Catalogue 1、Attention for safety........................2 1.1 Attention for installation and maintenance ................2 1.2 Attention for electric system.....................2 1.3 Attention for fireproofing ......................2 1.4 Attention for safe operation .....................3 1.5 Attention for safety in chemistry and biology ................3 2、Symbols and meanings.......................4 3、Application and scope........................4 4、Work theory..........................5 5、Classify and name........................5 5.1 Classify ............................5...
1、Attention for safety Attention for safety clarifies the requirements of safe operation in the instruction manual. Please carefully read it before installation, operation, maintenance and repairing. Understanding the warning and correct operation can avoid the personal injury and the damage of centrifuge. 1.1 Attention for installation and maintenance ·The rotor accessory and package may be placed in the centrifugal chamber.
1.4 Attention for safe operation ·Please use the rotor and accessories which are designed for this centrifuge. ·Make sure if the centrifugal chamber is cleaned before operation. ·Make sure if the screw (nut) of the rotor in the centrifugal chamber has been screwed down.
Form 1 3、Application and scope TDZ4-WS low speed centrifuge makes use of centrifugal settling to separate, purify and condense the granule. It can be widely used in the fields of clinical medicine, biochemistry, genetic engineering and radioimmunology. It is ideal...
·It belongs to low speed centrifuge according to the rotating speed. 5.2 Name T D Z 4 - W S tabletop DC brushless motor low speed self-balance max.speed identification No., unit is 1000r/min. TDZ4-WS tabletop low speed centrifuge whose Max.speed is 4×1000r/min, is driven by DC brushless motor.
TDZ4-WS low speed centrifuge is composed of the main machine and accessories. The main machine is composed of case, centrifugal chamber, drive system, control system and the part of manipulation display. The rotor and centrifugal tube(bottle) belong to accessory. 8.2 Control board...
Page 9
▲: Increase key. Press “▲” to accelerate the setting parameter value. ▼:Decrease key. Press “▼” to decelerate the setting parameter value. ENTER:Press [ENTER] to confirm the setup working parameter value TDZ4-WS LCD display speed and time setting. Speed setting: Press key, when “...
8.3 Accessory Model Rotor Max.speed(r/min) Capacity (ml) RCF(×g) 18×10 ml 12×20 ml TDZ4-WS Angle Rotor 4000 2200×g 4×50 ml Form 2 9、Installation and adjustment 9.1 Instrument installation and its requirements: 9.1.1 Installation requirements: Warning: Don’t install centrifuge in the place with flammable and explosive matters.
support. · Clear the centrifugal chamber. ·Check the main machine, accessory, attached tools and documents according to the packing list. ·You should confirm that the voltage of the power supply should be consistent with the voltage the centrifuge requires. (Check the marking beside the electric outlet on centrifuge back, and the instrument’s power requirement is AC 220V 50Hz with single phase and three wires.) Connect the power cord(three wires and 10A) to the electric outlet on centrifuge back first, then put the plug on the other end into the...
10.3 Adding liquid and placing of the centrifugal tube: Put centrifugal tubes with adding liquid into the centrifugal tube of rotor and tubes should be placed symmetrically in even number. Otherwise, there will be vibration and noise because of imbalance. 10.4 Close the door lid: Press down the door lid until you hear “clicking”...
automatically slows down and stops running, and the [STOP] indicator light is (3)、Stop by manual: In the running status (the working time isn’t counted down to “0”), press [STOP] to slow down and stop running, and the [STOP] indicator light is on. 10.7 Take out the rotor: When the rotor stops rotating, press the switch with the mark“○”...
Page 14
Some lubricating oil should be smeared on drive shaft for protection. · Replace fuse (1) Cut off the power, and unplug the plug the power cord connected with the external electric supply socket. Then draw another end of the power cord connected with the centrifuge.
12、The common trouble and trouble shooting The LCD display window lights on and displays the relative error code when the error/trouble appears. Error Reasons Solution code The door lid isn’t closed during run. Check and Close door lid door The limit switch of door lid lock is not Check and replace limit isn’t closed in place or damaged.
Warranty Respected customers, Thank you for purchase of our company’s products. To protect your legal rights and avoid your worries, we provide warranty service for one year with the main machines. If it is broken under warranty with inartificial reason, we will supply free maintenance service.
Packing List Item Name Quantity Unit Remark Check Main TDZ4-WS centrifuge machine 12×20ml angle rotor Option Power cord Fuse BGXP 5×20 Two in 250V 5A fuse case Attached tool Packing List copy Instruction Manual copy Attached file Qualified Certificate copy...
Page 18
Quality Certificate Product Name:TDZ4-WS Tabletop Low Speed Centrifuge Serial No.:______________________________ This production is tested to be qualified for delivery. Checker: Date:...
Need help?
Do you have a question about the TDZ4-WS and is the answer not in the manual?
Questions and answers
What is error 7