User’s Manual Abacus junior 5 1 INTRODUCTION 1.1 Intended Use hematology analyzer is a fully automated cell counter Abacus junior 5 designed for in vitro diagnostic use. The instrument was developed for use in hospitals and small to medium sized labs.
User’s Manual Abacus junior 5 1.2.3 Reagents Only reagents recommended by DIATRON MI Ltd. should be used with the analyzer, otherwise accuracy cannot be guaranteed. Isotonic saline solution used to dilute whole blood Diluent: specimens and to rinse the fluidic system between measuring procedures.
User’s Manual Abacus junior 5 1.4 Parts of the Analyzer hematology analyzer can be divided into three main parts: Abacus junior 5 Performs sampling, diluting, mixing, and lysing Fluidic System: functions. Generates regulated vacuum used for moving cells through the aperture during the counting process.
Page 12
User’s Manual Abacus junior 5 Function keys Below you can find all the possible icons and functions assigned to the so-called soft- keys (function keys) Function key Action triggered Exit from actual menu or action Leave data-entry menu without saving any changes made to it...
Specification (assay values and allowed tolerances along with expiry date) regarding these materials is always enclosed with the approved control material. 1.6 Accessories This list can also be referred as the “DIATRON Abacus junior 5 - pack” Hematology Analyzer...
User’s Manual Abacus junior 5 2.3 Unpacking and installation 1. Carefully remove the hematology analyzer Abacus junior 5 from the shipping carton. Inspect the instrument for any visible signs of damage incurred during shipping. If you find any damage, immediately file a claim with the carrier or your distributor. Check the accessories received against the packing list.
User’s Manual Abacus junior 5 2.3.1 Turning the Instrument ON a. In case you use an external printer (for information, read manual shipped with the printer) connect it and turn it on. b. Turn the instrument on by flipping the power switch (above external power supply connector) to I position.
User’s Manual Abacus junior 5 2.3.3 Preparing for shipment The second item in Exit (Shutdown) menu should be used when the instrument is to be shipped or left unused for a longer time. The instrument will ask you to utilize the cleaning tube kit and 100ml of distilled water.
Page 22
User’s Manual Abacus junior 5 arrow The arrow under the Printing short cut-key indicates Printing function is enabled. b. You can select the desired item with the a nd keys and press the OK key to enter or activate the highlighted item.
User’s Manual Abacus junior 5 4 OPERATING PRINCIPLES 4.1 Impedance Method The volumetric impedance method (a.k.a. Coulter method) counts and sizes cells by detecting and measuring changes in electrical impedance when a particle in a conductive liquid passes through a small aperture.
User’s Manual Abacus junior 5 4.4 Absolute and Linearity Ranges of Measured Parameters Within linearity range, the instrument is guaranteed to provide the specified accuracy. Beyond this linearity range, the instrument is able to display results, but may not guarantee accuracy defined in specifications.
User’s Manual Abacus junior 5 The sample rotor will turn into inside of the instrument and the needle draws 90µl of sample from the tube. The aspirating needle is retracted, while its outer surface is automatically rinsed with diluent. This insures a low carry-over between samples.
Page 30
User’s Manual Abacus junior 5 In the followings we summarize the warning flags and give an explanation of their possible cause and a few hints to overcome the problem: Uppercase letters refer to WBC or HGB problems: Flag Meaning Recommended user action...
User’s Manual Abacus junior 5 5.3 Measure Pressing the MEASURE short-cut key, the measurement screen is displayed. The default measurement screen is set to Human profile. Patient data entry screen is accessible by pressing the button. Patient data entry screen allows entering...
Page 34
User’s Manual Abacus junior 5 Parameter 1. No flag at parameter 2. * flag at result 3. E (error) flag at result 0-10 g/l 10 - 25 g/l > 25 g/l 0 - 0.5 x10 cells/µl 0.5 - 1.0 x10 cells/µl...
Page 36
User’s Manual Abacus junior 5 <DATABASE LOCAL MENU> SELECTION (2) Besides checking samples one-by-one, selection can also be done within Database local menu. SELECT ALL checks all boxes, DESELECT ALL clears all boxes. <DATABASE LOCAL MENU> SELECTION (2) SELECT BY DATE, TIME AND ID...
User’s Manual Abacus junior 5 7 UTILITIES 7.1 Maintenance By selecting item (1) of the UTILITIES you can access the MAINTENANCE menu. 7.1.1 Regular Maintenance Jobs Maintenance submenu offers tools including maintenance procedures such as cleaning, priming, draining. MAINTENANCE (1) Select the required submenu.
User’s Manual Abacus junior 5 WASTE HANDLING – VERY IMPORTANT Waste contains poisonous substances (because of possible cyanide content) and human origin substances causing biohazard. These substances are representing potential danger to environment. For this reason, safe handling of the waste liquid is very important.
Page 42
User’s Manual Abacus junior 5 7.1.6.1 Peristaltic pump maintenance The pump installed in the instrument is maintenance free. However, would you experience leakage from the pump, or get vacuum error displayed, you may replace the tube used. 1. Open the rear door.
User’s Manual Abacus junior 5 7.2.1 Factorial calibration If the CALIBRATION MODE has been previously set to Factorial Calibration, the factors can be set manually in the 0.80 - 1.20 range. UTILITIES CALIBRATION (2) CALIBRATE (1) (factorial) Enter previously calculated factors using numerical keys;...
User’s Manual Abacus junior 5 7.3 Quality Control By analyzing control materials, day-to-day reproducibility can be monitored. Both target values and acceptable ranges for each parameter can be specified for different QC levels herein. provides six different Quality Control levels. You can Abacus junior 5 configure up six individual reference sheets for each control material (e.g.
User’s Manual Abacus junior 5 UTILITIES QUALITY CONTROL (3) VIEW QC DIAGRAM (4) Means, standard deviations (StDev) and coefficients variation (CVar) calculated based on the QC analyses. The dotted lines delineate acceptable ranges on Levey-Jennings charts. You can look at Quality Control information and results in table format as well. To access this format, select item 3 View table of QC measures of Quality Control.
User’s Manual Abacus junior 5 7.5 Settings Selecting item four (4) of the UTILITIES you can access this menu. The SETTINGS sub-menu allows the user to set the fluid detector operation, printing parameters, dates, user modes and various data. 7.5.1 Printer Settings Make sure to select the appropriate printer mode for correct operation before printing.
Page 52
User’s Manual Abacus junior 5 Table format defines how a table-printout will look like. You can select normal or narrow view. A normal table is as wide as the paper, narrow format has parameters grouped such, that Printer cable defines which communication is to be used between instrument and printer.
Page 54
User’s Manual Abacus junior 5 7.5.1.1 Troubleshooting guide for printing problems Problem Possible reasons/remedies Printer does not respond, • Printer is off. Turn it on. no printout. • Printer is not connected to analyzer. Connect it to the parallel or USB port of analyzer.
Page 56
User’s Manual Abacus junior 5 7.5.2.3 Laboratory This menu allows entering laboratory information. This will be printed in the header of reports printed by the instrument. SETTINGS (4) CUSTOMIZE (2) LABORATORY (5) A maximum of 40 characters can be entered in each line. User can enter this...
Page 58
User’s Manual Abacus junior 5 7.5.2.4.1 Multi-user mode If the analyzer is in multi-user mode, a login screen will be displayed during start-up, asking for a user ID and a password. After the User ID is entered and confirmed with...
User’s Manual Abacus junior 5 7.6 Software Update Instrument software can be updated easily following the steps below: 1. Shut down the instrument, and switch it off. 2. Insert the upgrade floppy disk in the floppy drive. 3. Switch on the instrument and follow the on-screen instructions.
Page 62
User’s Manual Abacus junior 5 Full printout format with histograms: Header Sample and measurement info Report info Histograms Limits, or graphical location of the result within the range Location of of limits warning flags (not displayed now) Figure 16. Built-in printer printout...
Need help?
Do you have a question about the Abacus junior 5 and is the answer not in the manual?
Questions and answers