ER920W Wireless
Arrhythmia Event Monitor

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ER920W and is the answer not in the manual?

Questions and answers

Summary of Contents for Braemar ER920W

  • Page 1 Braemar, Inc. ER920W Wireless Arrhythmia Event Monitor Braemar Inc. Phone: 800.328.2719 1285 Corporate Center Drive, Suite 150 Fax: 651.286.8630 Eagan, MN 55121 USA E-Mail: Service@BraemarInc.com Copyright 2010, Braemar Inc. All rights reserved...
  • Page 3 (1) year from the date of shipment from Braemar to the original purchaser. This warranty does not apply The ER920W is intended for use in an electromagnetic environment in to any product which Braemar determines has been modified or which radiated RF disturbances are controlled.
  • Page 4 ER920W Wireless Event Monitor NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
  • Page 5: Table Of Contents

    For body worn operation, this device has been tested and meets the FCC 1 and 2 Channel (4 Lead) Electrode Placement ........6 RF exposure guidelines when used with the Braemar pouch and accessories supplied or designated for this product. See Page 22 for a list Monitor Preparation ..................7...
  • Page 6: Overview

    Event recording is activated by the patient or by automatic event detection. The ER920W (1 or 2 channel) event monitor provides up to 30 minutes of total recording time and will operate as a simple looping event recorder for a minimum of 30 days with the Lithium Thionyl battery pack.
  • Page 7: Precautions

    E. Device should be powered with Factory supplier battery packs only. Do not use any other batteries, mix batteries, or use other power supplies in the Braemar ER920W wireless arrhythmia monitor. F. Do not use cellular phone to transmit patient data.
  • Page 8: Monitor Components

    If ±8 kV air ±8 kV air floors are covered with other battery in the ER920W. Use of any other IEC 61000-4-2 synthetic material, the relative battery pack or mixing batteries may cause humidity should be at least 30%.
  • Page 9: Setup Steps

    This manual is designed to allow a technician to follow the instructions Sample rate 256 samples per second page by page to set up the ER920W. Here is the general layout: User interface LCD display and sound Connect leads and electrodes to patient.
  • Page 10: And 2 Channel (4 Lead) Electrode Placement

    E-mail: service@braemarinc.com Web: http://www.braemarinc.com 1 and 2 Channel Electrode Placement Contact Braemar for further technical information. Use only Channel 1 if you have a single channel cable. Channel 1 = White and Red (V5 vector) Channel 2 = Black and Brown (V1 vector)
  • Page 11: Monitor Preparation

    ER920W Wireless Event Monitor ER920W Wireless Event Monitor Monitor Preparation Service and Maintenance General setup: Remove the Patient Cable if it is connected to the Monitor. Open the Cleaning battery compartment by sliding battery door downward. Install the Routine cleaning should be performed before each patient use. Remove battery pack of AA Lithium Thionyl batteries.
  • Page 12 ER920W Wireless Event Monitor ER920W Wireless Event Monitor Enter programming mode: Event Markings A. Push and hold both program buttons on the front of the Monitor until an audible tone is heard. The following shows typical event markings that are generated by the B.
  • Page 13 ER920W Wireless Event Monitor ER920W Wireless Event Monitor Note about Arrhythmia and AFIB detection: A. To change Arrhythmia and AFIB settings you must access the Troubleshooting (page 2 of 2) CUSTOM program option. Displays “Stopped” at top Batteries were pulled during programming.
  • Page 14 ER920W Wireless Event Monitor ER920W Wireless Event Monitor Choose program: Choose either the DEFAULT or CUSTOM option. The CUSTOM Troubleshooting (page 1 of 2) option allows changes to all of the following settings. Symptom Recommended Solution NOTES: No display Ensure batteries are inserted with correct polarity.
  • Page 15 Default = ON Receiving a Text Message: ARRHYTHMIA Program Arrhythmia Detection The ER920W monitor can receive text messages from the technician. DETECTION ON/OFF. Messages up to 40 characters can be displayed on the LCD of the Select “ON” to program, Brady Default = ON recorder.
  • Page 16 ER920W Wireless Event Monitor ER920W Wireless Event Monitor Send Data via wireless transmission: User Defined options continued In most cases, events are automatically downloaded to the receiving center via digital cellular link. There is no patient interaction required for this transmission to occur. One event will be stored before it is sent to...
  • Page 17 Automatic Record: ECG signal must occur before the ER920W wireless arrhythmia monitor initiates monitoring. If Arrhythmia Detection is ON and an event is detected, the Monitor will beep at the start of the recording only if the wireless transmission option is not active.
  • Page 18 ER920W Wireless Event Monitor ER920W Wireless Event Monitor Patient Operating Instructions The Monitor is ready to record events. Display overview: A. The general settings of the Monitor are displayed during operation by the number of dots after the word “MONITORING” shown on the display.