Table of Contents

Advertisement

Instruction for use
UM009ENG LISA FULLY AUTOMATIC 201 10 Rev.4

Advertisement

Table of Contents
loading

Summary of Contents for W & H Lisa

  • Page 1 Instruction for use UM009ENG LISA FULLY AUTOMATIC 201 10 Rev.4...
  • Page 3: Table Of Contents

    Table of Contents INTRODUCTION..............................1 SCOPE ..............................1 FOREWORD ............................. 1 APPLICABLE EUROPEAN DIRECTIVES/NORMS ................... 1 INTENDED USE ............................1 SYMBOLS DISPLAYED ON THE STERILIZER ..................2 GENERAL AND SAFETY RECOMMENDATIONS ..................2 UNPACKING ................................ 3 UNPACKING THE STERILIZER ........................ 3 STANDARD ACCESSORIES ........................
  • Page 4 6.7.1 System info ..........................25 6.7.2 Formatting the memory card ...................... 25 USER AUTHENTICATION ........................25 FUNCTIONS AVAILABLE FOR THE ADMINISTRATOR ................ 26 6.9.1 Access the administrator functions .................... 26 6.9.2 Modifying a user name ......................27 6.9.3 Changing the administrator’s password ..................28 6.9.4 Resetting a user’s password to the default value (0000) ............
  • Page 5: Introduction

    The types of loads that can be sterilized with Lisa sterilizers are described in Table 1 of the reference tech- nical norm EN 13060. These loads include solid, porous, hollow loads type A and hollow loads type B, un- wrapped, single wrapped and double wrapped.
  • Page 6: Symbols Displayed On The Sterilizer

    SYMBOLS DISPLAYED ON THE STERILIZER Consult this table whenever one you see one of these symbols in this manual or on the sterilizer. ATTENTION Where this symbol is displayed on the sterilizer, the user must refer to this document. When shown in the user manual this symbol means ATTENTION IMPORTANT NOTES. To disregard the instructions given in this manual, incorrect use, poor maintenance or servicing by unauthorized personnel, clears the manufacturer of any responsibility for warranty and any other claims...
  • Page 7: Unpacking

    ON immediately. The sterilizer must be removed from the box and transported by two people. Lisa 517 40 kg Total weight: Lisa 522 50 kg Check the external condition of the box and the sterilizer. In case of any...
  • Page 8: Unit Description

    UNIT DESCRIPTION FRONT VIEW Water tank cover Sliding cover for Door seal manual filling of the clean water tank Chamber door Sterilization chamber Touch- screen Door pin Service door Dust filter SERVICE DOOR Connection to unlock the chamber door (ser- vice technicians only) Bacteriological filter Connection for the...
  • Page 9: Rear View

    REAR VIEW Male serial port Female serial port Parallel port (printer) Circuit breaker (manual reset) Mains plug socket Safety valve cover Mains cable guide Test connection Condenser vent Air filter Fitting for an external water supply system Fitting for continuous draining of used water...
  • Page 10: Description Of The Internal Water Tanks

    DESCRIPTION OF THE INTERNAL WATER TANKS The sterilizer is equipped with two independent water tanks, one for clean water and one for used water. The tank on the right side is called "clean water tank" and has to be filled with distilled or demineralized water required for the sterilization process.
  • Page 11: Chamber Rack

    The chamber usable space is the maximum volume of the chamber for accommodating a sterilization load. This volume is equivalent to a pipe with the following dimensions: Lisa 517 195 x 195 x 297mm (WxHxD); equal to the volume of 11.5 litres Lisa 522 195 x 195 x 390mm (WxHxD);...
  • Page 12: Installation

    The maximum weight of the sterilizer with a full clean water tank and the chamber fully loaded is: 52 kg (Lisa 517) 63 kg (Lisa 522) Leave a gap of 50mm in the back and 10mm on each side of the sterilizer to ensure adequate ventilation.
  • Page 13: Electrical Power Supply

    Select the printer type (§ 6.1.3). LisaPrint is the default setting (Print-p). All necessary data to document sterilization cycles is automatically printed (for details see Appendix 8). NOTE: Lisa 517/522 sterilizers offer the option to digitally save cycle data on removable memo- ry cards.
  • Page 14: Lisaware - Connecting The Sterilizer To A Computer (Optional)

    LISAWARE - CONNECTING THE STERILIZER TO A COMPUTER (optional) Connect up to four Lisa 517/522 sterilizers to a computer or computer network with LisaWare, a computer software that allows automatic saving of cycle data on a computer and the remote monitoring of the cycle progress on the computer screen.
  • Page 15: Automatic Water Supply

    AUTOMATIC WATER SUPPLY Your sterilizer can be connected to an external water supply system and to a continuous drain line. Once connected, the clean water tank will be automatically filled when empty, and the used water will drain con- tinuously. Thus, neither manual filling nor draining are required. The water supply system must deliver de- mineralized water meeting the requirements as listed in Appendix 7.
  • Page 16: Getting Started

    GETTING STARTED THE USER INTERFACE To power the sterilizer press the main switch located behind the service door (see image to the left). When the sterilizer is switched ON, the chamber automati- cally heats to 120° C. This process (pre-heating) takes approximately 10 minutes. If a sterilization cycle is initiated during the pre-heating phase, it will start automatically once the pre-heating tem- perature is achieved.
  • Page 17: Initial Warnings And Sleep Mode

    INITIAL WARNINGS AND SLEEP MODE This message will appear if the sterilizer is switched on and the clean water tank is empty. Fill the clean water tank with distilled or demineralized water (see § 5.4). If this message is displayed it is not possible to initiate a sterilization cycle. The mes- NOTE: sage disappears automatically once the clean water tank is filled.
  • Page 18: Filling The Clean Water Tank

    FILLING THE CLEAN WATER TANK 5.4.1 Manual filling Use only high quality distilled or demineralized water (see Appendix 7). • Slide the tank cover to the right to access the clean water tank inlet. • Remove the cap (1) from the tank inlet. •...
  • Page 19: Draining The Used Water Tank

    MEMORY CARD 5.6.1 Inserting / removing the memory card Lisa 517/522 sterilizers are equipped with a digital cycle data recording system. Cycle data are written and saved on removable/rewritable memo- ry cards.
  • Page 20: Programming

    PROGRAMMING Lisa 517/522 sterilizers allow the user to program a number of different features. The following section shows step by step how each of these features can be programmed. Press the menu icon to view the list of availa- The available menu options are shown ble options.
  • Page 21: Sleep Mode

    6.1.2 Sleep mode In “Sleep mode” the sterilizer reduces the Exit and return to Increase higher menu level power consumption to a minimum. The steri- time lizer will switch to sleep mode whenever it is idling without being used. Use this menu to program the time before the sterilizer automat- Confirm and return to ically switches to sleep mode.
  • Page 22: Date Format

    6.1.10 Date format Exit and return to Scroll up higher menu level Use this menu to change the date format. Press the “Scroll up” and “Scroll down” icons Confirm and return to select the format. Press the “Confirm” icon to previous screen to save the selection.
  • Page 23: Night Cycle Shortcut

    6.1.15 Night cycle shortcut If you frequently use the delayed cycle start option (see § 6.2) you can activate an icon on the main screen (cycle selection screen) that will serve as a shortcut to the delayed cycle start submenu. Exit and return to Scroll up Select the “Night cycle shortcut”...
  • Page 24: Connection To Pc/Log

    6.1.16 Connection to PC/Log Use this menu if you want your sterilizer to communicate with a computer in conjunction with the LisaWare software package. For further details please see § 4.4. Scroll up Exit and return to higher menu level Select the “Connection to PC/Log?”...
  • Page 25: Test Cycles

    TEST CYCLES Use this menu to run a variety of test cycles that might be mandatory in your country. Press the Exit and return Scroll up to higher "Menu" icon on the main screen and select menu level "Test cycles" from the menu (see § 6). Select the Test cycle by pressing the ”Scroll Selection up”...
  • Page 26: Delayed Cycle Start

    DELAYED CYCLE START Use this menu to delay the start of one or up to three consecutive cycles (e.g., if you want to load the steri- lizer in the evening and run a Vacuum Test followed by Bowie and Dick Test or a standard sterilization cycle early the next morning before office hours).
  • Page 27: Cycle History

    Return to pre- CYCLE HISTORY vious screen Scroll up Use this menu to view cycle data of the most recent 30 sterilization and/or test cycles. Cycle data is Saved cycles recorded on the main CPU board on a “first in – first out”...
  • Page 28: Maintenance

    MAINTENANCE Certain consumables have to be replaced periodically to guarantee a faultless operation of the sterilizer. Display messages will inform you whenever it is time to replace one of those components (see § 8). Press the "Menu" icon on the main screen and select "Maintenance" from the submenu (see § 6). Use this menu to view the number of cycles remaining before it is time to replace the: •...
  • Page 29: System Info

    USER AUTHENTICATION W&H Lisa 517/522 sterilizers, in combination with the optional LisaSafe package for label printing, provide an ultra-safe traceability concept in an extremely user-friendly way. They provide the following options, all separately selectable: •...
  • Page 30: Functions Available For The Administrator

    FUNCTIONS AVAILABLE FOR THE ADMINISTRATOR 6.9.1 Access the administrator functions To access the ADMINISTRATOR functions: 1) Press the “Menu” icon on the main screen that appears when the unit is switched ON. 2) Scroll the options until the “User authentication" option is highlighted and confirm. Abort and return to previous screen Scroll up...
  • Page 31: Modifying A User Name

    5) Select the desired option using the scroll buttons and confirm your selection: You can choose the following options: 1) Modify a user name Abort and return to previous 2) Change my password Scroll up screen 3) Reset a user’s passwords to the default value (0000) 4) Delete a user Confirm the choice...
  • Page 32: Changing The Administrator's Password

    6.9.3 Changing the administrator’s password The default password for the Administrator (and for any new user) is 0000. To change the Administrator’s password proceed as follows: 1) Scroll through the menu options and select the “Change my password” option and confirm. Abort and return to previous Scroll up...
  • Page 33: Resetting A User's Password To The Default Value (0000)

    6.9.4 Resetting a user’s password to the default value (0000) Use this function if a user has lost or forgotten his/her own password: it can be reset to "0000" so that the user can then enter with “0000” and save a new valid password. To reset a user’s password proceed as follows: 1) Scroll through the menu and select the “Reset password (0000)”...
  • Page 34: Deleting A User

    6.9.5 Deleting a user To delete the name of a user who no longer uses the sterilizer, proceed as follows: 1) Scroll through the menu options and select the “Delete user” option. Abort and return to previous Scroll up screen Confirm the choice Scroll down 2) Select the user to be deleted and confirm...
  • Page 35: Adding A New User

    6.9.6 Adding a new user This function allows to create a new user who is permitted to operate the sterilizer. 1) Scroll through the menu options, select the “Add new user” option and confirm. Abort and return to previous Scroll up screen Confirm the choice Scroll down...
  • Page 36: Setting The User Identification Options For All Users

    6.9.7 Setting the user identification options for all users Depending on the organization of the practice and the traceability requirements, the ADMINISTRATOR can set the following options: • - Identification and saving of the operator who has started the cycle; •...
  • Page 37: Functions Available To The Operator(S): Password Modification

    6.10 FUNCTIONS AVAILABLE TO THE OPERATOR(S): PASSWORD MODIFICATION Proceed as follows: 1) Press the “Menu” icon on the main screen that appears when the unit is switched ON. 2) Scroll the menu options and select the “User authentication” option. Abort and return to previous screen Scroll up Confirm and go ahead...
  • Page 38 5) The “Change my password” function is now available: select it by pressing the confirmation button. Abort and return to previous screen Confirm the choice 6) The password keypad appears. Enter your NEW four-digit password. WARNING! Abort and return The password is saved as soon as to previous the fourth digit has been entered.
  • Page 39: Starting And Ending A Sterilization Cycle With Identification Of The User

    6.11 STARTING AND ENDING A STERILIZATION CYCLE WITH IDENTIFICATION OF THE USER The instructions in this chapter are valid only if the “User Authentication” option is acti- NOTE vated. See § 6.8 6.11.1 Starting a sterilization cycle The steps required to start a sterilization cycle are basically the same as described in §7.1.1, with the fol- lowing changes: - After selecting and confirming the cycle, if the “Identification at cycle start”...
  • Page 40: End Of A Sterilization Cycle

    6.11.2 End of a sterilization cycle At the end of a cycle when the door is opened, if the “User name at load release” option has been acti- vated, a list of all authorized users will appear. Scroll through the list until your name is highlighted and confirm. Scroll up Abort and return to previous...
  • Page 41: Running A Sterilization Cycle

    RUNNING A STERILIZATION CYCLE THE AVAILABLE STERILIZATION CYCLES The following table shows the available sterilization cycles and the different type of loads that can be processed in the sterilizer. In total there are 3 type B sterilization cycles available; all offering the ECO- DRY feature (see below).
  • Page 42 See appendices tips,…) 6 and 7 Unwrapped, bagged, single/double wrapped Lisa 517: 4.5 kg solid items / 1.5 kg porous items MAXIMUM LOAD WEIGHT LIMITS: Lisa 522: 6.0 kg solid items / 2.0 kg porous items • Total type-B concept: All available sterilization cycles can sterilize and dry all types of loads: full solid, porous, hol- low A, hollow B, plastics, rubber, etc.;...
  • Page 43: Starting A Sterilization Cycle

    7.1.1 Starting a sterilization cycle Exit from sleep mode by pressing the Select a cycle by pressing the "Cycle selection" "Sleep mode" icon. icon and confirm the selection by pressing the "Se- lection confirmation" icon. Exit sleep mode Enter sleep mode Menu option Cycle selection...
  • Page 44: Cycle In-Progress

    7.1.3 Cycle in-progress While a cycle is in progress, the following parameters are displayed on the touch-screen: • Sterilization temperature and pressure – measured inside the chamber. • Elapsed time – time elapsed from the beginning of the sterilization cycle. •...
  • Page 45: Manual Stop

    MANUAL STOP A sterilization cycle in progress can be aborted manually at any time by pressing the manual stop icon. Once pressed, a warning screen will appear asking for a confirmation (see image below, top right). If the manual stop is confirmed, the alarm code A990 (see picture below) is displayed, and a two-minute reset phase starts to drain the pneumatic circuits and to depressurize the chamber (see image below, bottom left).
  • Page 46: Real-Time Cycle Data Information

    REAL-TIME CYCLE DATA INFORMATION Press the “Info” icon at any time during the sterilization cycle or when the sterilizer is idling to view a list of the most important technical parameters of the unit (see screen below, right). Scroll-up Info Manual stop Pause...
  • Page 47: Cycle Data Summary

    CYCLE DATA SUMMARY Press the "Info" icon at the end of a cycle before unlocking the chamber door (see image below, left) to view a summary of the cycle parameters (see images below, middle and right). The information screen lists the following parameters: •...
  • Page 48: Display Messages

    DISPLAY MESSAGES Once the sterilizer is switched on, a variety of messag- es can appear on the touch-screen. Messages show the status of the sterilizer; an example is shown aside. Message A message counter indicates the number of messages counter pending (e.g., message 01/02).
  • Page 49 If you press the “No” icon, the maintenance message will reappear after five cycles and the counter will keep counting up. Lisa sterilizers must be maintained according to the maintenance program (see § 11.1). Inadequate or inappropriate maintenance can lead to frequent malfunctions and...
  • Page 50: Alarms

    ALARMS Lisa sterilizers constantly monitor a variety of parameters during the sterilization cycle to assure a proper sterilization of the load. In case any of the parameters are off, potentially affecting the sterilization efficien- cy, the cycle is immediately interrupted and an alarm message is displayed.
  • Page 51: Alarm Code Table

    ALARM CODE TABLE Refer to the table below for troubleshooting assistance in case the sterilizer generates an alarm code. Code Description Action Power supply The load cannot be considered sterile. The cycle Power failure or significant drop in voltage occurred A010 must be repeated.
  • Page 52 Code Description Action Vacuum pump During a vacuum phase of the cycle, a vacuum of A310 –0.20 bar or lower could not be achieved. Clean, check or replace the door seal (§11.2) and the face side of the chamber; repeat the cycle. If the problem persists service.
  • Page 53: Maintenance

    MAINTENANCE There are two levels of maintenance: Maintenance performed regularly by the user. Maintenance carried out by authorized service technicians (see § 11.10). Before carrying out any maintenance on the unit, remove the mains cable. 11.1 MAINTENANCE PROGRAM Certain consumables have to be replaced periodically to guarantee a faultless operation of the sterilizer. Display messages will inform you whenever it is time to replace one of these components.
  • Page 54: Cleaning The Door Seal

    11.2 CLEANING THE DOOR SEAL ATTENTION! Operation to be carried out when the sterilization chamber is cold Clean the door seal and the chamber face side with a non ab- rasive cloth and a mild detergent solution. Rinse with clean wa- ter.
  • Page 55: Cleaning The External Sterilizer Surfaces

    11.5 CLEANING THE EXTERNAL STERILIZER SURFACES • Disconnect the mains cable. • Never use disinfectants or abrasive products. • Clean all external covers with a damp cloth and a mild detergent solution. • Do not use excessive amounts of water to wash the sterilizer as this may damage the electrical components and safety mechanisms.
  • Page 56: Cleaning The Water Tanks

    11.8 CLEANING THE WATER TANKS • Disconnect the mains cable. • Completely drain both tanks (see § 5.5). • Leave the drain tube attached to one of the drain quick connectors. • Turn the 5 screws of the tank cover a ½ turn counter- clockwise with the use of a screwdriver (or coin;...
  • Page 57: Replacing The Door Seal

    11.9 REPLACING THE DOOR SEAL ATTENTION! Operation to be carried out when the sterilization chamber is cold • Fully open the chamber door of the sterilizer. • Pull out the used door seal by hand. • Carefully clean the seal seat and the chamber face side with a cotton swab moistened with isopropyl alcohol.
  • Page 58: Service Conducted By An Authorized Service Technician

    11.10 SERVICE CONDUCTED BY AN AUTHORIZED SERVICE TECHNICIAN Regular service is imperative to ensure continuously effective sterilization. LISA 517/522 Sterilizers require servicing every 4000 cycles or 5 years by authorized service technicians. The service includes replace- ment of consumables and other important internal components, a safety and system check as well as tho- rough cleaning of areas that cannot be accessed by the user.
  • Page 59: Use Of The Memory Card

    USE OF THE MEMORY CARD Lisa 517/522 sterilizers come with a digital cycle data recording system. Cycle data is automatically saved on a removable/rewritable memory card. Data from memory cards can be managed with your PC or MAC. Memory cards are supplied with a USB cable and an external USB reader.
  • Page 60: Connecting The External Usb Card Reader To Your Pc/Mac

    Each cycle report is saved as a file in a HTM format. File names are generated automatically. Example of a file name: 01898 28-01-06 08_41_22 B&D _ HELIX LISA 517 02-0005.htm saving format depends on the Date-Clock setting (see § 6.1.10 and 6.1.11) 12.8 DIRECTORY NAME...
  • Page 61: Memory Card Management

    12.9 MEMORY CARD MANAGEMENT As soon as a memory card is inserted the unit automati- cally checks the card. In case the recorder detects that the memory card is damaged or defective, the following screen is shown. Confirm Press the “Confirm” icon and replace the card. If the card is not formatted or if it is formatted with a dif- ferent standard (e.g., FAT 12) the following screen is shown.
  • Page 62: Saving A File

    If the “Confirm” icon is pressed the formatting begins. During formatting the system also checks for errors. If it detects any errors the following screen is displayed. Press the “Confirm” icon to return to main menu and replace the card. Confirm 12.10 SAVING A FILE If a memory card is inserted, cycle data is saved automati-...
  • Page 63: Troubleshooting

    TROUBLESHOOTING Please find below a table of potential problems and easy solutions. For further troubleshooting information please also refer to § 8 (Display messages) and § 9 (Alarms). PROBLEMS POTENTIAL CAUSES SOLUTIONS The main switch or network cir- Activate the main switch or network circuit cuit breaker is OFF.
  • Page 64: Recycling / Disposal Instructions

    RECYCLING / DISPOSAL INSTRUCTIONS Lisa 517/522 sterilizers are mainly built from fiber-reinforced polymers, metals and electronic components. In case of disposal, separate the various components according to the materials they are made of. Drop the sterilizer with a company that specializes on recycling of related products. Do not abandon the sterilizer in unsecured places.
  • Page 65: Appendix 1 Technical Characteristics

    22 l / Ø 250 mm x D 440 mm (LISA 522) Usable space (identical for all cycles) 12 l / L 195 mm x H 205 mm x D 300 mm (LISA 517) 15.5 l / L 195 mm x H 205 mm x D 385 mm (LISA 522) Bacteriological filter 0.3 µm debris filter...
  • Page 66: Appendix 2 Sterilization Load Preparation

    APPENDIX 2 STERILIZATION LOAD PREPARATION Cleaning the instruments Thorough cleaning of all items prior to sterilization is imperative to ensure effective sterilization. If visible debris and bio-burden is not removed through cleaning prior to sterilization, it will interfere with microbial inactivation and compromise the sterilization process! •...
  • Page 67: Appendix 3 Maintenance Of Dental Handpieces

    APPENDIX 3 MAINTENANCE OF DENTAL HANDPIECES External disinfection This procedure reduces the risk of infection during cleaning and maintenance of the instrument. • Wear protective gloves during disinfection. • Refer to the instructions of the instrument manufacturer. • Avoid using abrasive disinfectants (ph-value 2.5 – 9; no chlorine based disinfectants). •...
  • Page 68: Appendix 4 Bowie And Dick Test

    APPENDIX 4 BOWIE AND DICK TEST The Bowie & Dick (B&D) Test device, also called Steam Penetration Test device, is representative of the small porous type load. It comprises several sheets of paper wrapped in a small packet in the middle of which there is a chemical heat-sensitive indicator sheet.
  • Page 69: Appendix 5 Helix Test

    APPENDIX 5 HELIX TEST The Helix Test device is representative of the type A hol- Indicator strip low load (EN 13060). It consists of a 1,500 mm long tube that is open on one side and closed with a capsule on the other side.
  • Page 70: Appendix 6 Vacuum Test

    APPENDIX 6 VACUUM TEST Use this test to validate the performance of the sterilizer in terms of leakage. During the test the following is checked: • Efficiency of the vacuum pump. • Tightness of the pneumatic circuit. The cycle profile specific for this test includes: •...
  • Page 71: Appendix 7 Water Quality

    APPENDIX 7 WATER QUALITY W&H Lisa sterilizers use distilled or demineralized water to generate steam for the sterilization process. Water quality sensors constantly monitor the quality of the water used for sterilization as water of too high mineral content can seriously damage the sterilizer and impair the sterilization process.
  • Page 72: Appendix 8 Example Of A Cycle Data Report

    APPENDIX 8 EXAMPLE OF A CYCLE DATA REPORT LISA 517 06-5790 0.0.0.0 -------------------------------------- -------------------------------------- -------------------------------------- -------------------------------------- Dr. Smith Cycle: B UNIVERSAL 134 Number: 01898 Date: 02/05/06 10:12:30 Phase Time part. T °C P Bar -------------------------------------- -------------------------------------- -------------------------------------- -------------------------------------- Start 00:00...
  • Page 73: Appendix 9 Accessories

    APPENDIX 9 ACCESSORIES DESCRIPTION ORDERING NUMBER Printer DP40 19721107 F523204x (Lisa 517) Perforated anodized aluminium tray F523205x (Lisa 522) Chamber racks: 190x43x300/375 * F523008x (Lisa 517) Reversible rack 190x43x300/375 * (Europe) F523009x (Lisa 522) 190x43x300/375 * 205x35x300/375 * Reversible rack...
  • Page 74 DESCRIPTION ORDERING NUMBER Door seal F460504x Dust filter F364502x Wall spacer F190107x USB MMC reader A801002x Multi Media Card A801001x Narrow bench securing bracket kit X051019x Drain tube mounting kit with fittings A812110X...
  • Page 75: Appendix 10 Helix Test Documentation Form

    APPENDIX 10 HELIX TEST DOCUMENTATION FORM Documentation For the documentation of sterilization cycles After the sterilization process the chemical indicator strip that was used in the “Helix Test” should be at- tached here (the strip comes with an adhesive part). The sterilization process was successful when the previously yellow marked spots turned to a consistent dark color.
  • Page 76 Authorized W&H service partners Find your nearest W&H service partner at http://wh.com Simply go to the menu option »Service« for full details. Alternatively please contact: W&H W&H W&H W&H UK LIMITED UK LIMITED UK LIMITED, 6 Stroud Wood Business Centre, Park Street, St.Albans, Herts AL2 2NJ UK LIMITED t +44 1727 874990 f +44 1727 874628...
  • Page 78 Manufacturer 93/42/EEC – Annex II W&H W&H W&H W&H Sterilization S.r.l UM009ENG LISA FULLY AUTOMATIC 201 10 Rev. 4 Italy, I-24060 Brusaporto (BG), Via Bolgara, 2 Subject to alterations t +39/035/66 63 000 f +39/035/50 96 988 wh.com wh.com wh.com...

Table of Contents