2.2 During Service ......................2-1 § 3 TROUBLESHOOTING ..............3-1 § 4 SUB TROUBLESHOOTING ............4-1 4.1 The LT-1000 does not Start..................4-1 4.2 The Color LCD Display is Abnormal................4-2 4.2.1 The back light does not light up............... 4-2 4.3 The Touch Panel is not Operational.
Page 4
§ 7 ADJUSTMENT ................7-1 7.1 Touch Panel Calibration .................... 7-1 7.2 Tracing Unit Calibration ..................... 7-1 7.2.1 Calibration with standard frames ..............7-1 7.2.2 Calibration with standard pattern ..............7-2 7.2.3 Torque calibration .................... 7-3 7.2.4 Full calibration ....................7-3 7.3 Displaying Parameter Setting Screen ................
LLT8A*RDA001A/E § 1 INTRODUCTION • This service manual is for the SATELLITE TRACER LT-1000. It does not include technical information for preparing lenses. • This manual is thoroughly understood before servicing. • Refer to the SATELLITE TRACER LT-1000 Operator’s Manual and Parts List.
• After part replacement, check that all parts are fixed in their original positions securely. • If the LT-1000 produces smoke or strange odors, immediately turn the power off and disconnect the electrical plug from the outlet. Usage of the instrument under such abnormal conditions may cause fire or electric shock.
Plug the power cord in the wall outlet and turn the power switch on. Does the LT-1000 start? 4.1 The LT-1000 does not start. Is the color LCD display normal? 4.2 The color LCD display is abnormal. Is the LT-1000 touch panel operational? 4.3 The touch panel is not operational.
LLT8A*RDA001A/E § 4 SUBTROUBLESHOOTING 4.1 The LT-1000 does not Start. Is the voltage from the wall outlet within the Use a wall outlet which meets the specifications. specifications? (100 V and 200 V systems: ±10%) Are the fuses blown? Replace the fuses (80402-02039).
LLT8A*RDA001A/E 4.2 The Color LCD Display is Abnormal. Does the back light light up? 4.2.1 The back light does not light up. Is there a break in the LCD cable (42203-E018)? Replace the LCD cable (42203-E018). Replace the color LCD module (40273-E020). Replace the BA01 board (43401-BA01).
LLT8A*RDA001A/E 4.3 The Touch Panel is not Operational. Is touch panel calibration possible? Is the symptom improved? I s t h e s i d e p l a t e ( 4 3 4 0 1 - M 0 0 4 ) grounded to the casing by the CA12 No cable (43401-CA12) and CA13 cable...
LLT8A*RDA001A/E 4.4 The Main Screen is not Displayed on the Color LCD. Does the inserted CF card contain the application Insert a CF card containing the application files files (It8a_autoload.bin and It8a_autoload.txt)? (It8a_autoload.bin and It8a_autoload.txt). Replace the BA01 board (43401-BA01). (Refer to 6.7.2.) No Use a different CF card on which the application...
LLT8A*RDA001A/E 4.5.1 Initialization does not start. Is the voltage between the P204 (J4) 1st and 6th Replace the BA02 board (43401-BA02). pins on the BA02 board (43401-BA02) DC + 24 (Refer to 6.7.3.) V? At this time, P204 is not connected. Is there a break in the CA06 cable (43401-CA06)? Replace the CA06 cable (43401-CA06).
LLT8A*RDA001A/E 4.5.3 No traced data is displayed. Is there a break in the CA06 cable (43401-CA06)? Replace the CA06 cable (43401-CA06). Is there a break in the CA07 cable (43401-CA07)? Replace the CA07 cable (43401-CA07). Are the DIP switches of the tracer ASSY. (43401- Set the DIP switches of the tracer ASSY.
LLT8A*RDA001A/E § 5 REMOVING PROCEDURE 5.1 Top Cover ASSY. 1. Unscrew SB3 × 6 (n = 4). 2. Disconnect the P501 (J1) connector from the BA05 board (43401-BA05). 3. Remove the top cover ASSY. (43401-2000). 4. Reassemble the removed parts in reverse order.
LLT8A*RDA001A/E 6.6 Inverter Replacement part: 42203-E017 1. Remove the LCD rear cover (43401-M307). (Refer to 5.3.) 2. Disconnect the CN2 connector from the color LCD module (40273-E020) and the CN1 connector from the inverter (42203-E017). 3. Unscrew PC2.6 × 4 (n = 2) to replace the inverter (42203-E017).
LLT8A*RDA001A/E 6.7.3 BA02 board Replacement part: 43401-BA02 1. Replace the switching power supply (80602- 00104). (Refer to 6.2.) 2. Disconnect the P201 (J1), P202 (J2), P203 (J3), and P204 (J4) connectors from the BA02 board (43401-BA02). 3. Unscrew the four attached screws. 4.
§ 7 ADJUSTMENT 7.1 Touch Panel Calibration 1. Push the touch panel calibration button on the rear side of the LT-1000 with a pencil point or such (φ6). Turn the power on while pressing the button. 2. A red square is first displayed in the upper- left corner of the screen as shown to the right.
LLT8A*RDA001A/E 5. Check the traced data (circumference) on the Circumference layout screen. Perform calibration again with the standard frames when the circumference is not in the range of 162.83 ±0.05 mm. 6. Remove the standard frames. 7.2.2 Calibration with standard pattern 1.
LLT8A*RDA001A/E 7.2.3 Torque calibration 1. Turn the power on while simultaneously pressing the right-eye and left-eye tracing buttons. 2. Confirm that the LED of the both-eye tracing button blinks. Open the upper and lower sliders fully. Press the both-eye tracing button while holding the sliders open.
Page 28
14. One short beep sounds when calibration is successfully completed. * Multiple beeps sound should calibration fail. The cause may be that the measured data or LT-1000 operation is in error. Check the inside of the tracer. 7 - 4...
LLT8A*RDA001A/E 7.3 Displaying Parameter Setting Screen 1. Press the MENU button on the layout screen. 2. The pop-up menu is displayed. 3. Pressing the Setting button. Select the screen Setting that includes the desired parameters. System Setting Network Adjustment Communication Maintenance 7 - 5...
2. Prepare the LT8A_AUTOLOAD.BIN and LT8A_AUTOLOAD.TXT files of the desired version and overwrite them in the CF card. 3. Insert the CF card and close the CF card slot cover (43401-M309). 4. Turn the power on. Check the software version after the LT-1000 software starts. 7 - 6...
1000. Cancel the error in the host PC. 8.5 Factory Setting of Tracing Unit Spacifications Machine Mode Protocol ID No. Server No. LT-1000 LT-1000 Meganet off-off-on-off * SW1 off-off-on-off: Communication rate 38400 bps off-off-off-off: Communication rate 9600 bps * SW2 on: The power is supplied from the first communication port (when the communication adapter is used).
LLT8A*RDA001A/E 8.6 Parameter Setting Table Screen Parameter name Factory setting Setting range ENGLISH Language ENGLISH is fixed. PTN Memory INTERNAL INTERNAL/EXTERNAL JOB# Input Interface KEYPAD TENKEY/KEYPAD PTN# Input Interface KEYPAD TENKEY/KEYPAD System Setting 1 Trace Data Auto Set ON/OFF VCA Data Display Rx/A,B VCA Layout Display NORMAL...