Mainline Medical TOF-Watch S Operator's Manual page 16

Table of Contents

Advertisement

Calibration button (22)
Press the CAL button
To go back to the previous parameters press
The following parameters can be changed in the set-up menu: Default settings are shown bold.
Set-up parameter:
TOF s flashing (TOF s repetition time)
TOF s repetition time can be adjusted between 1 and
I
3
:
60 minutes.
Surface electrode + stimulation flashing (stimulation units)
mA:
Surface electrode stimulation strength in milli-ampères.
µC:
Surface electrode stimulation strength in micro-coulomb.
Surface electrode + stimulation flashing (pulse width)
200 µs:
Default surface stimulation pulse width can be selected
to 200 or 300µs.
Surface electrode + stimulation flashing (stimulation size)
50 mA:
Default surface electrode stimulation strength can be
adjusted between 0 and 60 mA (0 and 12/18 µC).
Needle electrode + stimulation flashing (stimulation units)
µC:
Needle stimulation strength shown in micro-coulomb.
mA: Needle stimulation strength shown in milli-ampères.
Needle electrode + stimulation flashing (stimulation size)
0.0 µC:
Default needle electrode stimulation strength can be
adjusted between 0.0 and 6.0 mA/0.24 µC.
CAL flashing (calibration sequence selection)
2:
Determination of supra-maximal threshold, followed by
calibration of transducer sensitivity at "threshold + 10%"
( max. 60 mA or 12/18 µC).
1:
Calibration of transducer sensitivity at user set
current/charge.
16 | TOF-Watch S Operator manual 33.516/A
(22) to go to next parameter to be modified.
(4) and then
(22).