Buchi V-800 Instructions Manual

Buchi V-800 Instructions Manual

Vacuum controller

Advertisement

1
2
3
Function V-800
4
4.5 Installation with water jet jump B-767 with
5
Operation V-800
6
Functions V-805
7
Rotavapor R-200/R-205
7.6 Installation with water jet pump B-767
Read these instructions thoroughly before you
use the BÜCHI Vacuum Controller V-800/
V-805. Keep these instructions near the
apparatus, so that they are quickly accessible.
Chapter two contains important safety
instructions. This information is essential as
regards safe operation of the rotary evaporator.
Contents
2
8
Operation V-805
4
7
7
7
10
10
10
10
11
9
11
12
13
13
13
14
14
14
16
17
17
17
21
21
21
21
22
22
23
23
We reserve the right to make technical modifications without
prior notice. No part of these operating instructions can be
reproduced or processed, copied or distributed using electro-
nic or optical systems without permission in writing from BÜCHI
Labortechnik AG.
All rights reserved. © BÜCHI Labortechnik AG, 2000
en, Version H (50 pages)
V-800/V-805 Operating Instructions
Contents
24
24
24
25
25
25
27
29
30
31
32
33
34
36
38
39
39
40
43
44
48
Order number
96771
1

Advertisement

Table of Contents
loading

Summary of Contents for Buchi V-800

  • Page 1: Table Of Contents

    7.7 Installation with vacuum pump and valve unit Read these instructions thoroughly before you We reserve the right to make technical modifications without use the BÜCHI Vacuum Controller V-800/ prior notice. No part of these operating instructions can be V-805. Keep these instructions near the reproduced or processed, copied or distributed using electro- apparatus, so that they are quickly accessible.
  • Page 2: Scope Of Delivery

    BÜCHI Vacuum Controller V-800/V-805 1 Scope of Delivery Scope of Delivery Base apparatus V-800 Vacuum Controller 40736 230 V / 50-60Hz V-800 Vacuum Controller 40737 100/120 V / 50-60Hz V-805 Vacuum Controller 40738 230 V / 50-60Hz V-805 Vacuum Controller...
  • Page 3: Büchi Vacuum Controller

    BÜCHI Vacuum Controller V-800/V-805 1 Scope of Delivery Facultative accessories 1 Valve unit 37968 1 Holder for valve unit 40810 on Rotavapor R-200 /205 1 Remote control 40735 1 Communication set to 40235 rotary evaporator R-200/205 Remote control...
  • Page 4: Safety

    BÜCHI Vacuum Controller V-800/V-805 2 Safety Safety This apparatus has been built according to state-of-the-art and recognised safety rules. Still, it can be associated with risks and hazards: • If the apparatus is operated by insufficiently trained personnel; • If the apparatus is not used properly.
  • Page 5 BÜCHI Vacuum Controller V-800/V-805 2 Safety Improper use Applications not mentioned above are improper. Also, applications, which do not comply with the technical data are considered improper. The operator assumes sole risk for any damages, attributed to such improper use.
  • Page 6 These operating instructions must be available for operating personnel at the point of use of the apparatus as component of the Vacuum Controller V-800/V-805. This also applies to additional language versions of these instructions. These language versions can be ordered separately.
  • Page 7: Function

    BÜCHI Vacuum Controller V-800/V-805 3 Functions Functions V-800 Overview program structure V-800 BASIC STATE P Set MENU Choose hysteresis Enter Set Point MEMORY MODE SERVICE CONFIGURATION USER COMMUNI- MANUAL TIMER RECALL STORE DELETE SETTINGS CATION OFFSET CALIBRATION SERVICE TEST MESSAGE...
  • Page 8 BÜCHI Vacuum Controller V-800/V-805 3 Functions 3.2.4 Configuration - Communication In this menu, you defined whether: • Whether rotation of the evaporating flask is to start automatically when the controller starts (only in connection with Rotavapor R-200/R-205) • Whether the evaporating flask is to be lowered into the bath automatically when the controller starts •...
  • Page 9 BÜCHI Vacuum Controller V-800/V-805 3 Functions 3.2.6.2 Automatic hysteresis (dP = A) The hysteresis is automatically adjusted to the input set point and is 5 % of the set point, but a minimum of 2 mbar. Example: set point is 400 mbar, automatic hysteresis...
  • Page 10: Putting Into Operation Of

    BÜCHI Vacuum Controller V-800/V-805 4 Putting into operation Putting into operation of V-800 Look for damages when you unpack the unit. It is important that you detect any transport damages when you unpack. If necessary, you must prepare a status report immediately (informing postal company, railway company or transportation company).
  • Page 11: Installation With Water Jet Pump

    Screw on cooling water valve with sieve and seal to the water tap. Plug cable to the CW socket of the controller. Connect cooling water tubing. V-800/V-805 with Water jet pump B-764 Installation with water jet pump B-767 with integrated cooling water valve Screw on water jet pump with seal to the water tap.
  • Page 12: Installation With Vacuum Pump And Valve

    BÜCHI Vacuum Controller V-800/V-805 4 Putting into operation Installation with vacuum pump and valve unit Connect cable of valve unit with valve socket of the controller Connect control cable of the pump with pump socket of the controller Connect cable cooling water valve (optional) with CW socket of the controller Connect vacuum cover V.Contr.
  • Page 13: Operation

    BÜCHI Vacuum Controller V-800/V-805 5 Operation V-800 Operation V-800 Note that the apparatus must be properly put into operation in accordance with the instructions in Chapter 4, Putting into operation. Controls V-800 Main switch Function button, activates the respectively assigned...
  • Page 14: Fundamental Arrangement Of The Display

    BÜCHI Vacuum Controller V-800/V-805 5 Operation V-800 Fundamental arrangement of the display Information on active mode Set point Operating state Actual pressure Selected unit Graphical pressure display Illustration of functional buttons General information on control keys The [OK] key closes an input and puts the new value into the working memory.
  • Page 15 BÜCHI Vacuum Controller V-800/V-805 5 Operation V-800 5.5.2 Possible interventions into the regulation process [Pé] Increase the system pressure or interrupt pressure drop in the start phase. The apparatus goes into the hold mode and regulates further to the actual value. If necessary, the apparatus is partially ventilated.
  • Page 16: Configuration

    BÜCHI Vacuum Controller V-800/V-805 5 Operation V-800 5.5.5 Recall a program from the memory Recall the program from the memory: [MENU] ð [MEMORY] ð [RECALL ] ð [OK]. Choose the desired program number and confirm with [OK] . The display is changeing into the basic state automatically..
  • Page 17: Overview Of Program Structure

    BÜCHI Vacuum Controller V-800/V-805 6 Functions V-805 Functions V-805 Overview of program structure V-805 BASIC STATE P Set MENU Choose hysteresis Enter Set Point MEMORY CONFIGURATION SERVICE MODE USER SETTING COMMUNICATION MANUAL GRADIENT TIMER AUTODEST TEMPERATURE REPEAT AUTO. DEST. PARAMETER AUTO.
  • Page 18 BÜCHI Vacuum Controller V-800/V-805 6 Functions V-805 6.2.3 Distillation with timer function The set point of system pressure and distillation duration are input. When the distillation starts, the system pressure is reduced to the input value and kept constant during the selected period.
  • Page 19 BÜCHI Vacuum Controller V-800/V-805 6 Functions V-805 6.2.9 Configuration In this menu, you define: • Whether the rotation of the evaporator is to automatically start when the controller starts (only in connection with Rotavapor R-200/R-205) • Whether the evaporating flask is to be lowered into the bath automatically when the controller starts •...
  • Page 20 6 Functions V-805 BÜCHI Vacuum Controller V-800/V-805 6.2.11.2 Automatic hysteresis (dP = A) The hysteresis is automatically adjusted to the input set point and is 5 % of the set point, but a minimum of 2 mbar. Example: set point is 400 mbar, automatic hysteresis...
  • Page 21: Putting Into Operation

    BÜCHI Vacuum Controller V-800/V-805 7 Putting into operation Putting into operation V-805 Look for damages when you unpack the unit. It is important that you detect any transport damages when you unpack. If necessary, you must prepare a status report immediately (informing postal company, railway company or transportation company).
  • Page 22: Instalation Of Automatic Dual Temperature Sensor

    7 Putting into operation BÜCHI Vacuum Controller V-800/V-805 Installation and adjusting of automatic dual temperature sensor Insert the long end of the temperature sensor into the condenser, through the screw neck melted to the top of the condenser between the inner condenser coil and centrally melted vacuum tube in the condenser.
  • Page 23: Installation With Water Jet Pump

    BÜCHI Vacuum Controller V-800/V-805 7 Putting into operation Installation with water jet pump B-767 (with integrated cooling water valve) Screw on water jet pump together with seal to the water tap. Plug cable to valve socket of the controller. Connect water jet pump to the rotary evaporator via Woulff bottle.
  • Page 24: Operation

    8 Operation V-805 BÜCHI Vacuum Controller V-800/V-805 Operation V-805 Note that the apparatus must be properly put into operation in accordance with the instructions in Chapter 7, Putting into operation. Controls V-805 Main switch Control keys, activating assigned function Key for permanent vacuum, turns the pump 100 % on...
  • Page 25: Fundamental Arrangement Of The Display

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 Fundamental arrangement of the display Information on active mode Set point Operating state Actual pressure Selected unit Graphical pressure display Illustration of control keys General information on the control keys The key [OK] closes an input and puts the new value into the working memory.
  • Page 26: Büchi Vacuum Controller V-800/V

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 8.5.2 Possible interventions into the regulating process [Pé] Increase the system pressure or interrupt pressure drop in the start phase. The apparatus goes into the hold mode and regulates further to the actual value. If necessary, the apparatus is partially ventilated.
  • Page 27: Distillation With Pressure Gradient

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 Distillation with pressure gradient 8.6.1 Setting of the set point and Start, preparing a new program Choose gradient mode (if it is not already active) [MENU]ð[MODE]ð[GRADIENT]ð[PROG]ð[NEW]. The first segment can now be programmed. Each segment is defined by a start pressure, its end pressure and the duration.
  • Page 28 BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 8.6.2 Possible interventions in the regulating process (gradient mode) Increase the system pressure or interrupt pressure drop in the start phase ð [Pé]. The apparatus goes into the hold mode and regulates further to the actual value. If necessary, the apparatus is partially ventilated.
  • Page 29: Distillation With Timer

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 Distillation with timer 8.7.1 Setting the set point and start Choose timer mode (if it is not already active) [MENU] ð [MODE] ð [TIMER] ð [PROG] Define distillation time. Use [é] to increase the actual value and [ê] to decrease the actual value.
  • Page 30: Automatic Distillation

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 Automatic distillation (only with connected auto sensor) 8.8.1 Preconditions for an automatic distillation The automatic dual temperature sensor is based on a balance adjustment of the temperature in the condenser with consideration of the cooling material temperature. For trouble- free distillation until dryness, the following conditions must exist: •...
  • Page 31: Distillation At Constant Vapour Temperature

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 8.8.3 Possible interventions into the regulating process Increase the system pressure or interrupt pressure drop in the start phase ð [Pé]. he apparatus goes into the hold mode and regulates further to the actual value. If necessary, the apparatus is partially ventilated.
  • Page 32: Repeating The Last Distillation Performed

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 8.9.2 Possible interventions into the regulating process Increase the system pressure or interrupt pressure drop in the start phase ð[Pé]. The apparatus goes into the hold mode and regulates further to the actual value. If necessary, the apparatus is partially ventilated.
  • Page 33: Configuration

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 8.10.2 Possible interventions into the regulating process Increase the system pressure or interrupt pressure drop in the start phase ð [Pé]. he apparatus goes into the hold mode and regulates further to the actual value. If necessary, the apparatus is partially ventilated.
  • Page 34: Memory Of Methods

    BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 8.12 Method memory 8.12.1 General information Gradient programs and time programs can be saved in the methods memory. Manual distillation can only be saved via the repeat function. The save function . The save function includes conventional functions like saving, opening and deleting.
  • Page 35 BÜCHI Vacuum Controller V-800/V-805 8 Operation V-805 8.12.4 Deleting a program from the memory Call up memory function [MENU] ð [MEMORY] ð [DELETE] ð [OK]. Choose desired program number and confirm with [OK]. Back to basis state ð [ESC] or other program according...
  • Page 36: Information On Choosing Distillation Conditions

    BÜCHI Vacuum Controller V-800/V-805 9 Distillation conditions Information on choosing distillation conditions To achieve optimum distillation conditions when rotary evaporators are used, the energy, supplied for distillation from ∆T the heating bath, must be removed via the condenser. To (min. 20°C) guarantee this, operations should be performed according to this rule of thumb.
  • Page 37 BÜCHI Vacuum Controller V-800/V-805 9 Distillation conditions Solvent table Solvent Formula Molar Mass Evaporation Boiling point Density Vacuum in mbar for in g/mol energy in J/g at 1013 mbar in g/cm boiling point at 40°C Acetone C3H6O 58.1 0.790 n-amylalcohol, n-pentanol C5H12O 88.1...
  • Page 38: Control Function In Connection With Rotavapor

    10 Control function in connection with Rotavapor R-200/R-205 10.1 Start of rotation The Vacuum Controller V-800/V-805 makes it possible to automatically start rotation of the evaporating flask. This function is activated and deactivated in the configuration menu. The rotation speed corresponds to the respective rotary button setting of the rotary evaporator.
  • Page 39: The Ventilation Valve

    BÜCHI Vacuum Controller V-800/V-805 11 Ventilation valve/12 Remote control 11 The ventilation valve Automatic ventilation With standard configuration ex factory, when the controller stops, the connected apparatus is ventilated immediately and the valve remains open even after the apparatus has been turned off.
  • Page 40: Maintenance

    BÜCHI Vacuum Controller V-800/V-805 13 Maintenance 13 Maintenance You must observe all instructions, aimed at keeping the Rotavapor functional. This includes periodically cleaning and checking for damages. 13.1 Cleaning Attention: Before cleaning, make sure that the apparatus is unplugged from the current mains.
  • Page 41 BÜCHI Vacuum Controller V-800/V-805 13 Maintenance 13.4 Customer service Only authorised service personnel can perform work on or in the apparatus. These persons have a comprehensive technical training and knowledge about hazards, resulting when safety stipulations are not complied with. BÜCHI Customer Service Office have apparatus specific Service Manual, which can only be obtained by authorised personnel.
  • Page 42 BÜCHI Vacuum Controller V-800/V-805 13 Maintenance 13.6 Troubleshooting Malfunction Possible cause Remedy No display No current to apparatus Turn mains switch on Check mains connections Frequent switching of valve or pump System is untight Control all sealing points (tubes and...
  • Page 43: Taking Out Of Operation

    BÜCHI Vacuum Controller V-800/V-805 14 Taking out of operation 14 Taking out of operation You must remove dangerous substances and thoroughly clean the apparatus. As a result, people will not be injured from contact with dangerous substances. 14.1 Storage/transportation Clean the apparatus thoroughly. Residues of chemicals must be removed completely and glass components must be cleaned.
  • Page 44: Spare Parts And Accessories

    BÜCHI Vacuum Controller V-800/V-805 15 Spare parts and accessories 15 Spare parts and accessories Only BÜCHI original accessories and spare parts guarantee safety and that the apparatus functions correctly. You can only use spare parts and accessories from other companies if BÜCHI consents.
  • Page 45 BÜCHI Vacuum Controller V-800/V-805 15 Spare parts and accessories Name Order number Water jet pump B-767 31357 Set with 5 sieves and 10 seals 15860 1 5 8 6 0 Kalrez PP membrane 23347 Set of four O rings 37607...
  • Page 46 BÜCHI Vacuum Controller V-800/V-805 15 Spare parts and accessories Name Order number Valve unit, complete 37968 The valve unit combines the function of the Woulff bottle with that of the vacuum valve. Valve unit, complete Name Order number Automatic dual temperature sensor...
  • Page 47 BÜCHI Vacuum Controller V-800/V-805 15 Spare parts and accessories Name Order number Water control valve ½” complete (01308 + 03560 + 15860) 11606 15860 Set with 5 sieves and 10 seals 15860 Water control valve, complete Name Order number Vacuum tube 16/6 mm...
  • Page 48: Annex

    BÜCHI Vacuum Controller V-800/V-805 16 Annex 16 Annex 16.1 Technical data V-800 / V-805 Dimensions casing (W x L x D) 110 x110 x 70 mm Weight 0.9 kg Electrical supply 230 V / 50/60 Hz 100 - 120 V / 50/60 Hz...
  • Page 49 BÜCHI Vacuum Controller V-800/V-805 16 Annex 16.3 Declaration of conformity BÜCHI Labortechnik AG Postfach, CH-9230 Flawil, Switzerland Declare, solely under our responsibility, that the product: BÜCHI Vacuum Controller V-800/V-805 On which this declaration is based, complies with the following standards:...
  • Page 50: English

    BÜCHI Vacuum Controller V-800/V-805 16 Annex 16.4. FCC Deklaration English: This equipment has been tested and found to comply with the limits for a Class A digital device, pusuant to both Part 15 of the FCC Rules and the radio interference regulations of the Canadian Department of Communications. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.

This manual is also suitable for:

V-805

Table of Contents