Calibration button (22)
Press the CAL button
To go back to the previous parameters press
The following parameters can be changed in the set-up menu: Default settings are shown bold.
Set-up parameter:
TOF s flashing (TOF s repetition time)
TOF s repetition time can be adjusted between 1 and
I
3
:
60 minutes.
Surface electrode + stimulation flashing (stimulation units)
mA:
Surface electrode stimulation strength in milli-ampères.
µC:
Surface electrode stimulation strength in micro-coulomb.
Surface electrode + stimulation flashing (pulse width)
200 µs:
Default surface stimulation pulse width can be selected
to 200 or 300µs.
Surface electrode + stimulation flashing (stimulation size)
50 mA:
Default surface electrode stimulation strength can be
adjusted between 0 and 60 mA (0 and 12/18 µC).
Needle electrode + stimulation flashing (stimulation units)
µC:
Needle stimulation strength shown in micro-coulomb.
mA: Needle stimulation strength shown in milli-ampères.
Needle electrode + stimulation flashing (stimulation size)
0.0 µC:
Default needle electrode stimulation strength can be
adjusted between 0.0 and 6.0 mA/0.24 µC.
CAL flashing (calibration sequence selection)
2:
Determination of supra-maximal threshold, followed by
calibration of transducer sensitivity at "threshold + 10%"
( max. 60 mA or 12/18 µC).
1:
Calibration of transducer sensitivity at user set
current/charge.
16 | TOF-Watch S Operator manual 33.516/A
(22) to go to next parameter to be modified.
(4) and then
(22).
Set-up Display