandler, AZ 85225
800.528.7411 | sales@azic.com | www.azic.com
AZI P/N 700-0102-C
MADE IN THE USA

Advertisement

Table of Contents
loading

Summary of Contents for Arizona Instrument ZSP-150-F

  • Page 1: User Manual

    USER MANUAL ZSP-150-F ELECTRONIC BALANCE December 2016 ARIZONA INSTRUMENT LLC 3375 N Delaware Street | Chandler, AZ 85225 USA 800.528.7411 | 602.470.1414 | f 602.281.1745 www.azic.com Email: General – azi@azic.com International – intl@azic.com Customer Service – support@azic.com ARIZONA INSTRUMENT LLC 3375 N Delaware St | Chandler, AZ 85225 800.528.7411 | sales@azic.com | www.azic.com...
  • Page 2: Table Of Contents Zsp-150-F Electronic Balance

    ZSP-150-F Electronic Balance Setup and Operation Manual Thank you for choosing a Zeta Series SP-150 (ZSP-150-F) Electronic Balance from Arizona Instrument LLC. AZI is pleased to provide you with a balance designed and manufactured for years of reliable service and proudly made in the USA.
  • Page 3: Table Of Contents

    Leveling the Balance ......................5 Connecting to an Electrical Outlet ..................5 The On/Off Button (I/O) ......................6 RS-232 Interface Cable ......................6 ® ® Connecting the ZSP-150-F to the Vapor Pro / Vapor Pro Rx ...........7 ® ® Using the ZSP-150-F with the Vapor Pro / Vapor Pro Rx ..........7...
  • Page 4: Unpacking And Setup

    Therefore, place your balance on a solid, sturdy surface that is free of air currents and vibration and not in direct sunlight. The surface should not be magnetic and should be located away from doors, windows, heaters, air conditioners and fans. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 4 of 22...
  • Page 5: Leveling The Balance

    Please see the calibration instructions on page 10. AZI recommends that the balance be plugged into an electrical outlet at all times. This ensures that the balance is always warmed up and ready to use. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 5 of 22...
  • Page 6: The On/Off Button (I/O)

    Your balance is equipped with a bi-directional RS-232 compatible interface. The included Null Modem Cable (AZI P/N 200-0069) is a custom cable manufactured to the following diagram. Only this cable will provide the connections necessary for the ZSP-150-F Balance to communicate with the ®...
  • Page 7: Connecting The Zsp-150-F To The Vapor Pro / Vapor Pro Rx

    ® ® Connecting the ZSP-150-F to the Vapor Pro / Vapor Pro Rx ® The balance is configured to interface directly with the Vapor Pro series of Moisture Analyzers. ® ® Weight readings at the balance can be transmitted directly to the Vapor Pro or Vapor Pro through the null modem cable connection.
  • Page 8: Connecting The Zsp-150-F To The Vapor Pro Xl

    ® Connecting the ZSP-150-F to the Vapor Pro ® The balance is configured to interface directly with the Vapor Pro series of Moisture Analyzers. ® Weight readings at the balance can be transmitted directly to the Vapor Pro XL through the null modem cable connection.
  • Page 9: Zsp-150-F Front Panel Display And Controls

    ZSP-150-F Front Panel Display and Controls Figure 4 - Balance Front Panel The ON/OFF (I/O) Button The ON/OFF (I/O) button is located in the upper left corner of the balance’s front panel. When the balance is off, pressing the button will start the turn-on sequence shown in Figure 3. During this sequence the balance is doing an automatic systems checkout to ensure it is functioning properly.
  • Page 10: The Range/Function Button

    “OK”, then display the calibration weight including decimal places. The balance is now calibrated and in the normal weighing mode. 7 Remove the calibration weight. The display returns to zeros. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 10 of 22...
  • Page 11: Cal 2 - Linearity

    SAVEd is displayed followed by returning to normal operation. To exit the selection mode and return to normal operation, turn the balance off and then back on again at any time. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 11 of 22...
  • Page 12: Zsp-150-F Rear Panel

    ZSP-150-F Rear Panel The rear panel of the ZSP-150-F provides the power connector, an RS-232 connector, a USB connector and a security bracket. The security bracket may be used to provide a convenient method of securing the balance. RS-232 Power...
  • Page 13 ® ® Verify the balance’s ability to communicate with the Vapor Pro / Vapor Pro Rx as detailed in ® Connecting the ZSP-150-F to the Vapor Pro on page 7. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 13 of 22...
  • Page 14: Troubleshooting Guide

    Partially lit display, “Live” cord plugged into balance Unplug from wall outlet, wait 5 “weird”/”garbage” seconds, re-plug into wall outlet. characters on the display AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 14 of 22...
  • Page 15: Zsp-150-F Specifications

    ZSP-150-F Specifications Weighing Mode Single Range Capacity 150 g Readability 0.001 g Tare Range 0-150 g Repeatability (Standard Deviation) 0.001 g Linearity ±0.0015 g Stabilization Time User Selectable Weighing Speed Adaptive Update Interval 200 ms RS-232 bidirectional interface is standard. Baud rates...
  • Page 16: Fcc Compliance

    (1) year from the date of delivery. During the warranty period, Arizona Instrument will repair, or, at its option, replace at no charge, components that prove to be defective. The equipment must be returned, shipping prepaid, to Arizona Instrument's product service facility.
  • Page 17: Returned Goods Policy

    Returned Goods Policy Should it become necessary to return any item to Arizona Instrument for any reason, please contact our Customer Service Department at (800) 235-3360 or (602) 470-1414 to obtain a Return Materials Authorization (RMA) Number. When you call, please be ready to provide the serial number and a description of the problem.
  • Page 18: Appendix A - Front Panel Parts Counting

    MODE button. If you want to exit the counting You are now in the weighing mode you used just before mode, press the MODE button selecting pieces. again. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 18 of 22...
  • Page 19: Appendix B - Weight Checking

    If the part is too light, LO and the weight are displayed. The limits are stored until the balance is turned off. To exit weight checking, press the MODE button and the balance will return to normal operation. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 19 of 22...
  • Page 20: Appendix C - Live Animal/Severe Environment Weighing (T)

    RANGE/FUNCTION button This resets the time averaging mode. 8 If you want to exit the time averaging The balance returns to normal operation. weighing mode, press the MODE button. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 20 of 22...
  • Page 21: Appendix D - Percent Weighing (%)

    Step 2. If you want to exit the percent weighing The balance returns to normal operation. mode, press the MODE button. AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 21 of 22...
  • Page 22: Trademark And Copyright Protection

    , Arizona Instrument , AZI and the stylized AZI logo are all registered trademarks of Arizona Instrument LLC. Instrument firmware is copyright protected. Copyright 2007-2016 Arizona Instrument LLC. All Rights Reserved. Arizona Instrument LLC Zeta Series ZSP-150-F Electronic Balance User’s Manual...