PRODUCT INFORMATION The device features blood pressure measurement, pulse rate measurement and the result storage. Readings taken by this device are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method. This manual contains important safety and care information, and provides step by step instructions for using the product.
INDICATIONS FOR USE The LEICKE Sharon Blood Pressure Monitor is a digital monitor intended for measuring blood pressure and heartbeat rate with wrist circumferences ranging from 13.5 cm to 21.5 cm ( about 5¼ inches to 8½ inches ).
Page 6
CAUTION: * Intended for adult indoor use only. Pregnant women, neonatal patients, pre-eclamptic patients and patients with severe obesity must not use the device. If in need, please consult a professional doctor. * Intended for non-invasive measuring and monitoring of arterial blood pressure. Not intended for use on extremities other than wrist or for functions other than obtaining a blood pressure measurement.
LCD DISPLAY SIGNAL SYMBOL DESCRIPTION EXPLANATION Systolic Blood Pressure High blood pressure Diastolic High blood pressure Blood Pressure Beat/minute Pulse Low Battery Low battery - please charge Measurement unit of blood pressure mmHg Unit (1mmHg=0.133kPa) IHB Detector Irregular Heartbeat Detector Bluetooth Successful Bluetooth Connection Error...
Charging the Device 1. The battery of the monitor is a built-in rechargeable lithium-ion battery, the battery current is 420 mAh. Please use the USB cable to charge the battery, just like the following pictures: Method 1 Method 2 Charging the power under following circumstances: •...
CAUTION: 6. When charging, do not touch charging connector and patient simultaneously. 7. Do not attempt to replace your blood pressure monitor’s battery. It is built-in and not changeable. 8. Only charge the battery in accordance with the user instructions supplied with the blood pressure monitor.
Page 11
Setting Date and Time Please proceed to time setting before your initial use so as to ensure each piece of record is labelled with a time stamp. (Year Range: 2012-2052; Time Format: 12 Hours) 1. When the monitor is OFF, press and hold “SET”...
Page 12
Setting Date and Time 4. Repeat steps 2 and 3 to confirm [MINUTE]. 5. Repeat steps 2 and 3 to confirm [MONTH], [DAY] and [YEAR]. 6. After confirming [YEAR], the LCD will display “dONE” and then shut off.
Positioning the Cuff 1. Remove all accessories (watch, bracelet, etc.) from your left wrist. If your physician has diagnosed you with poor circulation in your left wrist, use your right wrist. 2. Roll or push up your sleeve to expose the skin. 3.
Page 14
PAIRING THE BLOOD PRESSURE MONITOR WITH YOUR MOBILE DEVICE Download and install the MedM Health APP to your mobile device which supports Bluetooth 4.0 technology from the APP Store or Google Play. 1. Turn on Bluetooth and the app. Make sure both are ON when pairing is proceeding.
Page 15
Start Measurement 1. After correctly positioning the cuff, press START/STOP button to turn on the monitor, and it will complete the measurement process. Adjust to zero. Inflating and measuring. Display and save the measuring result.
Page 16
2. This device will proceed to data transmission after measurement. The Bluetooth symbol blinking on the LCD indicates data transmission. 3. If the data is successfully transmitted, the LCD will display “dONE”. If data transmission fails, the LCD will display “ ”...
Recall the Records 1. Press “MEM” button to access the memory when the monitor is off. The monitor will display the calculated value of the latest readings first. 2. Press “MEM/UP” button or “SET/DOWN” button to rotate the history records. “MEM/UP”...
Delete the Records When you did not obtain the accurate measurement, you can clear all the measuring results by following the steps below. 1. Under Memory Recalling Mode, press and hold both the “MEM” button and the “SET” button for 3 seconds. 2.
Page 19
Tips for Measurement Measurements may be inaccurate if taken in the following circumstances. Immediate measurement Within 1 hour after tea, coffee, smoking after dinner or drinking When talking or moving Within 20 minutes your fingers after taking a bath When you need to pee urgently In a very cold environment...
Don’t open or repair the device by yourself. 3. Please report to LEICKE if any unexpected operation or events occur. 4. Cleaning: Dust environment may affect the performance of the unit. Please use a soft cloth to remove dirt from the device and cuff before and after use.
About Blood Pressure What are systolic pressure and diastolic pressure? When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
About Blood Pressure Irregular Heartbeat Detector: An irregular heartbeat is detected when the heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure. During each measurement, this equipment records the heartbeat intervals and works out the standard deviation. If the calculated value is larger than or equal to 15, the irregular heartbeat symbol appears when the measurement results are displayed.
Page 23
About Blood Pressure Why do I get a different blood pressure at home compared to the hospital? The blood pressure is different even throughout the day due to weather, emotion, exercise etc. Also, there is the “white coat” effect, which means blood pressure usually increases in clinical settings. What do you need to pay attention to when you measure your blood pressure at home: If the cuff is tied properly.
Page 24
Possible Cause/ Problem Description Symptom Possible Solution • Power is exhausted. Display is dim or Please charge. No Power will not light up. Shows on • Battery is low. the display Batteries Please charge. • Data communication has failed Shows •...
Page 25
Possible Cause/ Problem Description Symptom Possible Solution • Relax for a moment and Error 21 shows then measure again. Message • The treatment of the measurement failed. • EExx,shows on • Retake the measurement. the display. If the problem persists, •...
Page 26
EMC Guidance Table 1 Guidance and MANUFACTURER’s declaration – ELECTROMAGNETIC EMISSIONS- for all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer’s declaration – electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
EMC Guidance Table 2 Guidance and MANUFACTURER’s declaration – ELECTROMAGNETIC IMMUNITY – for all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment Electromagnetic IMMUNITY test...
EMC Guidance Table 4 Guidance and MANUFACTURER’s declaration – ELECTROMAGNETIC IMMUNITY – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Page 29
EMC Guidance Table 6 Recommended separation distances between portable and mobile RF communications equipment and the ME EQUIPMENT or ME SYSTEM – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the device. The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
FCC Interference Statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
Page 31
CONTACT If you have any questions, please feel free to contact us. PRODUCTION & WHOLESALE LEICKE GmbH Dohnanyistr. 28 04103 Leipzig Germany Web: www.leicke.com Mail: info@leicke.com...