Access to the programming menu
When pressed during the pump operation phase, it cyclically displays the programmed values on
the display; When pressed at the same time as the
dependent on the selected operating mode. During programming it carries out an "enter" function,
meaning that it confirms entry to the various menu levels and modifications within the same.
Starts and stops the pump. In the event of a level alarm (alarm function only), flow alarm and active
memory alarm, it deactivates the signal on the display.
Used to "exit" the various menu levels. Before definitively exiting the programming phase, you will be
asked if you wish to save any changes.
Access to the pump calibration menu. If in Off mode, the calibration menu is not activated.
Used to run upwards through the menu or increase the numerical values to be changed. Can be
used to start dosage in Batch mode
Used to run downwards through the menu, or decrease the numerical values to be changed.
Flashing green LED during dosage
Red LED that lights up in various alarm situations
B
EM00136063
Control Panel – TEKNA TPR
Electrical connections
rev. 1.0
keys, it increases or lowers a value
1
Alarm relay
2
3
Pole +
4
Pole -
5
Remote control input (start-stop)
6
7
Temperature probe input
8
9
Flow sensor input
10
B
Input level control
Exit 4-20 mA
500 Ω max load
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TEKNA TPR 600 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Seko TEKNA TPR 600

  • Page 1 Control Panel – TEKNA TPR Access to the programming menu When pressed during the pump operation phase, it cyclically displays the programmed values on the display; When pressed at the same time as the keys, it increases or lowers a value dependent on the selected operating mode.
  • Page 2: Setting Menu

    Programmiermenü Tekna TPR You can access the programming menu by pressing the key for over three seconds. The keys can be used to run through the menu items, with the key being used to access changes. The pump is programmed in constant mode in the factory. The pump automatically returns to the operating mode after 1 minute of no activity.
  • Page 3: Operation

    Setting the Language Programming Operation PROG Makes it possible to select the language. The pump is set in Pump Functions English in the factory. Max flow rate P100% Changes can be made by pressing the key, then using keys to set the new value. Press Alarm Relay N.Open confirm and return to the main menu...
  • Page 4: Display During Operation

    Paragrafo 2 – Dosaggio Proporzionale alla misura del pH (impostazione di fabbrica) Programming Operation The pump measures and controls the pH of a solution, programming sequence: set-point, set-point type, PROG proportional band and alarm band Set-point type: acid Configuration P roportional Pump Funtions Band flow-rate...
  • Page 5 Paragraph 3 – Dosage Proportional to the Potential Redox Measurement (O.R.P.) Programming Operation The pump measures and controls the pH of a solution, programming in sequence: set-point, set-point type, proportional band and alarm band Set-point type: maximum PROG P roportional Band flow-rate Configuration...
  • Page 6 Paragraph 5 – Setting the Alarm Relay Programming Operation PROG This is used to set the alarm relay in the absence of an alarm situation, if open (default) or closed. Pump Functions Changes can be made by pressing the key, then using Max flow rate P100% keys to set the new value.
  • Page 7 Paragraph 8 - Password Programming Operation PROG By entering the password, you can enter the programming menu and see all the set values. The password will be requested whenever you seek to modify them. The flashing line indicates the number than can be modified. Password Use the key to select the number (from 1 to 9), and the...
  • Page 8 pH Calibration Menu Pressing the CAL key for 3 seconds takes you into the calibration menu. If calibration was excluded during programming, the following appears on the display: Calibration If calibration is active: Calibration Calibration Automatic Cal. Wait..Automatic Automatic Buffer Sol.
  • Page 9: Manual Calibration

    Potential Redox Calibration Menu (O.R.P.) Pressing the CAL key for 3 seconds takes you into the calibration menu. If calibration was excluded during programming, the following appears on the display: Calibration If calibration is active: Calibration Calibration Automatic Cal. Wait..Automatic Automatic Buffer Sol.
  • Page 10 Alarms Display Cause Interruption Fixed alarm LED level alarm, without Restore the liquid level. Flashing word “Lev” interrupting pump operation I.e. P100% Fixed alarm LED End of level alarm, with interruption Restore the liquid level. Flashing words “Lev” and “stop” to pump operation I.e.