Summary of Contents for NOMACORC NomaSense CO2 P2000

  • Page 1 User Guide...
  • Page 2 Introduction About this User Guide Safety instructions Personnel operating and maintaining the device must be familiar with all aspects of its Intended use operation and be proficient in maintenance. • This User Guide describes the common Such personnel should review the following use and maintenance of the NomaSense precautions to promote safety awareness: P2000 device.
  • Page 3 Specifications - NOMASense CO P2000 General • Min. 2000 measurements on fully charged battery • Sample flow 130 - 165 ml/min standard • Auto power-off after 1 minute • Memory function for 10 measurements • Display hold function (also after power off) •...
  • Page 4 Overview ON/OFF key MEASURE key Starts a measurement cycle of Switches the device ON/OFF. Key is also used to interrupt a running 10  seconds. Pressing the key measurement or the display of during measurement will extend the measurement with collected measurements. another 10 seconds.
  • Page 5 TEMP./MEMORY key Activates temperature selection function. Key is also used to scroll through the saved measurements. (1) Needle holder Place the needle (5) in the holder when not in use. (2) Main display Shows the measurement result and the temperature setting. (3) Connector for external power supply/ battery charger “External...
  • Page 6: Operation

    Operation Switching ON/OFF When required temperature setting has been selected, press the (ON/OFF) key to Press the (ON/OFF) key to switch the device save the setting and return to measuring ON. The device beeps twice and the display mode (this will happen automatically (2) briefly shows the version number of the after app.
  • Page 7 Clear memory 6. Penetrate the septum (10) with the needle (5). Pressure will be equalized between If the memory is full, it must be cleared before the bottle and the air. new values can be stored. Press and hold 7. Close the valve (7) by turning left it ¼. The (TEMP ./MEMORY) key.
  • Page 8 CLICK Device messages 3. To remove the adapter simply press the lock (13) downwards and hold it while “P .On” Power on. External power supply removing the adapter. connected. “P .OF. “ Power off. External power supply Battery charging disconnected. When the display (2) shows “Lo.BAt.
  • Page 9: Cleaning And Maintenance

    Right-hand Left-hand Cleaning and maintenance Left side Right side Left side Right side General Battery pack Battery pack CAUTION! Personnel performing CAUTION! When using Gas OUT Gas IN Gas IN Gas OUT any cleaning or maintenance compressed air for cleaning, the must familiarize themselves with hose must be disconnected from the “Safety Instructions”...
  • Page 10 Calibration Performing 0% calibration The device should be calibrated if it indicates a value that deviates by more than 2-3% when • Prepare the device with a proper 0% CO measuring gas with a known CO content. calibration gas. • Press and hold the (MEASURE) key.
  • Page 12 Nomacorc LLC | 400 Vintage Park Dr. | Zebulon, NC 27597 | U.S.A. | Tel : +1 919-460-2200 | nomacorc.com Nomacorc | ZI Les Plénesses | Chemin de Xhénorie 7 | B-4890 Thimister-Clermont | Belgium | Tel : +32 87 63 88 20 | nomacorc.com...