2 additional, freely 1 additional analog output configurable relays for the measured temperature value Oxi 170 RT RS as RT option as RT option 1 x RS 485 1.2.2 Freely configurable relay (RT and RT RS option) Both potential-free K1 and K2 relays of the RT and RT RS instrument...
Oxi 170 Overview 1.2.4 RS 485 digital interface (RT RS option) The monitor can communicate with other instruments via the RS 485 digital interface of the RS instrument versions. The following operating modes are possible: Master mode ( mode): Unidirectional mode for the output of measured values.
This operating manual contains essential instructions that have to be followed during the commissioning, operation and maintenance of the Oxi 170 monitor. Thus, it is essential for the operator to read this component operating manual before carrying out any work with the system.
Safety instructions Oxi 170 Authorized use The authorized use of the Oxi 170 monitor consists solely of stationary measurement in water and wastewater applications, seawater, brackish water and aquacultures. Read the technical specifications in accordance with chapter 8 T . Only operation and use ECHNICAL DATA according to the instructions in this operating manual is authorized.
Netz Option Option Mains Fig. 3-1 Mounting boreholes of the Oxi 170 The clearances of the boreholes are given in the dimensional drawing in the chapter 8 T . For wall mounting, use the suitable ECHNICAL DATA material (screws, wall plugs, etc.).
Installation Oxi 170 Electrical terminal strip The terminal strip of the Oxi 170 is accessible after the housing cover has been opened: 13 12 11 Sensor/Probe Rec2 Rec1 g f e d Netz Option Option Mains Fig. 3-2 Terminal strip Oxi 170 inside the housing...
Seal all open cable glands with matching blind plugs. WARNING Free wires must not protrude into the housing of the Oxi 170. Danger of short-circuit and fire! Touch-proof circuits can also come into contact with dangerous voltages if touched and lead to life threatening situations when working with the Oxi 170.
Installation Oxi 170 3.3.2 Line power connection WARNING If the power supply is connected incorrectly, it may represent a danger to life from electric shock. Pay attention to the following points during installation: The monitor must only be connected by an electrician.
Oxi 170 Installation 3.3.3 Relay contacts and current outputs WARNING If external electrical circuits that are subject to the danger of physical contact are incorrectly connected to the relay contacts, there may be a danger of life threatening electric shock. Electrical...
Installation Oxi 170 Connecting the sensors 3.4.1 Connecting the oxygen (D.O.) sensor WTW online D.O. sensors have a permanently mounted connection cable with a 7-pin plug adapter. Connection Connect the plug adapter of the sensor connection cable with the sensor input jack of the monitor:...
Oxi 170 Installation 3.4.2 Cable extension The EK 170 cable (order no. 108 206) can be used in conjunction with the KI/S terminal box (order no. 108 606) for cable extension between the sensor/electrode and the monitorand the K 160 cable plug (order no.
Operation Oxi 170 Operating keys The monitor is operated via the following six keys below the display: Key functions Function Call up measuring mode <M> Start calibration (only active in measuring mode) <C> In measuring mode: Display the sensor slope Confirm current selection <ENTER>...
Operation Oxi 170 Configuration 4.4.1 Call up configuration level Call up the configuration level from the operating level as follows: 1 Press <DOWN>. 2 Press <ENTER>. 3 The menu for the password prompt appears. When delivered, no password protection is set up. In this case proceed to the next operating step by pressing <ENTER>...
Oxi 170 Operation 4.4.2 Configuration setting table General operating The starting point for the settings is the start display of the configuration instructions level (see section 4.4.1): Swap to the first setting with <ENTER>. Operation: All settings of the following setting table are run through step by step.
Page 24
Operation Oxi 170 Setting table: Setting Selection/Values Explanation (values in bold typeface = values on delivery) Switches the salt content correction on (On) or off (Off, delivery condition). The salt content correction allows for the salinity equivalent entered in the parameter level when calculating the measured value (see section 4.5.2)
Page 25
Oxi 170 Operation Setting Selection/Values Explanation (values in bold typeface = values on delivery) mg/l / % Measured value range of the REC 1 analog output (D.O. value). After confirming the reC (mg/l or %) selection with <ENTER> you can adapt the upper final value.
Page 26
Operation Oxi 170 Setting Selection/Values Explanation (values in bold typeface = values on delivery) °C Measured value range of the REC 2 analog output (temperature value). After confirming the reC (°C) selection with (only with RT and RT RS option) <ENTER>...
Oxi 170 Operation Parameterization 4.5.1 Calling up the parameterization level Call up the parameterization level from the operating level as follows: 1 Press <DOWN>. 2 Press <ENTER>. 3 The menu for the password prompt appears. When delivered, no password protection is set up. In this case proceed to the next operating step by pressing <ENTER>...
Operation Oxi 170 4.5.2 Setting table of parameterization Overview of the settings Instrument version (main level) Display Description Stan- RT RS dard Salinity equivalent Current output 1 Current output 2 Relay limits Air pressure Site altitude RS 485 interface Password protection...
Page 29
Oxi 170 Operation specific configuration settings. Settings that are not available will be skipped. From any point of the main level press <ESC> to return to the starting point of the parameterization level. From here change to the operating level with <M> or <ESC> or to the configuration level with <UP><DOWN>.
Page 30
Operation Oxi 170 Setting Selection/Values Explanation (values in bold typeface = values on delivery) Current range and attenuation of the REC 1 current output in rEC operating mode (analog output for the D.O. value). Operating notes: – The setting is carried out in the order: =1 ...
Page 31
Oxi 170 Operation Setting Selection/Values Explanation (values in bold typeface = values on delivery) 500 ... 1100 Entry of air pressure for air pressure compensation (in hPa or mbar). Notes: This is the long term average value of the actual air pressure (QFE) at the location.
Page 32
Operation Oxi 170 Setting Selection/Values Explanation (values in bold typeface = values on delivery) -=5 ... =5 Comparison of the measured temperature value against a reference thermometer. Notes: – Due to the thermal capacity of the sensor, it is necessary to place it in a container with at least 2 liters of water.
Oxi 170 Operation Freely configurable relays (RT and RT RS option) 4.6.1 Configuration Relay functions The two potential-free relays of the RT and RT RS instrument variants can be freely configured. The following functions are possible: Note The relay functions are selected in the configuration level under the setting Crc (see section 4.4.2).
Operation Oxi 170 Function Description CS (Cleaning System) The relay controls an external sensor cleaning unit. The relay works as a closer (i.e. normally open). During the cleaning process the measured value is frozen. The parameterization of the relay is described in section 4.6.3.
Page 35
Oxi 170 Operation Relay state Example: Monitoring of limit values with one or two relays Relay 1 Relay 2 Time Measured value Relay 1 Hysteresis HS1 Hysteresis HS1 Hysteresis HS2 Hysteresis HS2 Relay 2 Time Fig. 4-2 Switching points of relays working as limit indicators The selected limit value (UL/LL) is exceeded/undercut.
Page 36
Operation Oxi 170 Setting table for limit indicators: Setting Selection/Values Explanation UL or LL Any within the Upper or lower limit. measurement range Setting tolerance according to the measured value display. Operating note: – Set the first (flashing) digit with <UP><DOWN>...
Oxi 170 Operation 4.6.3 Timer for external sensor cleaning system Function The cleaning function enables the timing of a sensor cleaning system through one relay of the monitor. The relay always works as a closer (normally open). Setting table for sensor cleaning:...
Operation Oxi 170 4.6.4 Display of the relay states in the operating level For an active relay the REL 1 or REL 2 status display appears in the operating level. Depending on the selected switching behavior the relay contact is either open or closed.
Page 39
Oxi 170 Operation Calibration routine 1 In the measured value display: C A L Call up the calibration routine with % mg/l <C>. The CAL display appears. Note: The states of all relays and Baud ° current outputs linked with the...
Page 40
Operation Oxi 170 7 Press <ENTER>. mg/l The display switches to the normal measured value display. The states of all relays and current ° outputs linked with the measured oxygen value again follow the measured value. ba24105e08 11/2010...
Oxi 170 Operation Display of instrument information The following information can be called up: Configuration data. For this, switch to the configuration level. Press <ESC> during the password prompt to view all settings in view mode (see section 4.4.1 C ALL UP CONFIGURATION LEVEL Parameterization data.
Operation Oxi 170 Test mode General information Test mode can be used for the following purposes: Set specific current values on the current outputs for test purposes Switch the relays on and off for test purposes (RT and RT RS...
Page 43
Oxi 170 Operation Test of current output 2 t s t (only for RT and RT RS option). 6 Press <ENTER>. REL1 REL2 Here a fixed current value can be set on current output 2: – Set 0.1 mA with <M>...
Page 44
Operation Oxi 170 The tSt display flashes. The RS 485 communication test is active. The RS 485 interface works as a repeater, i.e. all received blocks are sent back again. Special commands: REL1 REL2 – The monitor sends the instrument identification according to the RS command "RSID"...
Oxi 170 Maintenance and cleaning Maintenance and cleaning Maintenance The measurement instrument is maintenance free. Cleaning Occasionally wipe the outside of the measuring instrument with a soft, damp cloth. CAUTION Do not use high-pressure water blasters for cleaning (danger of water penetration!).
Page 46
Maintenance and cleaning Oxi 170 ba24105e08 11/2010...
Page 47
Oxi 170 What to do if ... What to do if ... CAL Err Cause Remedy (Calibration error) – There is no valid calibration – Recalibrate Display OFL Cause Remedy – The measured D.O. or – Check the measurement temperature value is outside...
Page 48
What to do if ... Oxi 170 ba24105e08 11/2010...
Cable and terminal box Description Model Order for cable extension Cable plug K170 109 508 Cable EK/170 108 206 Passive terminal box KI/S 108 606 Note More accessories are available in the WTW catalog or on the Internet under www.WTW.com. ba24105e08 11/2010...
Oxi 170 Technical data Technical data General data Front view: Dimensions Oxi 170 mg/l ° REL1 Rear view: Side view: Fig. 8-1 Dimension drawing of the Oxi 170 (dimensions in mm) Test certificates ba24105e08 11/2010...
Page 52
Technical data Oxi 170 Mechanical Housing material Glass fiber reinforced Polycarbonate construction Membrane keyboard Polyester Weight approx. 2,2 kg Type of protection IP 66 Ambient conditions Temperature Operation -25 °C ... +55 °C (-13 ... 131 °F) Storage -25 °C ... +65 °C (-13 ... 149 °F) Relative humidity ≤...
Page 53
Oxi 170 Technical data Electrical connections 13 12 11 Rec2 Rec1 Sensor/Probe g f e d Netz Option Option Mains Fig. 8-2 Terminal strip Connection technique Connections – 7-pin sensor input jack on the underside of the housing – Screwless plug-in connections in the...
Page 54
Technical data Oxi 170 Current outputs Output Galvanically separated from the sensors Current range Adjustable: – 0 - 20 mA – 4 - 20 mA Accuracy 0.1 % of current value ± 50 µA, load max. 600 Ω Functions – 1 x analog output for measured D.O. value –...
Technical data Oxi 170 8.2.2 Temperature measurement Temperature sensor NTC (integrated in the sensor) Measuring range -5.0 ... 50.0 °C Resolution 0.1 K ± 1 digit Accuracy 0.2 K Adjustment method Manual via comparative measurement of ± 0.5 K FCC Class A Equipment Statement...
Oxi 170 Indexes Indexes This chapter provides additional information and orientation aids. List of display abbreviations (see section 9.1) Definition of terms (see section 9.2) Index (see section 9.3) Display abbreviations Configuration Password Starting point of the configuration level Salt content correction...
Page 58
Indexes Oxi 170 Temperature balancing Cleaning system Parameterization Upper limit Lower limit relay Hysteresis Switching delay Cleaning interval Cleaning duration Adjustment duration Parameterization Attenuation of current outputs Miscellaneous Software version Test mode ba24105e08 11/2010...
Oxi 170 Indexes Glossary The glossary briefly explains the meaning of the specialist terms. However, terms that should already be familiar to the target group are not described here. Adjusting To manipulate a measuring system so that the relevant value (e.g. the...
Page 60
Indexes Oxi 170 Slope (relative) Designation used by WTW in the D.O. measuring technique. It expresses the relation of the slope value to the value of a theoretical reference sensor of the same construction type. Test sample Designation of the test sample ready to be measured. Normally, a test sample is made by processing the original sample.