Summary of Contents for Analytical Industries Palm Co
Page 1
2855 Metropolitan, Pomona, CA 91767 USA. Monoxide Monoxide Monoxide Tel: 909-392-6900, Fax: 909-392-3665 Email: Diveaii@aii1.com, Web: www.aii1.com Analyzer Analyzer Analyzer This manual may not be reproduced in whole or in part with- out the prior written consent of Analytical Industries Inc.
1.1 Indications for Use Warranty Safety Data Sheet (SDS) The Palm CO Carbon Monoxide Analyzer is intended to measure and display the concentration in the range of 0 to 25 PPM of carbon monoxide in com- 10.1 Disposal pressed breathing air tanks, intended for scuba diving. It provides continuous, fast, reliable, and accurate carbon monoxide measurements.
DO NOT use another manufacturer’s sensor inside this analyzer: The 6.3. design of the electronics is intended for only the Analytical Industries Inc. ALWAYS contact the factory if you are in doubt while troubleshooting. OSV-22-AF carbon monoxide sensors. Use of a different carbon monoxide...
3.3 Bump Testing: Prerequisite Gas Tests Prior to Start-Up and Use Bump testing is recommended before each use of the Palm CO ana- If the analyzer is located in a clean source of ambient air and the display lyzer. To conduct the test, one exposes the sensor to a gas mixture of...
1. The analyzer should be ON and located in a clean source of ambient air. attached to the bottom of the 2. PRESS the ZERO () key for 3 seconds. Palm CO. 3. The analyzer display will read “Zro”. Blocking the “vent” hole causes a 4.
6.2 Battery Replacement Changes in humidity will produce transients in sensor output that last for about The Palm CO Carbon Monoxide Analyzer is powered by two ten minutes. Therefore, it is recommended that at least ten minutes be pro- 1.5V AA alkaline batteries with an approximate life of 450 vided before taking measurements after a sudden humidity change, which may hours.
Page 7
6.2.1 Procedure: 6.3 Carbon Monoxide Sensor Replacement Place the analyzer face down on a flat surface and remove the four (4) DO NOT attempt to open or probe sensor. In event the sensor leaks, Phillips screws from the rear of the enclosure, FIG 2. refer to the Safety Data Sheet in Section 10.
To ensure proper operation after replacing the sensor or batteries check that: If the recommended corrective action does not resolve the problem return the The overview of the Palm CO Analyzer reassembly, FIG 7, is followed. device to the factory for service. ...
Limitations Carbon monoxide sensors and batteries should be disposed of in accordance Analytical Industries Inc. shall not be liable for losses or damages of any kind; with local regulations for batteries. loss of use of the analyzer; incidental or consequential losses or damages;...