  • Page 1 CCT-3300 Series Conductivity Controller Operation Manual...
  • Page 2 Introduction Thanks for your choosing CCT-3300 series conductivity/resistivity controller manufactured by Hebei Create Instrumentation Technologies Co., Ltd. Correct sensor installation and parameter setting would show great performance and advantage of this instrument for your good usage. So please carefully read this manual before installation.
  • Page 3 D).The relevant damage caused by choosing the wrong type. E).The physical damage caused by ultimate load F).The meter is out of operation caused by improper storage and transportation (refer to SJ/T10463-93 standard) G).Consumable material is out of maintenance service and need purchase separately..
  • Page 4: Table Of Contents

    Content Ⅰ.General ....................1 1.1 Applications....................1 1.2 Classifications................... 1 1.3 Technicial Features................... 1 Ⅱ.Installation..................2 2.1 Electrical connection................2 2.2 Installation of electrode............... 3 Ⅲ.Settings .................... 5 3.1 Mode settings....................6 3.1.1 Electrode type setting..............7 3.1.2 Electrode constant setting............7 Ⅴ.Complete set..................
  • Page 5: Ⅰ.general

    Ⅰ.General CCT-3300 series is a new developed conductivity/resistivity integration controller manufactured by Hebei Create Instrumentation Technologies Co., Ltd. Please select the suitable electrode in application. You could only set the electrode type in the menu, and the instrument could run in conductivity or resistivity measurement automatically.
  • Page 6: Ⅱ.installation

    parameter Conductivity 0.50μS/cm~20.00mS/cm 0.01μS/cm Resistivity 0.05MΩ·cm~18.25MΩ·cm 0.01MΩ·cm 0.25ppm~10.00ppt 0.01ppm ±0.5℃ Temp. 0~50℃ 0.1℃ Temp. element NTC10K Isolated, reversible, adjustable, instrument/transmitting mode 4-20mA for selection output Loop resistance 400Ω(Max)DC24V Control SPDT relay, Load capacity AC230V/5A(Max) contact Note: CCT-3320T has no relay Storage Temp:(-20~60)℃...
  • Page 7: Installation Of Electrode

    CCT-3300E CCT-3310E CCT-3320E/CCT-3320 CCT-3320T Wiring instruction : Connect white line of cell (WHITE) Connect green line of cell(GREEN) Connect yellow line of cell (YELLOW) Connect red line of cell (RED) I+/I- Instrument mode(powered by instrument) T+/T- Transmitting mode(powered by conditioning module) RELAY Relay contact(SPDT NO/NC)...
  • Page 8 Correct installation modes of probe Please strictly select installation position and installation mode. Since the incorrect installation will cause the reading error. 1) The electrode should be installed in a place in the circulating and closed pipeline where the stream is steady and air bubbles are hard to generate. 2) No matter the concentric electrode is horizontally, slant or vertically installed, the front head should be deeply inserted into the moving water and face to water flow direction installation.
  • Page 9: Ⅲ.settings

    2. Check the temperature/mA reading under measurement status. 1.Enter parameter setting under main menu Enter key 2.Save the parameters and enter next menu Two operation modes for CCT-3300 series: ◆ measurement mode:normal display/transient display ◆ setting mode:for parameter setting switch diagram:...
  • Page 10: Mode Settings

    3.1 Mode settings Some parameters have been set before leaving factory. If the test environment changes(such as replacement of electrode, reset the alarm setting), please check the parameter which is in different menus. The specific content and operations as follows: Normal display Setting mode Electrode selection...
  • Page 11: Electrode Type Setting

    3.1.1 Electrode type setting When the meter shows “C=”, you could select electrode type. The electrode type could be circulated display for your selection by pressing . Choose your wanted type and press to save your settings and enter into the next menu setting automatically.
  • Page 12 Enter C=0.010, electrode constant value is sensor’ s label value X 10(range: 0.0801- 0.1199) Digital Digital C=0.100, electrode constant value is sensor’ s label value X 10(range: 0.801- input shifter 1.199) C=1.000 electrode constant value is sensor’ s label value X 10(range: 0.801-1.199)...
  • Page 13 Output voltage of voltage of mA mode terminal terminal cable Meter mode I+/I- >12V DC Transmitter T+/T- DC24V mode General fault inspection and trouble shooting: Possible factor phenomenon trouble shooting methods A. No power supply A. Check the wire connection of power No reading connection supply...
  • Page 14: Ⅴ.complete Set

    NOTE: High pure water and ultrapure water could not select open sampling measurement method. If the high pure water is exposed to the air, a lot of carbon dioxide would dissolve into the water immediately and the impurity inside the container and dusts in the air would also dissolve into the water which could cause serious errors.